ACCESSORIES

PT-VDET PROOF TESTER

®

FOR

V-DETECT AC VOLTAGE DETECTOR

The PT-VDET Proof Tester Voltage Detector Tester is for use on the V-DETECT AC Voltage
Detector. This tester generates high voltage AC for testing the V-DETECT AC Voltage
Detector. To use, turn on the V-DETECT Detector and set it for the lowest range, hold
the Tester in one hand and the V-DETECT  Detector in the other. Apply
the top of the V-DETECT Detector to the metal end plate
on the Tester. Press and hold the TEST button on the Tester.
A properly operating V-DETECT AC Voltage Detector
will indicate voltage with continuous flashing lights and
beeping sound. For best results, hold the PT-VDET
with a tight grip and wrap your fingers around it.
If the V-DETECT AC Voltage Detector does
not indicate voltage, do not use it and return it
to HD Electric Company for service.

CAUTION:

This Tester generates high voltage

AC for testing V-DETECT AC Voltage Detectors.
There is no danger of electric shock when this
tester is used as directed. Discontinue use and
return to HDE for service if the housing is cracked
or broken, or if the battery cover is lost.

CAUTION: High Voltage

WARNING:

Do not use this tester except as directed.

Do not use to test equipment other than specified
V-DETECT AC Voltage Detectors. Do not apply to energized circuits or equipment.
Do not operate this Tester without the battery cover and do not open the housing. Refer
all servicing to the factory. Failure to follow these instructions may lead to electric shock,
severe injury or death.

MAINTENANCE AND CARE

STORAGE

- It is recommended for protection that the instrument and its accessories

are stored in the carrying bag or case.

CLEANING INSTRUCTIONS

- To clean, wipe with a damp cloth with water. 

Do not use harsh chemicals or solvents.

CLEANLINESS

- The molded housing is very rugged, but it should be kept clean and free

of dirt, grease and any other foreign materials. If the housing surface integrity has been 
compromised in any way, remove from service and return to factory for repair or replacement.

DAMAGE

- If you suspect any mechanical or electrical damage, do not use the 

instrument and arrange for repair by returning to the factory.

CALIBRATION & TESTING

- Regular calibration is not required.

There is no calibration adjustment.

SERVICE

- Return to HD Electric Company for service.

MANUFACTURING LOCATION - 

HD Electric Company • Waukegan, IL. 60085, USA

Содержание V-DETECT

Страница 1: ...ng Instruction Manual Making the Invisible Visible V DETECT and ACCESSORIES AC VOLTAGE DETECTOR 1475 Lakeside Drive Waukegan Illinois 60085 U S A 847 473 4980 fax 847 473 4981 website www HDElectricCompany com ...

Страница 2: ...2 ...

Страница 3: ...T AC Voltage Detectors have user selectable voltage ranges covering the range from secondary voltages to the highest transmission voltages Accessories may include a holster charging cord for charging the unit from a car or truck a 115 240VAC wall charger adapter a carrying bag and a carrying case OPERATIONAL IMPAIRMENT If the V DETECT AC Voltage Detector is used in a manner not described in this i...

Страница 4: ...r working order by pressing the ON button and checking that all voltage range lights turn on in turn and that the unit stays on following the self test Never allow another high voltage or grounded conductor to contact the instrument during use Keep the housing free and clear of all structures at all times Bridging the housing from line to ground or line to line may cause a fault and arc The V DETE...

Страница 5: ...est it has passed the self test and is ready to use If it turns off after this test then it has not passed the self test and must be recharged or serviced While the instrument is on it will emit a very short beep and flash of the lights once every FOUR SECONDS 3 SELECT A VOLTAGE RANGE Use the up and down arrows to select a voltage range The lowest range setting is indicated by the lowest voltage r...

Страница 6: ...ested as sensing distance will be reduced or diminished NEVER allow prolonged arcing between an energized conductor and the instrument This will damage the housing of the unit NEVER bridge between conductors with the housing This may result in an arc and or fault NEVER select a voltage range above the voltage of the conductor to be tested The instrument may not detect voltage on the conductor bein...

Страница 7: ...ours When the charge is complete only the highest voltage range light will remain on After three hours this light will turn off The battery cannot be overcharged A quick fifteen minute charge to a low battery will allow one day of use under typical conditions A full charge will allow more than one week of continuous use under typical conditions If the instrument will not turn on try charging it fo...

Страница 8: ...DIMENSIONS 11 L x 3 W x 2 D 28 cm x 7 6 cm x 5 cm WEIGHT 11oz 0 31 kg OPERATING FREQUENCY 25Hz 400Hz BEEPER SOUND PRESSURE LEVEL 100 db BATTERY LIFE One week of typical daily use after full charge BATTERY CHARGING CURRENT 0 5A maximum at 12VDC ENCLOSURE MATERIAL Supertough Nylon UL 94 HB PRINTED CIRCUIT BOARDS FR 4 UL94V 0 ENVIRONMENTAL CONDITIONS CONDITIONS Indoor and outdoor use ALTITUDE Up to 6...

Страница 9: ...r if the battery cover is lost CAUTION High Voltage WARNING Do not use this tester except as directed Do not use to test equipment other than specified V DETECT AC Voltage Detectors Do not apply to energized circuits or equipment Do not operate this Tester without the battery cover and do not open the housing Refer all servicing to the factory Failure to follow these instructions may lead to elect...

Страница 10: ...CARRYING HOLSTER Carrying holster with belt clip and restraining strap D CS DET CARRYING CASE Carrying case holds V DETECT Detector PRX HS PT VDET Tester and charge cords E VDET BAG CARRYING BAG Padded carrying bag with strap holds V DETECT Detector PRX HS PT VDET Tester and charge cords F HOT STICKS A range of hot sticks are available Contact HD Electric for more details A F D C E B ...

Страница 11: ...11 ...

Страница 12: ... beyond HDE s reasonable control Deliveries or other performance may be suspended for an appropriate period or cancelled by HDE upon notice to Buyer in the event of any of the foregoing but the balance of this agreement shall otherwise remain unaffected If HDE determines that its ability to supply the total demand for the Goods or to obtain material used directly or indirectly in the manufacture o...

Отзывы: