23
Z7A/...
HBM
A1847-1.1 de/en/fr
4
Specifications
T
ype
Z7A
Accuracy class according to OIML R 60
Maximum number of load
cell verification intervals (n
LC
)
D1
1000
C3
3000
Max. capacity (E
max
)
t
0.5; 1; 2; 5; 10
2; 5; 10
Min. load cell verificat. interval (v
min
)
% of
E
max
0.0357
0.0100
Sensivity (C
n
)
mV/V
2
Sensivity tolerance
%
0.1000
0.0500
Temperat. effect on sensivity (TK
C
)
1)
% of
0.0350
0.0117
Temperat. effect on zero signal (TK
0
)
% of
C
n
/10 K
0.0500
0.0140
Hysteresis error (d
hy
)
1)
0.0500
0.0170
Non-linearity (d
lin
)
1)
% of C
n
0.1000
0.0333
Creep (d
cr
) in 30 min.
n
0.0735
0.0167
Input resistance (R
LC
)
Ω
350
Output resistance (R
O
)
Ω
356
0.2
356
0.12
Reference excitation volt. range (U
ref
)
V
0.5...12
Max. excitation voltage (B
U
)
V
18
Nominal temperature range (B
T
)
−
10...+40 [+14...104]
Operating temperature range (B
tu
)
°
C [
°
F]
−
30...+70 [
−
22...158]
Storage temperature range (B
tl
)
[ ]
−
50...+85 [
−
58...185]
Safe load limit (E
L
)
% of
E
150
Breaking load (E
d
)
E
max
300
Max. capacity (E
max
)
t
0.5
1
2
5
10
Relative static lateral force limit (E
lq
)
2)
% of
100
50
25
(100)
2)
15
(100)
2)
18
(100)
2)
Permissible dynamic load (F
srel
)
(vibra-
tion amplitude acc. to DIN 50100)
% of
E
max
70
Nominal displacement
(
s
nom),
approx.
mm
0.25
0.30
0.35
0.45
0.70
Weight (G), approx.
kg
2.3
5
8
Degree of protection, acc. to IEC529
IP 67
Material
Measuring body
Cable gland
Cable sheath
Steel, galvanized nickel
Stainless steel, Viton
R
PVC
1)
The data for linearity error (d
lin
) hyseresis (d
hy
) and temperature effect TK
c
) on sensitivity are typical val-
ues. The sum of these data meets the requirements of OIML R60
2)
The values indicated in parentheses refer to mounting conditions, with which the transducer foot is pre-
vented by stops from a movement. Here, major effects of errors have to be expected.