HM1734-WM STRAIN GAGE INPUT MODULE
HM1734-WM AOP rev418. ACD
CALIBRATING WITH KNOWN LOAD (AUTO-CAL) Ver. 4.18
1)
Set known weight for channel at scale parameter.
2)
Tare-0 (with no weight on cell/scale).
3)
Set module to AUTOCAL mode (BIT).
4)
Apply known load (test weight) to load cells/ scale.
5)
Set AUTO TUNE BIT on for CH1 (CH2 if applicable).
6)
To ensure accuracy repeat the steps above.
7)
MAKE SURE AUTOCAL MODE BIT STAYS ON.