20/44 Assembly
Hengst Filtration GmbH
, OPM II, RE 51460-B/04.2021
Observe the ΔP of the online particle monitor dependent on the viscosity of the
hydraulic fluid.
Fig. 7: Δp-Q characteristic curve for different viscosities
Based on this, you can estimate the required pressure level for the required flow
of 50 ... 400 ml/min.
▶
Identify a place of installation complying with the criteria specified under chap-
ter 7.2.2 "Installation position".
▶
Depressurize the system.
▶
Connect two fluid lines to the two threaded couplings.
▶
Fasten the online particle monitor using the mounting points, see fig. 6:
Dimensions, on the device back side.
7.3.2 Electrically connecting the online particle monitor
WARNING
Faulty energy supply!
Danger to life! Risk of injury!
▶
Always observe country-specific regulations.
▶
Design the voltage supply according to EN50178, SELV, PELV, VDE0100-410/A1.
▶
Use the Y distributor if a plug-in power supply unit and an USB CAN adapter are
to be connected at the same time.
More information on the accessories for the online particle monitor is contained
in data sheet 51460, see chapter 1.2 "Required and amending documentation".
▶
De-energize the system for the installation and connect the online particle moni-
tor as described in the following.
Estimation of the required
pressure level
Procedure