16
CALIBRATION
This scale has been factory calibrated, and does not require calibration prior to use.
Calibration of your 3001KL-AM/3001KL-AMX scale can be performed using calibrated
kilogram (kg) or pound (lb) weight standards. The required calibration standard will be
the Unit of Measure (UOM) for which the scale is configured.
Because the scale is
shipped from the factory in “KG” mode, the requested UOM for
calibration will be the KG standard (e.g., 400kg). If the unit is to be calibrated with LB
weight standards, the scale must first be configured to the
“LB” UOM. Refer to the
Unit
of Measure Configuration
section on page 12 for configuration details.
Note:
Only weights certified and traceable to national standards are to be used for
calibration.
Warning: Always ensure that the scale is returned to the correct UOM after scale
calibration.
Operator Action
Display
1. Ensure the display is powered off, then press and hold the
HOLD/RELEASE
and
ZERO
button while pressing the
ON/OFF
button.
(Two alphanumeric characters will appear next to “U”
followed by 4-5 characters.)
U”XX” followed by a 4-5 digit
‘idle scale count’ number.
2. Press the
ENTER
button.
“CAL”.
3. Make sure there is nothing on the scale, then press the
ENTER
button.
“Clear” followed by "400.0 KG"
(in KG mode) or
“0400.0 LB”
(in LB mode)
4. Use the ► and ▲ buttons to set the calibration weight to a
value other than the default weight. The calibration weight
should be as high as possible, with 270 KG/400 LB as a
minimum calibration weight.
Shows "400.0 KG" (in KG
mode) or
“0400.0 LB” (in LB
mode) or entered calibration
weight.
5. Press the
ENTER
button.
Displays
“Load” followed by
"400.0 KG" (in KG mode) or
“0400.0 LB” (in LB mode)
6. Load the scale with the required weight.
Display increments showing
raw weight data.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the
ENTER
button.
“SUCC” or “Err1”.
8. If the calibration is successful, the display will show SUCC
and power off. If the calibration is not successful, the display
will show Err1 and power off. Contact customer service for an
“Err1” display (+1 800-638-3722).
-------