8
ASSEMBLY INSTRUCTIONS
4. Tip the platform and handrails back to
an upright position. Fasten the handrail
to the platform using the (8) included M8
1.25 x 20 bolts (4 bolts per each
handrail support). Use the included
13mm box wrench to tighten the bolts.
5. Carefully tilt the scale forward to access
the underside of the platform. Plug the
load cell connector cable into the port.
6. Carefully tilt the scale back to an upright
position. Obtain the display head and
align the four holes on the back of the
display head with the four mounting
holes on the display support. Using the
(4) included M4 x 0.7 x 10mm screws,
fasten the display head onto the display
support using a Phillips head screwdriver
to tighten the screws.
7. Plug the load cell connector cable into
the load cell port on the rear of the
display head.
The scale is now assembled. See
Powering the Scale on page 9 for
battery/AC adapter instructions. If
powering the scale via AC adapter, see
the next step for attaching the adapter
holder.
8.
Optional Step
– For powering the
scale via AC power adapter.
Attach the
adapter holder to the handrail using the
(2) included sheet metal screws. Use a
Phillips head screwdriver to tighten the
screws.
Load cell connector cable port