Heinrichs PIT Series Скачать руководство пользователя страница 10

Heinrichs Messtechnik GmbH 

 

 

 

        Operating Manual PIT-UMF2 (B) 

Page 10 of 14 

Power supply / electrical connection 

See name plate or Operating Instructions of the corresponding transmitter. 

CE-Mark 

The  measuring  system  complies  with  the  legal  requirements  of  the  following  EU  Directives:  Low 
Voltage Directive 2006/95/EC, Electromagnetic Compatibility Directive 2004/108/EC  
 
Heinrichs Messtechnik confirms compliance with the directives by attaching the CE mark to the device. 
 

Standards and directives, certificates and approvals 

Certified to DIN-EN-ISO 9001:2008 
Production in accordance with AD2000 guidelines 
TÜV approval for welding requirements in accordance with EN 3834-2 

 
General standards and directives 

EN 61010  

– Safety requirements for electrical measuring, control and laboratory devices  

EN 60529:2010 

– Degrees of protection through housing (IP code) 

 

EMC Directive 2004/108/EC

 

EN 61000-6-2:2005  

 

immunity for industrial environments  

EN 61000-6-3:2007+A1:2011   emissions residential environments 
EN 55011:2009+A1:2010  

group 1, class B (emitted interference) 

 

10  Safety information 

10.1  Intended use 

The  PIT-UMF2  (B) flow  meter  may  be  used  only  for  flow  measurements  of fluids  whose  conductivity 
exceeds 20µS/cm. The manufacturer shall not be liable for damages that may result from unintended 
or inappropriate use.

 

 
When dealing with an aggressive medium, clarify the material durability of all wetted parts.   
 

10.2  Installation, start-up and operating personnel 

Only  trained  specialists  authorized  by  the  system  operator  may  carry  out  the  installation,  electrical 
installations,  start-up  maintenance  and  operation.    They  must  read  and  understand  the  operating 
manual and follow its instructions. Basically, the national conditions and provisions must be followed. 
 

11  Packaging, storage and transport 

Carefully unpack the device to avoid damaging it. 
 
Storage and  installation must be  done in a clean and dry room so that contamination 

– especially of 

the interior of the fitting 

– is avoided. The ambient temperature ranges must be observed.  

 
With  the  help  of  the  delivery  note  enclosed  in  the  packaging,  check  whether  all  technically  relevant 
data coincide with your requirements. 
 
When  transporting  the  device  to  a  remote  mounting  location,  we  recommend  that  you  reuse  the 
factory-issued packaging and the transport protection. 
 

12  Maintenance 

The device requires no maintenance if used according to its intended purpose.  Cleaning might be 
necessary due to deposits and dirt on the electrodes or the sensor head.  
 

Содержание PIT Series

Страница 1: ...Page 1 of 14 Magnetic Inductive Flow Velocity Sensor PIT UMF2 B Operating Manual...

Страница 2: ...1 Ambient temperature ranges 7 5 3 2 Storage temperature 7 5 3 3 Climatic category 7 5 3 4 Ingress Protection 7 5 3 5 Shock resistance vibration resistance 7 5 4 Medium temperature and pressure 8 5 4...

Страница 3: ...notice Unless otherwise agreed on the rejected parts must be made available to us in case a claim is made 1 Identification 1 1 Manufacturer supplier Heinrichs Messtechnik GmbH Robert Perthel Str 9 D...

Страница 4: ...nd E2 at right angles to the direction of flow and the magnetic field B generated by the field coils 3 The size of this measuring circuit voltage is proportional to the mean flow velocity and thus the...

Страница 5: ...he electrode cable must be shielded and protected against disturbing interferences 3 2 3 Sensor versions PIT S Wetted parts stainless steel hastelloy PTFE Transmitter neck connection flange stainless...

Страница 6: ...ing to EN 29104 Measurement of Fluid Flow in Closed Conduits Methods of Evaluating the Performance of Magnetic Inductive Flow meters Based on this standard the minimum straight run of pipe ahead of th...

Страница 7: ...the PIT A sensor will be supplied an earthing If the sensor does not have an earthing electrode ensure that a suitable earthing disc will be used 5 2 Use in hazardous areas The flow meters PIT UMF2 B...

Страница 8: ...o 140 C 40 bar 6 Dimensions weight 6 1 PIT UMF2 B dimensional drawing for remote transmitter All dimensions in mm Weight 3 6 kg DN Nominal Diameter T Length of sensor d1 Diameter of sensor L Length of...

Страница 9: ...imensions in mm Weight 5 5 kg DN Nominal Diameter T Length of sensor d1 Diameter of sensor L Length of socket weld fitting ET Immersion depth in of pipe diameter Version DN T d1 L PIT A PFA 150 600 16...

Страница 10: ...flow measurements of fluids whose conductivity exceeds 20 S cm The manufacturer shall not be liable for damages that may result from unintended or inappropriate use When dealing with an aggressive med...

Страница 11: ...using this device it is important to ensure that the sensor head will not be damaged by closing the valve For a detailed description of this process see the Additional Operating Instructions PIT Druc...

Страница 12: ...m for remote installation The outer shield hast o be connected to the metalized cable glands at both ends The inner shields are connected to each other and are plugged into the terminal with the label...

Страница 13: ...ed medium Since this measured medium is dangerous in water poisonous corrosive flammable we have checked that all hollow spaces of the device are free of these materials neutralized and flushed all ho...

Страница 14: ...e 2004 108 EC Niederspannungsrichtlinie 2006 95 EG Low Voltage Directive 2006 95 EC EN 61000 6 2 2005 St rfestigkeit Industriebereich immunity industrial environment EN 61000 6 3 2007 A1 2011 St rauss...

Отзывы: