Unpacking and Installation

 

The centrifuge is first packed in a styrofoam and then put in a carton box for shipping. 
Please examine the shipment upon receipt and retain all packing until the unit has been 
tested.  The warranty excludes any damage caused by shipping. 
 
Inspect for any possible shipping damage.  If you note any damage, please contact 
HENRY SCHEIN.  Save the original packaging.  If it becomes necessary to return the 
instrument to us, we ask that it be returned in the original packaging. 
 
Unpack the shipping carton carefully and verify that the shipment is complete.  It should 
include the following items: 
 
 

PowerSpin™ centrifuge (LX)  

 

1 unit 

 

Power cord 

 

 

 

 

1 pc. 

 

Operation manual (LX version) 

 

1 pc. 

 
No on-site installation is required.  Follow simple procedures below to set the centrifuge 
up: 

 

Take the centrifuge from the styrofoam and place it on a sturdy flat surface. 

 

Press the lid latch release button located on the right side and open the centrifuge 
cover.  A special packing styrofoam disc is packed inside the lid of the centrifuge to             
prevent the rotor or shields from moving during shipping.  REMOVE THIS 
PACKING DISC PRIOR TO OPERATION OF THE CENTRIFUGE 

 

The centrifuge rotor is pre-installed.  No rotor installation and calibration is required. 

 

The power inlet is located in the back of the centrifuge.  Plug in one end of the power 
cord to the inlet and the other end to a power outlet.  

 

Turn the speed control counterclockwise to “0” rpm. 

 

CAUTION

Make sure the voltage of the power source matches the voltage input of 
centrifuge.    

 
Operation 

1.

 

Set the timer control to “off”. 

2.

 

Plug the centrifuge into a power source outlet. Make sure the voltage is correct. 

3.

 

Turn the power rocker switch in rear of unit to “on” position. 

4.

 

Open the lid by pressing the lid-opening button. 

5.

 

Place the specimen tubes into the tube shields of the rotor. 

 
 

CAUTION:  Make sure that the specimen tubes are balanced; see safety 

section below

.

 

 

6.

 

Close the lid firmly. 

7.

 

Set the speed control knob at or close to “0” rmp. 

8.

 

Set the timer clockwise to the desired time.  

9.

 

Rotate the speed control knob slowly to the appropriate speed. 
For example, Blood at 3,400 or Urine at 2,000- 2200 rpm. 

Содержание PowerSpin 112-6907

Страница 1: ...e rotor Item 112 6909 with 6 place rotor OPERATION MANUAL Revision Date 2014 03 Distributed by in US only manufacturer HENRY SCHEIN INC 135 Duryea Road Melville NY 11747 USA Henry Schein U K Holdings Ltd Medcare House Centurion Close Gillingham Business Park Gillingham ME8 0SB U K ...

Страница 2: ...es Can spin 2 ml and 4 5 ml tubes with available adapters Linear speed control allows use from 500 rpm 4 000 rpm Dependable mechanical 30 minute timer with bell signal at end of run High impact plastic housing for years of dependable use Safety switch in lid shuts off motor when lid is unlatched Latch release located on the right front side of unit Power indicator light tells when unit is powered ...

Страница 3: ...ing styrofoam disc is packed inside the lid of the centrifuge to prevent the rotor or shields from moving during shipping REMOVE THIS PACKING DISC PRIOR TO OPERATION OF THE CENTRIFUGE The centrifuge rotor is pre installed No rotor installation and calibration is required The power inlet is located in the back of the centrifuge Plug in one end of the power cord to the inlet and the other end to a p...

Страница 4: ...ng is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfecting liquid and allowed to dry Do not drench or immerse the unit in any liquid Be sure to disconnect the power before attempting to clean Routine Rotor Inspection and Maintenance All centrifuge ro...

Страница 5: ...ge Do not operate with a damaged cord Biohazard If a tube breaks a biological hazard may exist Unplug the instrument Clean using approved laboratory procedures Always wear the proper protective clothing and gloves Use a suitable disinfectant and dispose of the sample properly Refer to the guidelines set forth by OSHA CLIA etc Service If you feel that your centrifuge needs factory service please ca...

Страница 6: ... turned on b Lamp needs to be replaced c Fuse is blown 4 Fuse Replacement CAUTION DISCONNECT THE POWER CORD BEFORE REPLACING FUSE Follow the procedures below to replace the fuse a Unplug the power cord Turn the on off switch to off b Remove the power cord from the power inlet on the back of the centrifuge c Locate the fuse holder The fuse holder is a part of the power inlet On off switch Inlet Fus...

Страница 7: ...__________500 4 000 rpm Speed accuracy 100 rpm 3000 RPM Width________________________________10 2 inches Depth________________________________11 8 inches Height________________________________8 inches Weight________________________________18 lbs ...