2.B.2
Cables:
A high grade shipboard approved instrument cable with woven shield is recommended to
reduce noise interference. Other types of cable are available at additional cost.
C
System Capacities:
2.C.1
Transducer:
Transducers are rated at full range. Units 10K and under are capable of withstanding a 30
percent over pressurization without damage to the system. 16K units are capable of
withstanding a 10 percent over pressurization without damage to the system. System
capacities are as follows:
6,000 PSI
10,000 PSI
16,000 PSI
20,000 PSI* (Developmental)
SECTION III
Installation
:
3.A
Components:
3.A.1
Transducer:
The transducer is to be mounted by the customer at the locations deemed most suitable for
such equipment.
3.A.2
Cables:
Maximum recommended cable length is 200'. Cable lengths, other than standard, shall have
been specified by the customer to the factory in advance. Other cable types are available upon
request. Cabling will be installed using acceptable industry standards for such installations.
3.B Mounting Components:
3.B.1
Pressure Transducer:
The pressure system transducer comes from the factory as a complete assembly and only
requires mounting of the transducer housing using standard practices for such installations.
Note
: Special care should be taken to insure that the transducer pill mounted in the center of
the transducer housing, is not allowed to make contact with any hard and/or sharp object as
the accuracy and the sensitivity could be affected. The transducer pill is vulnerable to this
type of contact. Any evidence of mishandling or abuse shall
void
the warranty. We have