19
U93
A2081-2.2 en/de/fr
HBM
Parameters
Explanation
Rights
level
Unit
Value
1)
Impedance of each
bridge element
345.0
ohm
ID
Input resistance according to the
HBM data sheet
Response time
1.0000000
u
sec
ID
Not relevant to HBM transducers.
Excitation level
(nominal)
5.0
V
ID
Nominal (rated) excitation voltage
according to the HBM data sheet
Excitation level
(minimum)
0.5
V
ID
Lower limit for the operating range of
the excitation voltage according to
the HBM data sheet.
Excitation level
(maximum)
12.0
V
ID
Upper limit for the operating range of
the excitation voltage according to
the HBM data sheet.
Calibration date
1‐Jun‐2006
CAL
Date the manufacturing certificate
was created at HBM (or date of last
calibration or entry of TEDS data).
Format: day‐month‐year.
Calibration initials
HBM
CAL
Initials of the calibrator or calibration
laboratory concerned (max. three
characters).
Calibration period
(days)
730
days
CAL
Time before recalibration, calculated
from the date specified under
Calibration Date.
Measurement location
ID
0
USR
Identification number for the
measuring point (assigned by the
user). Possible values: 0 to 2047
(numbers only). The HBM Channel
Comment template is also available
as a supplement.
1)
Typical values for an HBM U93/2kN type force transducer, manufactured on 1.6.2006
The user can also write other templates, such as the
Signal Conditioning
template.