20
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
The 04-630-001, 04-630-006, 04-631-001 and 04-631-006 are intended for use in
the electromagnetic environment specified below. The customer or the user of the
04-630-001 and 04-630-006 should assure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromagnetic
environment
guidance.
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.
Содержание 04-630-001
Страница 26: ...SAMPLE BLOOD PRESSURE LOG...
Страница 52: ...52 EJEMPLO DE REGISTRO DE PRESI N ARTERIAL...
Страница 53: ...53 91 630 631 05 15 2015 HealthSmart International...
Страница 54: ...54...