HBS 100 Varex
EN
N
◼
The neutral electrode is not detected or not detected correctly
Check all terminals. If the neutral electrodes is still not recognized, exchange it where
appropriate.
◼
It is not possible to switch the machine on or off
Check whether the power cable has been correctly connected and that the socket is intact. If
the error persists, inform the specialist personnel.
◼
The system works perfectly, but the settings cannot be saved.
Inform the specialist personnel.
7 Guideline value for power settings
Recommended
settings
Monopolar
Cutting
Smooth cuts with thin needle electrode
5 - 40 W
Cutting with knife electrode, lancet or wire loop
50 - 100 W
Monopolar
coagulation
Coagulation of large surfaces
40-100 W
Coagulation of small surfaces
20-50 W
Bipolar
application
Coagulation
30
– 60 W
Bipolar Varex
coagulation
Coagulation
0
– 25 W
When performing the power settings, the user should generally
take into account the cutting speed, geometry of the electrode
and properties of the tissue.
!
The recommended settings serve as guideline values only and must
be adjusted to the relevant situation.