CAUTIONS AND WARNINGS 

INTENDED USE 

This Health o meter

®

 Professional platform scale is intended to be used in a professional medical environment 

by trained medical staff. This product was designed to weigh patients who are safely positioned and standing 
on the platform. Do not modify the product or use it for anything other than its intended purpose. 

 

To prevent patient/caregiver injury or damage to your scale, please follow the instructions in this 
user manual very carefully. 

 

 

Do not exceed the weight capacity specified for this scale. 

 

For accurate weighing, this scale must be placed on a flat, stable surface. 

 

For accurate weighing, verify before each use the proper operation according to the procedure 
described in this manual. 

 

Do not use in the presence of flammable or explosive materials. 

 

If the scale becomes damaged, it should not be operated until properly serviced.  

 

Do not wheel or pull the scale down stairs, doing so may damage internal parts. 

 

Ensure that the AC adapter and scale do not come in contact with liquids, excessive temperature, 
or excessive humidity. 

 

Patient/caregiver safety 

 

This scale is designed for static weighing of patients only. No scale should be used for patient 
transfer.  

 

To prevent patient injury, the patient must be attended throughout the entire weighing event.  
Caregivers should ensure that the patient is stable and provide support as needed when getting 
onto and off of the scale. 

 

Patient should wear socks or other form of lightweight foot covering during weighing event. 

 

 When operating the scale with the AC adapter: 

 

Operate this scale exclusively with the AC adapter that is provided (ADPT30). Use of an 
unspecified adapter will void the warranty, and can pose a serious safety hazard.   

 

Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent 
plug prongs. 

 

Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated 
voltage appropriate for operation.   

 

Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.  

 

Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the 
scale.   

 

Operating this scale at voltages and frequencies other than specified can damage the equipment 
and will void the warranty. 

 

When operating the scale with batteries: 

 

If the scale’s battery icon indicates that the batteries are low replace the batteries or switch to the 
AC adapter as soon as possible.  

 

When replacing / inserting batteries, be sure to use all new batteries. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 
93/42/EEC (EN 55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable 
protection against harmful interference in a typical medical installation. The equipment generates, uses 
and may radiate radio frequency energy and, if not installed and used in accordance with these 
instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that 
interference will not occur in a particular installation. If this equipment does cause harmful interference with 
other devices, which can be determined by turning the equipment off and on, the user is encouraged to try 
to eliminate the interference by one or more of the following measures: 

 

Reorient or relocate the receiving device. 

 

Increase the separation between the equipment. 

 

Connect the equipment into an outlet on a circuit different from that to which the other device(s) is 
connected. 

 

Consult Customer Service or a field service technician for help. 

 

 

In no event whatsoever shall Pelstar, LLC be liable for damages or injuries whatsoever arising out 

of or connected with the assembly, use, or misuse of its products.

 

Содержание 3001KG-AM

Страница 1: ...MX Antimicrobial Digital Platform Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM3001KL AM_3001KL AMX_3001KG AM_3001KG AMX_E_R...

Страница 2: ...al 5 Assembly Instructions 6 Set up Instructions 9 Everlock Instructions 13 Operating Instructions 14 Maintenance and Cleaning 20 Calibration 21 Troubleshooting 23 Warranty 24 Note This scale has been...

Страница 3: ...cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ensure that th...

Страница 4: ...ale Specifications Capacity and Resolution 3001KL AM 3001KL AMX 1000 lb x 0 2 lb 454 kg x 0 1 kg 3001KG AM 3001KG AMX 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15W...

Страница 5: ...se contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal This Health o meter Professional scale must be disposed of prope...

Страница 6: ...king materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts...

Страница 7: ...e completely hand tightened 2 Turn the platform over so it is facing up Locate the handrail and untie the load cell cable connector Position the handrail over the handrail supports on the platform and...

Страница 8: ...d with the four mounting holes on the display support Using the 4 included M4 x 0 7 x 10mm screws fasten the display head onto the display support using a Phillips head screwdriver to tighten the scre...

Страница 9: ...ead by unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professi...

Страница 10: ...the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button to briefly disp...

Страница 11: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Страница 12: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Страница 13: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Страница 14: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Страница 15: ...by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 4 To rewei...

Страница 16: ...may take several seconds for the scale to lock onto the weight 3 After obtaining the patient s weight and with the patient remaining on the scale press 4 The scale will prompt entry of the patient s...

Страница 17: ...he tared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as...

Страница 18: ...nt The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain...

Страница 19: ...parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Страница 20: ...and Disinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit...

Страница 21: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Страница 22: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Страница 23: ...n the feet on the underside of the platform to ensure all 4 feet are properly balanced 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing proce...

Страница 24: ...ty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make...

Отзывы: