Operation Manual 

 
 
 

CCT-3300 Series Conductivity 

Controller 

Содержание CCT-3300

Страница 1: ...Operation Manual CCT 3300 Series Conductivity Controller ...

Страница 2: ...any problems during installation and usage Check the actual product with complete set after you receive the package and contact us if any missing or damage Our serious promise 1 The meter s quality guarantee is one year from the date of purchasing During this period if the meter has quality problems manufacture is responsible for maintenance work for free or changes it 2 We provide lifelong mainte...

Страница 3: ... 93 standard G Consumable material is out of maintenance service and need purchase separately Please pay more attention when this symbol appears in this manual which refers to safe installation product functions and usage Without the influence on the operation any small changes or improvement on the products by the manufacture will not be notified separately Please make the object as the standard ...

Страница 4: ...ications 1 1 3 Technicial Features 1 Ⅱ Installation 2 2 1 Electrical connection 2 2 2 Installation of electrode 3 Ⅲ Settings 5 3 1 Mode settings 6 3 1 1 Electrode type setting 7 3 1 2 Electrode constant setting 7 Ⅴ Complete set 10 Ⅵ Order Directory 10 ...

Страница 5: ... high performances in high salt water treatment concentrated solution recirculating cooling water reclaimed water and ect 1 2 Classifications model Power supply frequency Hz Current loop modes control mode Constant selection cm 1 CCT 3300E DC24V Instrument Transmitter SPDT relay 0 01 10 0 CCT 3310E AC110V 50 60 Instrument Transmitter SPDT relay 0 01 10 0 CCT 3320 AC220V 50 60 Instrument Transmitte...

Страница 6: ...sistance 400Ω Max DC24V Control contact SPDT relay Load capacity AC230V 5A Max Note CCT 3320T has no relay Storage environment Temp 20 60 relative humidity 85 RH none condensation Working environment Temp 0 50 relative humidity 85 RH none condensation Dimension 48 mm 96 mm 80mm H W D Installation 44mm 92mm Panel mounted fast installation Ⅱ Installation Dimension Do not place the LCD display screen...

Страница 7: ... RED I I Instrument mode powered by instrument T T Transmitting mode powered by conditioning module RELAY Relay contact SPDT NO NC 24V A 24V B DC power supply automatic internal recognition polarity 0V 110V AC 110V input interface 0V 220V AC 220V input interface Electromagnetic compatibility ground protection terminal connect the earth NC Empty terminal no connection inside 2 2 Installation of ele...

Страница 8: ...t circuit they should not be connected to the same group of cable joint or terminal board with the power line or control line 4 The cable of electrode is special cable and has standard length before leaving factory Any customization for the cable length it s recommended to make an agreement with the factory before placing an order 5 Please keep the measuring part of electrode clean and do not dire...

Страница 9: ... 3300 series measurement mode normal display transient display setting mode for parameter setting switch diagram Select key 1 select thousand hundred ten and unit in circulate under parameter setting interface 2 switch to display conductivity TDS resistivity under measurement condition Add key 1 Adjust the value under parameter setting status 2 Check the temperature mA reading under measurement st...

Страница 10: ...ws Electrode selection C blinks on screen Setting mode Normal display timing Constant input C or 10 blinks Unit selection ppm or S cm blinks 4mA setting 4mA blinks 20mA setting 20mA blinks Hi limit alarm setting Hi blinks Lo limit alarm setting Lo blinks Enter and esc Decimal digital setting decimal digital blinks Long press For 3 seconds The setting value of 4mA and 20mA can not be equal The deci...

Страница 11: ...ode is special used for high pure water measurement For C 10 00 1 000 0 100 electrode they are used for pure water clear water and polluted water measurement The engineer unit could be modified 3 1 2 Electrode constant setting After constant selection the instrument would enter into the constant correction menu and this screen would display constant level Digit position blinks means the value coul...

Страница 12: ... is zero and stable the meter is good The problem can be initially identified from the electrode installation 1 Connect the white and green terminals on the short circuit meter and check the resistivity reading if the reading is 0 0MΩ cm the meter is good The problem can be initially identified from the electrode installation 2 Meter mode or transmitter mode of 4 20 mA judgment Before the judgment...

Страница 13: ...he pipeline or select anther measurement point C Use stable water quality to check the problem reason Serious error of reading A Incorrect constant Setting B The electrode constant is changed C Inappropriate flow speed of measured point D Incorrect electrode installation A Reset the electrode constant B Replace a new electrode or reset the electrode constant C Choose a new installation position D ...

Страница 14: ...eck the below items to choose the suitable power supply cell constant material fixed method for your applications Model power supply cell constant connection electrode material CCT 3300E DC 24V CCT 3310E AC 110V CCT 3320E AC 220V CCT 3320 AC 220V CCT 3320T AC 220V 0 01 cm 1 0 10 cm 1 1 00 cm 1 10 00 cm 1 thread ferrule 316L titanium 316L 316L plastic graphite NOTE The standard cable length is 5 me...