After the centrifuge rotor has come to a complete stop (following the end of
the run bell), open the lid.
Safety
Your PowerSpin™ LX centrifuge is a carefully designed and built instrument. When
operated properly, it presents no safety hazards. However, please take note of these safety
precautions.
Lid
NEVER OPEN THE LID UNTIL THE ROTOR HAS COME TO A
COMPLETE STOP!
Load Balance Always make sure that the load is balanced before operating
the instrument
.
It is very important that test tubes with equal loads are placed opposite each other in the
rotor. Also, when spinning a single specimen tube, another specimen tube of the same
size filled with water must be installed directly across from the specimen tube to maintain
balance. If the centrifuge is not properly balanced, damage could occur. If excessive
noise or vibration is noticed during a run, stop the centrifuge and recheck the balance.
Maintenance
This centrifuge is designed to be relatively maintenance free. The motor is a brushless
type and has permanently sealed bearings so oiling is not necessary. A surface cleaner
such as Formula 409 can be used to clean the outside of the instrument. The rotor
chamber and cover can be disinfected with a mild solution of bleach or other disinfecting
liquid and allowed to dry. Do not drench or immerse the unit in any liquid! Be sure to
disconnect the power before attempting to clean.
Routine Rotor Inspection and Maintenance
All centrifuge rotors should be removed and inspected for damage, defects or other
problems:
-
Before initial use.
-
Each time the rotor is removed for cleaning.
-
At least every 6 months.
Plastic rotors and tube shields should be replaced every 12 months (once each year).
What to do if your inspection finds rotor or tube shield damage, defects or other
problems?
If you detect any problems or damage, or defects in your centrifuge rotor or
tube shields discontinue use of the centrifuge and rotor immediately!
Do not under any
circumstances continue to use a rotor that you think has damage
as it could create a
safety problem.
Contact support at HENRY SCHEIN immediately at 800-588-9776 or 732-274-1155.
Please be prepared with the serial # (found on the rear of the machine near the electrical
plug) and any information you may have about where and when it was purchased.