Health O Meter 3001KL-AM Скачать руководство пользователя страница 15

15 

MAINTENANCE

 

 

Maintenance 

The following pages provide instructions for maintenance, cleaning, calibration, and 
troubleshooting for the 3001KL-AM/3001KL-AMX scales. Maintenance operations other 
than those described in this manual should be performed by qualified service personnel. 

Caution: Before first use, or after long periods of non-use, check the scale for 
proper operation and function. If the scale does not operate correctly, refer to 
qualified service personnel. 

 
1.  Check the overall appearance of the scale for obvious damage, wear, and tear. 
2.  If applicable, inspect the AC adapter for cord cracking, fraying, and for broken/bent 

prongs. 

 
 
Cleaning 

Proper care and cleaning is essential to ensure a long life of accurate and effective 
weighing. 

Caution: Disconnect the scale from the AC adapter power source prior to 
cleaning the unit.   
Note:

 Routine cleaning does not affect the antimicrobial properties of the handrail 

coating.   
 
1.  Clean all external surfaces with a clean damp cloth or tissue. Dry with a clean soft 

cloth. 

2.  Do not immerse the scale into cleaning or other liquid solutions. 
3.  Do not use Isopropyl Alcohol or other solutions to clean the display surface.  
4.  Do not use abrasive cleaners. 
 

 
Replacement Parts 
 

Part # 

Description 

B2266701-0 

PLASTIC FOOT  ASSEMBLY 

DISPLAY1100 

DISPLAY HEAD 

LKIT1100 

ELECTRONICS KIT 

 
 
 

Содержание 3001KL-AM

Страница 1: ...1 Model 3001KL AM 3001KL AMX Antimicrobial Digital Platform Scale User Instructions P N UM3001KL AM 3001KL AMX Rev20170921_EngOnly...

Страница 2: ...ecifications 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Set up Instructions 10 Operating Instructions 11 Maintenance 15 Calibration 16 Troubleshooting 17 Warranty 18 Note This sc...

Страница 3: ...ation Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source Unplug the AC adapter and carefully store both the adapter and adapter cord before moving the scal...

Страница 4: ...factory locked in KG mode Refer to Page 12 to learn more about configuring the scale Antimicrobial Powder Coating Unique to the 3001KL AM 3001KL AMX scales the handrails and transport handle have bee...

Страница 5: ...lable during normal business hours at 1 800 638 3722 Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or local regula...

Страница 6: ...ly remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any...

Страница 7: ...e completely hand tightened 2 Turn the platform over so it is facing up Locate the handrail and untie the load cell cable connector Position the handrail over the handrail supports on the platform and...

Страница 8: ...forward to access the underside of the platform Plug the load cell connector cable into the port 6 Carefully tilt the scale back to an upright position Obtain the display head and align the four hole...

Страница 9: ...ell port on the rear of the display head 8 Attach the adapter holder to the handrail using the 2 included sheet metal screws Use a Phillips head screwdriver to tighten the screws The scale is now asse...

Страница 10: ...ified AC adapter Only use the Health o meter Professional AC adapter model ADPT30 designed for this scale 2 For battery operation see Battery Installation below Battery Installation 1 If connected unp...

Страница 11: ...culate patient s Body Mass Index BMI RECALL Recalls the previous weight TARE PRETARE TARE Deducts the weight of an object on the platform PRETARE Prompts entry of a weight value that will be deducted...

Страница 12: ...turn on the scale 3 Wait until the 0 0 LB or 0 0 KG appears on the display 4 Ask the patient to step on the scale The scale will begin to display the patient s weight and will lock on the patient s w...

Страница 13: ...manually by using the keypad or automatically as outlined in this section Tare 1 With all weight off the scale press the ON OFF button to turn on the scale 2 Wait until the 0 0 LB or 0 0 KG appears o...

Страница 14: ...s weight 4 Press the TARE button The scale will zero out and the word TARE will appear at the bottom center of the display indicating that there is a tare value for the parent s weight 5 Ask the paren...

Страница 15: ...ble inspect the AC adapter for cord cracking fraying and for broken bent prongs Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconne...

Страница 16: ...and ZERO button while pressing the ON OFF button Two alphanumeric characters will appear next to U followed by 4 5 characters U XX followed by a 4 5 digit idle scale count number 2 Press the ENTER bu...

Страница 17: ...y balanced 3 Place the scale on a level floor Tighten the feet on the underside of the platform to ensure all 4 feet are properly balanced 4 Scale is out of calibration 4 Check weight with known weigh...

Страница 18: ...contrary to the operating instructions abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire flood hurri...