Table 5-1
Failure symptoms, causes, and solutions
Symptom
Possible causes
Solution
Output voltage = 0 V
11.5 to 26.5 V power supply cable is
improperly connected or faulty.
Repair or replace power supply cable
(see page 10).
Pressure reading is too high.
• Conductance in connection to
vacuum chamber is inadequate.
• Plumbing to module leaks or is
contaminated.
• Chamber pressure is too high due to
leak, contamination, or pump
failure.
• Power supply or output cable is
improperly connected or faulty.
• If conductance is inadequate,
reconnect Convectron gauge port to
vacuum chamber (see page 9).
• If plumbing leaks or is
contaminated, clean, repair or
replace plumbing.
• If pump failed, repair or replace it.
• If cable is improperly connected or
faulty, repair or replace cable (see
page 10).
Pressure reading is inaccurate.
• Module is not calibrated for the
process gas that is being used.
• Module is not mounted horizontally.
• Convectron gauge or differential
pressure sensor is damaged (for
example, by reactive gas) or
contaminated.
• Temperature or mechanical
vibration is extreme.
• If Convectron gauge is out of
calibration, recalibrate it for the
process gas that is being used (see
page 34 or page 56).
• If module is not mounted
horizontally, re-mount it (see
page 8).
• If Convectron gauge is damaged,
replace it (see page 67).
• If Convectron gauge is
contaminated, return it to factory
(see pages 68 and 70).
• If temperature or vibration is
extreme, relocate module or
eliminate source of heat or
vibration.
Indicated pressure is different than
pressure indications from other
measurement devices.
• Process gas is a not the gas that the
user anticipated using in the system.
• Convectron gauge is defective.
• If the process gas is not what was
anticipated, calibrate Convectron
gauge for gas that is being used (see
page 34 or page 56).
• If Convectron gauge is defective,
return it to factory (see pages 68
and 70).
Relay LED indicator is ON, but relay is
not functioning.
Relay contacts are defective.
Make sure relay load is within
specified rating and is non-inductive
(see page 74).
• Relay will not activate.
• Output voltage is < 0.10 V.
• A circuit board is faulty.
• Convectron gauge sensing wire is
open.
Return module to factory (see
Содержание GRANVILLE-PHILLIPS Convectron ATM
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 12: ...4 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 80: ...72 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 90: ...82 Convectron ATM Vacuum Gauge Module Instruction Manual ...
Страница 97: ......