82
360093-1/H
Platelet Incubator Operation Manual
9.3.1.7
Batteries
►
(i.Series models only) Six 1.5 V D-cell non-rechargeable alkaline (or equivalent) batteries are required
for backup power to the i.Center monitoring system
►
(Horizon Series models only) One 9 V non-rechargeable alkaline (or equivalent) battery is required for
backup power to the power failure alarm
►
(PC4200 models only) One 9 V rechargeable Nickel-Metal Hydride (NiMH) battery (or equivalent) is
required for backup power to the motion alarm system
►
(Models with a chart recorder only) One 9 V non-rechargeable alkaline (or equivalent) battery is
required for backup power to the chart recorder
9.3.2
Weight
PC100h,
PC100i
PC900h,
PC900i
PC1200h,
PC1200i
PC2200h,
PC2200i
PC3200h,
PC3200i
PC4200h,
PC4200i
105 lb
159 lb
208 lb
363 lb
431 lb
663 lb
48 kg
72 kg
94 kg
165 kg
196 kg
301 kg
9.3.3
Size
All dimensions are for the overall exterior and include items that protrude from the main unit, unless
specified otherwise.
Model
Width
Height
Depth
PC100h, PC100i
21.25 in (54 cm)
25.00 in (64 cm)
23.50 in (60 cm)
PC900h, PC900i
26.50 in (67 cm)
30.25 in (77 cm)
27.75 in (71 cm)
PC1200h, PC1200i
40.75 in (104 cm)
30.25 in (77 cm)
27.75 in (71 cm)
PC2200h, PC2200i
40.25 in (103 cm)
60.00 in (153 cm)
30.25 in (77 cm)
PC3200h, PC3200i
40.25 in (103 cm)
75.50 in (192 cm)
30.25 in (77 cm)
PC4200h, PC4200i
40.25 in (103 cm)
75.50 in (192 cm)
30.25 in (77 cm)
9.3.4
Operating conditions
This agitator is designed for indoor use only.
Altitude (maximum): 2000 m
Ambient temperature range: 15 °C to 35 °C
Relative humidity (maximum for ambient temperature): 80%
Temperature control range: 20 °C to 35 °C
9.3.5
Agitation speed
On PC4200 models, the agitation speed varies by input voltage and frequency. The speed tolerance is ±10% of
the nominal speed.
Model
Input voltage and
frequency
Agitation speed
(cycles per minute)
PC4200i, PC4200h
115 V 60 Hz
60
230 V 50 Hz
60
230 V 60 Hz
72