10.

 

After the centrifuge rotor has come to a complete stop (following the end of 
the run bell), open the lid. 

 

Safety 

Your PowerSpin™ LX centrifuge is a carefully designed and built instrument.  When 
operated properly, it presents no safety hazards.  However, please take note of these safety 
precautions. 

 

Lid 

NEVER OPEN THE LID UNTIL THE ROTOR HAS COME TO A 
COMPLETE STOP! 

 

Load Balance  Always make sure that the load is balanced before operating 

the instrument

.  

      

It is very important that test tubes with equal loads are placed opposite each other in the 
rotor.  Also, when spinning a single specimen tube, another specimen tube of the same 
size filled with water must be installed directly across from the specimen tube to maintain 
balance.  If the centrifuge is not properly balanced, damage could occur.  If excessive 
noise or vibration is noticed during a run, stop the centrifuge and recheck the balance. 

 
Maintenance 

This centrifuge is designed to be relatively maintenance free.  The motor is a brushless 
type and has permanently sealed bearings so oiling is not necessary.  A surface cleaner 
such as Formula 409 can be used to clean the outside of the instrument.  The rotor 
chamber and cover can be disinfected with a mild solution of bleach or other disinfecting 
liquid and allowed to dry.  Do not drench or immerse the unit in any liquid! Be sure to 
disconnect the power before attempting to clean. 
 

Routine Rotor Inspection and Maintenance 

All centrifuge rotors should be removed and inspected for damage, defects or other 
problems: 

-

 

Before initial use.

 

-

 

Each time the rotor is removed for cleaning.

 

-

 

At least every 6 months.

 

 
Plastic rotors and tube shields should be replaced every 12 months (once each year). 
 

What to do if your inspection finds rotor or tube shield damage, defects or other 
problems?  

If you detect any problems or damage, or defects in your centrifuge rotor or 

tube shields discontinue use of the centrifuge and rotor immediately!  

Do not under any 

circumstances continue to use a rotor that you think has damage

 as it could create a 

safety problem.   

 

Contact support at HENRY SCHEIN immediately at 800-588-9776 or 732-274-1155.  
Please be prepared with the serial # (found on the rear of the machine near the electrical 
plug) and any information you may have about where and when it was purchased.   

 
 

Содержание PowerSpin 112-6907

Страница 1: ...e rotor Item 112 6909 with 6 place rotor OPERATION MANUAL Revision Date 2014 03 Distributed by in US only manufacturer HENRY SCHEIN INC 135 Duryea Road Melville NY 11747 USA Henry Schein U K Holdings Ltd Medcare House Centurion Close Gillingham Business Park Gillingham ME8 0SB U K ...

Страница 2: ...es Can spin 2 ml and 4 5 ml tubes with available adapters Linear speed control allows use from 500 rpm 4 000 rpm Dependable mechanical 30 minute timer with bell signal at end of run High impact plastic housing for years of dependable use Safety switch in lid shuts off motor when lid is unlatched Latch release located on the right front side of unit Power indicator light tells when unit is powered ...

Страница 3: ...ing styrofoam disc is packed inside the lid of the centrifuge to prevent the rotor or shields from moving during shipping REMOVE THIS PACKING DISC PRIOR TO OPERATION OF THE CENTRIFUGE The centrifuge rotor is pre installed No rotor installation and calibration is required The power inlet is located in the back of the centrifuge Plug in one end of the power cord to the inlet and the other end to a p...

Страница 4: ...ng is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfecting liquid and allowed to dry Do not drench or immerse the unit in any liquid Be sure to disconnect the power before attempting to clean Routine Rotor Inspection and Maintenance All centrifuge ro...

Страница 5: ...ge Do not operate with a damaged cord Biohazard If a tube breaks a biological hazard may exist Unplug the instrument Clean using approved laboratory procedures Always wear the proper protective clothing and gloves Use a suitable disinfectant and dispose of the sample properly Refer to the guidelines set forth by OSHA CLIA etc Service If you feel that your centrifuge needs factory service please ca...

Страница 6: ... turned on b Lamp needs to be replaced c Fuse is blown 4 Fuse Replacement CAUTION DISCONNECT THE POWER CORD BEFORE REPLACING FUSE Follow the procedures below to replace the fuse a Unplug the power cord Turn the on off switch to off b Remove the power cord from the power inlet on the back of the centrifuge c Locate the fuse holder The fuse holder is a part of the power inlet On off switch Inlet Fus...

Страница 7: ...__________500 4 000 rpm Speed accuracy 100 rpm 3000 RPM Width________________________________10 2 inches Depth________________________________11 8 inches Height________________________________8 inches Weight________________________________18 lbs ...

Отзывы: