MP85A(DP) FASTpress, MP85A(DP)
−
S EASYswitch
22
A2073-5.0 en/de/fr
9
Error messages/operating state (LED display)
The relevant device status and error overview is clearly displayed in the
Sta-
tus overview
window. Open the window via the “Status” button in the
Mea-
sured value display
window.
The meaning of the displays and possible corrective measures for error mes-
sages are listed in the online help (use F1 to call the Help). Further helpful tips
and setting information can be found in the online help in the FAQs section.
Status
Depending on the display mode, various error messages can be displayed
instead of the measured value in the LCD-display of the MP85A process con-
troller (or the PME Assistant).
Current errors are displayed continuously. Press the
+
key to go to the “ER-
ROR” display mode.
Error message
Cause
Remedy
TransdErr
Input signal overflow
Transducer not connected
Incorrect transducer connection
Amplifier does not match transducer
type
No sense leads connected
Connect transducer
See pin assignment
Connect sense leads
ADC Ovfl.
Input signal of measurement chan-
nel A/D converter too large
Adapt hardware measuring range
(menu: Preparing for measure-
ment/Amplifier/Transducer)