![HEINE D-008.73.103 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/heine/d-008-73-103/d-008-73-103_manual_2119191037.webp)
37/44
MED 234913 2021-02-03
MED 234913 2021-02-03
Immunity test levels
Test
Test level
IEC 60601-1-2 test levels
Compliance test levels
Electrostatic Discharge
(IEC 61000-4-2)
Contact Discharge: ±8 kV
Air Discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Radiated RF EM fields
(IEC 61000-4-3)
3 V/m
80 – 2700 MHz
80 % AM at 1 kHz
Electrical fast transients / bursts
(IEC 61000-4-4)
± 2 kV
100 kHz repetition frequency
Not applicable, because of
internally powered mode
Surges
(IEC 61000-4-5)
± 0.5 kV, ± 1 kV
Conducted disturbances
induced by RF fields
(IEC 61000-4-6)
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
(IEC 61000-4-3)
385 MHz; Pulse Modulation: 18 Hz; 27 V/m
450 MHz, FM: ±5 Hz deviation: 1 kHz sine; 28 V/m
710, 745, 780 MHz; Pulse Modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse Modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse Modulation: 217 Hz; 28 V/m
2450 MHz; Pulse Modulation: 217 Hz; 28 V/m;
5240, 5500, 5785 MHz; Pulse Modulation: 217 Hz; 9 V/m
Power frequency magnetic fields
(IEC 61000-4-8)
30 A/m; 60 Hz
Voltage dips
(IEC 61000-4-11)
0 % UT; 0.5 cycle; at 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 % UT;
25/30 cycles Single phase:
at 0°
Not applicable, because of
internally powered mode
Short interruptions
(IEC 61000-4-11)
0 % UT; 250/300 cycles