8
Description of the Hardware
2.3.2
Rear-panel
Figure 2.3:
EPC 10 USB
Rear Panel
Power:
Connect the enclosed power cord to this connector. It is not necessary to switch between different voltage
ranges manually. The built-in power supply detects and covers the two major voltage ranges of 90-120 V and
210-250 V at 50-60 Hz.
Digital Out / Digital In:
TTL-level, digital input and output lines are available here for the control and
monitoring of digital signals (see chapter 12.2 on page 62).
Digital I/O:
This connector can be used to connect an EPC 8 patch clamp amplifier or a TIB 14 trigger interface
box to the
EPC 10 USB
amplifier.
Trigger:
This port does not exist any more. Use
Trigger In
on the front panel of the
EPC 10 USB
.
USB:
This port is the connection to the USB 2.0 port in the host computer, that allows the computer to commu-
nicate with the
EPC 10 USB
.
Slave Sync / Master Sync:
2 CAT5 connectors for synchronization of a second amplifier (
EPC 10 USB
only)
or interface (LIH 8+8 only) system.
AUX DAC / Telegraphing Inputs:
These ports are for inspection purposes only.
Sound:
A 3.5 mm jack for connecting phones or speakers. The volume and resistance/frequency ration can be
adjusted by the
Patchmaster
software.
2.4
EPC 10 USB DOUBLE, TRIPLE, and QUADRO
The
EPC 10 USB
Double, Triple, and Quadro contain two, three and four independent
EPC 10 USB
amplifiers
in one case, respectively. All amplifiers share a single AD/DA board. The AD/DA board provides 4 DA channels
which are sufficient to allow simultaneous stimulation of every individual amplifier. The 8 available AD channels
can independently read all voltage- and current-outputs of the individual amplifiers.
The number of freely available DA and AD channels for the different
EPC 10 USB
patch clamp amplifiers are
listed in the table below.
Amplifier
DA channel
AD channel
EPC 10 USB SINGLE
3
5
EPC 10 USB DOUBLE
2
3
EPC 10 USB TRIPLE
1
1
EPC 10 USB QUADRO
-
-
Table 2.1: Number of free DA/AD channels
One amplifier is always selected as the ”active” amplifier. The ”active” amplifier is the one which is enabled to
http://www.heka.com
Содержание EPC 10 USB
Страница 1: ...Hardware Manual Version 2 8 EPC 10 USB Computer controlled Patch Clamp Amplifier...
Страница 6: ......
Страница 10: ...4 Introduction http www heka com...
Страница 16: ...10 Description of the Hardware http www heka com...
Страница 22: ...16 Installation http www heka com...
Страница 32: ...26 Verifying and Testing the EPC 10 USB http www heka com...
Страница 44: ...38 The control software http www heka com...
Страница 48: ...42 Operating Modes http www heka com...
Страница 54: ...48 Compensation Procedures http www heka com...
Страница 58: ...52 Patch Clamp Setup http www heka com...
Страница 64: ...58 Using the Patch Clamp http www heka com...
Страница 74: ...68 Appendix II Probe Adapters http www heka com...
Страница 76: ...70 Appendix III S Probe http www heka com...
Страница 81: ......