Connections and displays

20

A4722-1.0

  

HBM: public

CX22B-R

Important

The data recorder hosts a small buffer battery. This
battery keeps date and time during power interruptions.
The battery shall only be replaced by HBM service.

4.2

Ground connection and grounding

Lay the signal and data leads separately from current-
carrying power lines. Cable ducts made of sheet metal
with an internal partition are ideal.

If there are differences in potential in or to the connected
measurement system, you must install a potential equal­
ization line (recommended value: highly flexible stranded
wire, line cross-section 10mm

2

).

4.3

Connection of MX modules

The simplest way to connect measurement modules to
the CX22B‒R Data Recorder is via FireWire, or alterna­
tively via Ethernet.

FireWire has the following advantages:

S

Voltage supply, time synchronization, data communi­
cation and real time in one cable and thus little effort
and low cost

S

Every module has 2 FireWire connections, so flexible
topologies are possible: star, line or even hybrids with
an appropriate hub, also supplied with power via
FireWire.

Ethernet has the following advantages

Содержание SomatXR CX22B-R

Страница 1: ...Operating Manual English CX22B R catman Data Recorder...

Страница 2: ...0 Fax 49 6151 803 9100 info hbm com www hbm com Mat 7 2002 4722 DVS A4722 1 0 HBM public 12 2017 E Hottinger Baldwin Messtechnik GmbH Subject to modifications All product descriptions are for general...

Страница 3: ...21 4 5 USB 2 0 interfaces 22 4 6 Digital inputs and outputs 22 4 7 LED displays 23 5 Connecting measurement modules 26 6 Operating the Data Recorder 29 6 1 Operation with monitor and keyboard mouse 29...

Страница 4: ...performance 69 10 3 EWF safeguarding your settings against change 69 11 Operating the CX22B R as gateway 71 11 1 Description 71 11 2 Activate deactivate gateway mode 71 11 3 Connect to a QuantumX modu...

Страница 5: ...unting and careful opera tion Everyone involved with siting starting up or operating the module must have read and understood the operating manual and in particular the technical safety instructions I...

Страница 6: ...mounted on a surface or in a box Please consider following conditions at the place of installation S Observe the maximum permissible ambient tempera tures given in the specifications S Minimize devic...

Страница 7: ...when open is not allowed Once the device has been opened it must be professionally inspected in accor dance with EN 60950 before starting it up again The module is delivered from the factory with a fi...

Страница 8: ...h water pressure to the unit for clean ing Give the module sufficient time to dry before starting it up again General dangers of failing to follow the safety instructions The module is a state of the...

Страница 9: ...s are in operation the following measures need to be taken to reduce the risk of incineration at sur faces Operating personnel must be advised of the possi bility of incinerations when touching the su...

Страница 10: ...c tionality HBM recommendation Greenline shielding design downloadable from the Internet at http www hbm com greenline The measurement module and devices used in automa tion equipment must be covered...

Страница 11: ...personnel means persons entrusted with siting mounting starting up and operating the product who possess the appropriate qualifications for their function This includes people who meet at least one of...

Страница 12: ...tandards It is also essential to comply with the legal and safety requirements for the application concerned during use The same also applies to the use of accessories Maintenance and repair work on a...

Страница 13: ...ur attention to important infor mation about the product or about handling the prod uct Tip This marking indicates application tips or other infor mation that is useful to you Emphasize See Italics ar...

Страница 14: ...les the manufacturer to guarantee that the product complies with the requirements of the rele vant EC directives the Declaration of Conformity can be found on the HBM support website www hbm com Statu...

Страница 15: ...istered trademark of Hottinger Baldwin Messtechnik GmbH All trademarks and brands used in this document are belonging to the respective product or the manufacturer owner Hottinger Baldwin Messtechnik...

Страница 16: ...odule S What can be connected and what to look out for S How to start your first measurement and DAQ jobs and how to get to the measurement data The following documentation is also available S SomatXR...

Страница 17: ...cket 4 pin for connecting SomatXR or QuantumX modules S Voltage supply ODU socket 4 pin S WLAN for a sma antenna The following interfaces and options can be found on the front S 2 x Gigabit Ethernet 2...

Страница 18: ...B R 4 Connections and displays 2x Gigabit Ethernet Status LEDs 1 2 2x USB 2 0 Monitor DVI D Protection caps Fig 4 1 CX22B R connections on the front Supply voltage DC 10 V 30 V WLAN antenna 2x Fire Wi...

Страница 19: ...ruptible power supply UPS to buffer short voltage drops It also allows a regu lar storing of the measurement data respectively a proper shut down of the module When running on the UPS no power will be...

Страница 20: ...ted measurement system you must install a potential equal ization line recommended value highly flexible stranded wire line cross section 10 mm2 4 3 Connection of MX modules The simplest way to connec...

Страница 21: ...ules via WLAN 4 4 Communication with the Data Recorder You can work with the Data Recorder in the following way S Directly connected peripherals Monitor or touchscreen DVI Mouse and keyboard to USB S...

Страница 22: ...4 6 Digital inputs and outputs Three inputs and three outputs are available at connector X6 ODU 14 pin The inputs and outputs are TTL com patible 0 5 to 5 5V Inputs The level for the inputs is 5 volts...

Страница 23: ...eset to LOW if in catman EASY in the options for the CX22B R system group you activate the setting Reset all CX22B R digital outputs at the start of a DAQ job 4 7 LED displays Fig 4 3 Position of the...

Страница 24: ...Connections and displays 24 A4722 1 0 HBM public CX22B R Fig 4 4 Pin assignment digital In and Output...

Страница 25: ...being recorded If a start trigger was set the LED flashes yellow while waiting for the trigger event The ERROR LED lights up red if the software detects an error e g when no connection could be set up...

Страница 26: ...nection X102 of the CX22B R In the configuration shown below the module supply volt age is looped through via FireWire max 1 5 A via FireWire The power consumption of a module is ex plained in the da...

Страница 27: ...ASY checks the firmware automatically and if necessary prompts you to update the firmware The relevant firmware package is supplied with the Data Recorder and is located in the sub directory C Program...

Страница 28: ...Ethernet If you have more than two mod ules you will also require an Ethernet switch For syn chronization an additional FireWire connection or the use of a PTPv2 capable Ethernet switch e g EX23 R is...

Страница 29: ...ll connections Connection via VGA is not possible not even with a DVI VGA adapter plug as the leads required for VGA are not as signed Switch on the power supply and wait until the boot process is com...

Страница 30: ...can be updated free of charge for 12 months from the date of delivery To run an update you must first register with HBM in order to benefit from software maintenance To do this simply send us the reg...

Страница 31: ...via DHCP Dynamic Host Con figuration Protocol For a detailed description see Section 6 2 2 Page 36 S The CX22B RW Data Recorder is connected via WLAN ad hoc to a PC or hotspot integration into lo cal...

Страница 32: ...at least or work with the direct pe ripherals 6 2 1 A direct Ethernet line between PC and Data Recorder Connect your PC and CX22B R via an Ethernet ca ble category CAT5e or higher Install the HBM Devi...

Страница 33: ...can configure the CX22B R The sim plest setting is DHCP APIPA Suitable IP addresses are automatically organized between the two devices and you can start work immediately Alternatively you can use fix...

Страница 34: ...xample for IP address and subnet mask IP address of the PC 192 168 0 1 subnet mask 255 255 255 0 Possible combination IP address of the PC 192 168 0 2 sub net 255 255 255 0 Click on Apply and wait unt...

Страница 35: ...indows dialog setting up the connection Depending on the configuration of your Windows operating system another query whether the connec tion should be set up may be displayed identity of the remote c...

Страница 36: ...a Recorder 36 A4722 1 0 HBM public CX22B R Fig 6 3 Identity of the remote computer display The connection to the CX22B R Data Recorder will now be implemented and you must register with the device Win...

Страница 37: ...Password hbm Important If you change the IP address and subnet mask this is not yet permanently saved So in the start screen click on Commit changes and reboot or Save and restart if you have yet to c...

Страница 38: ...ncluding the Data Recorders will appear in the list The IP addresses and subnet masks of the PC inter faces are displayed in the Settings of my computer section on the left In the TCP IP settings for...

Страница 39: ...e the CX22B R into an existing wire less network This requires a configured Access Point and information regarding the network name login and password Setting up the CX22B R for Adhoc operation Make t...

Страница 40: ...the CX22B R This is done convention ally with a monitor and keyboard or remotely via the LAN cable Exit to Windows and click on the small pictogram at the bottom right Connect to the network and if n...

Страница 41: ...tion that a suitable program is installed on the particular de vice The screen resolution should be 1000 x 750 pixels as otherwise only parts of the CX22B R Data Recorder screen will be visible and yo...

Страница 42: ...s operating system In Win dows 7 the program can be found under All programs Accessories Enter the IP address of the CX22B R here Fig 6 5 Enter the IP address of the SomatXR CX22B R Procedure Enter th...

Страница 43: ...ogram To do this open the CX22B rdp file in the installation directory of the HBM Device Manager software Important The group policy of the network administrator can limit or totally prevent changes b...

Страница 44: ...4722 1 0 HBM public CX22B R Activate data exchange via the clipboard Fig 6 6 Remote desktop connection local resources tab Click on Next in the Local resources tab Enable the required drives of your P...

Страница 45: ...Enable drives for use in the CX22B R Fig 6 7 Dialog for enabling additional resources Close the dialog Save the configuration via the General tab either under a new name Save as or as a default settin...

Страница 46: ...and password to be saved Save settings Fig 6 8 Saving the remote desktop configuration If you double click this configuration file or a link to it a connection to the CX22B R is immediately set up and...

Страница 47: ...C and the CX22B R S Have you waited long enough for the PC to specify its address If the PC cannot find a server in the network the search for the server looks in the DHCP settings Obtain an IP addres...

Страница 48: ...try deactivating all the other Ethernet interfaces S If the HBM Device Manager cannot find the CX22B R try setting up a standard remote desktop connection S If you are using the CX22B R in a larger n...

Страница 49: ...the WLAN network con nect PC and modules via the Ethernet network 6 6 UPS settings and shut down catman offers the option to define events and certain actions if the CX22B R switches into the UPS oper...

Страница 50: ...Operating the Data Recorder 50 A4722 1 0 HBM public CX22B R Notice To shut down CX22B R please unplug the power con nection Otherwise the data recorder will reboot after a few seconds automatically...

Страница 51: ...R start screen CX22Shell you can S Launch the catman EASY program S Change important CX22B R settings Settings You can use this to change the device name CX se rial number and the password for example...

Страница 52: ...g the remote desktop window S Switch to the system level Windows Exit to Win dows This is necessary for example when you want to ac tivate the CX22B R WLAN S Shut down the CX22B R Data Recorder Shut d...

Страница 53: ...g transducers with TEDS Transducer Electronic Data Sheet assign the connected sensors transducers to the channels find your sensors in the sensor list Sensor database component window then drag drop e...

Страница 54: ...tart via Trigger and Digital input for Trigger mode so that you can use the digi tal inputs of the CX22B R Start Stop Pre Trigger Post Trigger Fig 8 1 Starting and stopping measurement with a trigger...

Страница 55: ...will start The RUN LED flashes while waiting for the start trigger and glows continuously as soon as measurement starts Important The ERROR LED lights up if there is an error in cat man EASY e g if t...

Страница 56: ...ed Sensor database The sensor database is an easy way to adjust the mea suring chain to the sensor being used You only need to enter the characteristic values of a sensor once after which you can clic...

Страница 57: ...atabase under a different name to drive D or E and open it again catman EASY will automatically save the new path and file name and use it the next time the program is launched If you want to keep the...

Страница 58: ...t screen click on Exit to Windows and use the operating system Windows Explorer to copy the files 2 Data exchange via the clipboard Configure your remote desktop connection so that you can exchange da...

Страница 59: ...2 Entering the IP address of the CX22B R in Windows Explorer In the logon dialog enter the name of the CX22B R the user name and the password In the factory set tings the device name is CX serial num...

Страница 60: ...e name E There is no partition E listed in the standard setup However this letter will be assigned to the first external memory that is connected 8 3 Data analysis There are basically two ways to eval...

Страница 61: ...yMonitoring license or a catman AP license Parallel data recorders A data recorder is similar to a DAQ job in that is has its own start stop conditions storage file and other settings But while ordina...

Страница 62: ...e the video files can also be uploaded FTP configura tion is part of the DAQ job settings register tab Remote data storage Push Notifications Push notifications are short messages which can be send vi...

Страница 63: ...R for self contained measurements CX22B R A4722 1 0 HBM public 63 9 Configuring the CX22B R for self contained measurements Catman can be launched via the start screen Settings button Fig 9 1 CX22 cat...

Страница 64: ...edure Configure your DAQ project as described in Section 8 1 Set up the DAQ jobs so that the data are saved with the Automatically on DAQ termination setting Data saving in the Settings group Data sto...

Страница 65: ...will be reloaded and measurement continued when the power supply is restored The module restart up to the continuation of measurement will require approx 3 minutes depending on the number of connecte...

Страница 66: ...Configuring the CX22B R for self contained measurements 66 A4722 1 0 HBM public CX22B R...

Страница 67: ...onal change partition C EWF filter S 180 GByte are readily available partition D C Windows 8 Embedded including catman EASY D available 60 GB SSD SSD 180 GB Fig 10 1 Partitioning of the available memo...

Страница 68: ...s a rough guideline data throughput should be calcu lated in MByte s A signal sample can be calculated with 8 bytes The opportunity is also available to experi mentally check whether the chosen config...

Страница 69: ...nly stored temporarily and are deleted again once the sys tem is switched back on If you want to permanently save settings or programs on C you must use Commit changes and reboot in the start screen C...

Страница 70: ...ernet or WLAN In this mode the CX22B R act like a CX27 Important When the CX22B R is in gateway mode it cannot be used as a recorder This means that catman should run while the CX22B R is configured i...

Страница 71: ...mode 11 3 Connect to a QuantumX module behind a CX22B R To connect to a QuantumX SomatXR module being routed by a CX22B R look for this module in the cat man s device manager module list The modules...

Страница 72: ...ing the CX22B R as gateway 72 A4722 1 0 HBM public CX22B R Important note You can connect only time simultane ously to a QuantumX SomatXR module being routed by the CX22B R Successvie connections will...

Страница 73: ...ollowing table shows the different tests we made to benchmark the performances of the gateway mode in the CX22B R Maximum cumulative data througput gateway mode LAN Gigabit 1 WLAN interface adhoc 2 MS...

Страница 74: ...ctionality of the CX22B R Data Recorder can no longer be guaranteed and HBM can no longer provide support for any faults arising in this connection In practice drivers have already been successfully i...

Страница 75: ...ot in the start screen Also see EWF Chapter 10 3 on Page 68 12 2 Changing system settings Settings Important All changes in the Settings dialog are only permanently saved and active the next time you...

Страница 76: ...ge 63 S Waiting tine period before catman starts up S The language version of catman EASY language of the DAQ software the operating system and key board input General tab Select EN for the English in...

Страница 77: ...from a USB flash drive or from a network drive You can also update the MX Assistant software at any time 12 4 System restoration Recovery If the changes you have made to the system result in the syste...

Страница 78: ...nnection but not com pliant with the Ethernet specification If you experience problems the network search scan will not function Sensor failure or removing a sensor during measuring mode You have the...

Страница 79: ...mmunications Operation of this equipment in a residential area is likely to cause harmful interfe rence in which case the user will be required to cor rect the interference at his or her own expense A...

Страница 80: ...his documentation can cause harmful interferences for communication Operation of this device in residential areas can cause harmful interferences In this case the user is obliged clearing off the radi...

Страница 81: ...with national and local environmental protection and material recovery and recy cling regulations old devices that can no longer be used must be disposed of separately and not with normal household g...

Страница 82: ...product will comply with general hazardous sub stances limits for at least 20 years and will be ecological lly safe to use during this period as well as recyclable This is documented by the alongside...

Страница 83: ...ng 8 Connections 18 26 29 CX22 shut down 52 start screen 51 CX22 Start screen 51 D Device name change 51 DHCP 31 Digital inputs 22 Digital output reset 23 Digital outputs 22 E ERROR LED 25 EWF 68 I In...

Страница 84: ...77 Remote desktop close 52 Remote Desktop for Mobiles 41 Repair 7 S Sensor database 56 Settings 75 Stand alone mode 64 Start screen siehe CX22 Start Screen 51 Status LED operating status 25 Switching...

Страница 85: ...Index CX22B R A4722 1 0 HBM public 85...

Страница 86: ...www hbm com HBM Test and Measurement Tel 49 6151 803 0 Fax 49 6151 803 9100 info hbm com measure and predict with confidence A4722 1 0 7 2002 4722 HBM public...

Отзывы: