SPECIFICATIONS
22
23
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
About 22cm~42cm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Temperature:-20
0
C-60
0
C
Relative Humidity: 10%RH-93%RH
Atmospheric Pressure: 50kPa-106 kPa
Temperature:5
0
0
C to 40 C
Relative humidity:
85%RH
≤
Atmospheric pressure: 86kPa to 106kPa
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Net Weight
Approx. 225g (excluding the dry cells and cuff)
External dimensions
In the box
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
IP21
Accuracy
Normal working condition
Storage & transportation
condition
Software Version
V01
Pressure:
5
0
0
C-40 C within±0.4kPa(3mmHg)
pulse value:±5%
Rated cuff pressure:
0kPa - 40kPa (0mmHg~300mmHg)
Measurement pressure:
(0~280mmHg)
0-37.30kPa
pulse value: (40-199) beat/minute
Digital LCD V.A. 40 x 60 mm
Approx 118 (W) x 110 (L) x 52 (H) mm
Classic BP-120, 4xAAA batteries, BP Cuff,
User manual & Warranty card
Battery powered mode:
6VDC 4×AAA batteries
AC adaptor powered mode: 6V 1A
(Please only use the recommended AC adaptor
model). (excluded)
TROUBLESHOOTING
PROBLEM
REMEDY
No Power
Check the battery
power. Check the
battery polarity position
Replace with new sets
of batteries. Proper
installation of batteries
polarities
No inflation
Check the air hose plug
Tightly insert plug in
air socket
CHECK THIS
Err and stop working
Movement of arm or
body while measurement
Talking while measurement
Keep the body peaceful
Keep quiet when
measuring
Cuff leak
Whether the cuff wrap is
too loose. Whether the cuff
is broken
Wrap the cuff tightly
Change to new cuff
Please contact HealthSense Customer care if you can't solve the problem,
do not disassemble the unit or perform unauthorized repair by yourself
SYMBOL
CAUSE
CORRECTION
No display appears
Weak battery
orimproper placement
Replace both batteries
with new ones. Check
the battery installation
for proper placement
of the battery polarities.
Er 2
Monitor could not
detect pulse wave
Make sure the air plug
is properly inserted in
the unit
Er 4
It appears during the
process of inflating
Wrap the cuff properly
Er 5
Pipe blocked
Make sure the air flow
is smooth
Er 6
Heart rate area display
“Hi”
Heart rate is more
than 200 beats/Min.
Er 7
Heart rate area display
“Lo”
Heart rate is less than
40 beats/Min.
Er 1
Sensor abnormal
Send it to the local
distributor
Er 3
Measurement result is
abnormal (SYS<45mmHg
or DIA<24mmHg)
Occasionally - measure
for one more time/Always -
Send it to local distributor