Measuring Amplifiers on Eurocards ME30

The Eurocards ME30 is a 600 Hz carrier frequency amplifiers for strain gauge
transducers. An aluminium front panel (4 divs. wide) is delivered as standard.
Aplasticcs handle can be mounted instead of it if ordered separately.

All plug-in cards are delivered without housing without power supply unit so
that a 19” rack frame can be set up as required. Owing to the narrow width of
the plug-in cards (4 divs = 20.32mm) up to 21 cards can be inserted into a 19”
rack. The dimensions and connecting details correspond to the IEC standards
48D and 297.

All controls are accessible from the front.

Individual adjustments are described in chapter four; ex-works settings are
specified on the back of the Eurocard.

Notes for the diagram on the
right
S21:
calibration – measuring – zero
S22/23:
zero balancing (coarse)
S24/1+2:
MR1 fine balancing
internal/external
S24/3:
switch for measuring frequency
range 60Hz to 2Hz
S24/4:
MR changeover from MR1 to
MR2
S25:
switching the bridge excitation
voltage
S26/27:
setting of measuring range 
(coarse)
P21/22:
fine balancing of MR
P23:
zero balancing (fine)

Cal. +1mV/V
zero

MR1
fine balancing
external

5k

50k

zero balancing
external

zero
calibr.
MR1–2

wh

gy

bk

bu

gn

14.5...15.5V <50mA
14.5...15.5V <50mA

+ 9V...+35V
 (opt. DC/DC converter)

U

B

 +14.5...+15.5V

U

B

 –14.5...–15.5V

U

a

 

10V >5kOhm

option I

a

 

20mA

2.33V

ss

synchr.-voltage
or synchr. input

+

Fig. 1.1:

Block diagram

Содержание ME30

Страница 1: ...600 Hz Measuring Amplifier for strain gauge transducers measuring frequency 0 60 Hz ME30 Operating manual B 31 ME30 11 e ...

Страница 2: ......

Страница 3: ...12 4 Individual Settings 14 4 1 Operational Setting 14 4 2 Setting the Measuring Range 14 4 3 External Fine Balancing of the MR1 15 4 4 Bridge Fine Balancing 16 4 5 Calibration signal 17 4 6 Measuring Frequency Range Cut off Frequency fco 17 4 7 Synchronization 17 5 Options 18 5 1 Stabilized Power Supply 18 5 2 DC DC Converter 18 5 3 Current Output Module 19 5 4 Safety Barriers for use in Hazardou...

Страница 4: ...g to the speci fications given in this manual When using the transducer the legal and safety regulations for the respective application must also be observed The same applies if accessories are used General dangers in the case of non observance of the safety instruc tions The ME30 complies with the state of the art and is operationally reliable If the device is used and operated inappropriately by...

Страница 5: ...ddition de sign realise and take responsibility for the measuring system s safety such that potential residual dangers are minimized The respective regulations must in any case be observed Residual dangers regarding the measuring system must be specified explicitly If there is any risk of remaining dangers when working with the it is pointed out in this introduction by means of the following symbo...

Страница 6: ...e to property or lead to some form of physical injury Symbol NOTE Means that important information about the product or its handling is being given Symbol Meaning CE mark The CE mark is a guarantee by the manufacturer that his product meets the requirements of the relevant EG Directives see declaration of conformity at the end of these operating instructions ...

Страница 7: ...aced use original HBM components only Qualified personnel The device may be used by qualified personnel only the technical data and the special safety regulations must in any case be observed When using the device the legal and safety regulations for the respective application must also be observed The same applies if accessories are used Qualified personnel means personnel familiar with the insta...

Страница 8: ... are accessible from the front Individual adjustments are described in chapter four ex works settings are specified on the back of the Eurocard Notes for the diagram on the right S21 calibration measuring zero S22 23 zero balancing coarse S24 1 2 MR1 fine balancing internal external S24 3 switch for measuring frequency range 60Hz to 2Hz S24 4 MR changeover from MR1 to MR2 S25 switching the bridge ...

Страница 9: ...in 30a sensing line to 32a pin 32c bridge supply voltage pin 32a measuring signal pin 30c bridge supply voltage pin 28a sensing line to 28c pin 28c contact alternatively 29a c Fig 2 1 Bridge connection The required bridge excitation voltage set by the factory is 5V symmetrical about ground It can be switched to 2 5V by switch S25 UB Bridge resistance Switch S25 2 5V RB 60 4000Ω 5V ex works setting...

Страница 10: ... at a low value b unstabilized double pole supply voltage If a stabilized voltage is not available the amplifier can be operated by means of a stabilizer unstabilized supply or a DC DC converter single pole supply battery see chapter 5 1 and 5 2 2 3 Output output voltage 10V RL 5 kΩ operating voltage zero The output is provided for the connection of a display and or recorders A current output is p...

Страница 11: ...ensing line gn 29c operating voltage zero 30c measuring signal wh 31c operating voltage zero ye 32c sensing line gy St 21 1a not occupied 2a not occupied 3a not occupied 4a DC DC converter positive pole 5a DC DC converter negative pole 6a not occupied 7a output Uadd 14 5 15 5V 50mA 8a not occupied 9a not occupied 10a not occupied 11a not occupied 12a not occupied 13a internally connected 14a inter...

Страница 12: ...ridge Zero Balancing Carry out bridge zero balancing by means of the screwdriver potentiometers P23 fine and switches S22 coarse and S23 polarity See also chapter 4 4 3 3 MR Fine Adjustment The fine adjustment of the measuring range is made seperately for both ranges by means of the screwdriver potentiometers P21 measuring range 1 and P22 measuring range 2 The potentiometers in the front panel per...

Страница 13: ...13 ME 30 P21 P22 ME30 S21 P23 S22 S23 Fig 3 1 Front view ME30 ...

Страница 14: ...ied to the output This function enables amplifiers to be replaced in measuring chain without the need for rebalancing the bridge with the unloaded transducer All that is required is that the amplifier balance values are noted and then this value is set if necessary with P23 S22 and S23 4 2 Setting the Measuring Range Measuring range 1 is get ex works Measuring range 2 can be activated by switching...

Страница 15: ...o x 0 90 1 28 o o o x o o x o 1 18 1 69 o o o x o o o x 1 56 2 23 o o o o x o x o MR 1 2mV V 2 07 2 94 o o o o x o o x 2 73 3 88 o o o o o x x o 3 61 5 13 o o o o o x o x o open x closed 4 3 External Fine Balancing of the MR1 Connections potentiometer pin 9c wiper pin 8c potentiometer finish pin 7c An external 5kΩ potentiometer can be connected for fine adjustment of the range This will enable abo...

Страница 16: ...g A balance range of 0 12mV V set by the manufacturer can be passed with an externally connected potentiometer of 50 kΩ Pin assignment potentiometer pin 26c wiper pin 21c potentiometer finish pin 24c 4 5 Calibration signal Switching switch S21 to the right calibration detunes the amplifier by a de fined amount 1mV V The measuring range can be adjusted without hav ing to load the connected transduc...

Страница 17: ... ex works setting fco 80Hz 3dB 0 2Hz 1dB fco 2 5Hz 3dB 4 7 Synchronization When operating several ME30 amplifiers in a common housing one of them has to be switched as the master the carrier frequency of which supplies the clock pulses the others are to be switched as slave Pins 11c of all amplifiers have to be joined for synchronization The master amplifier is selected by switch S28 Synchronizati...

Страница 18: ...pole pin 16a supply zero pin 19c 5 2 DC DC Converter The manufacturer can mount a DC DC converter onto the board for the oper ation of a single pole supply voltage battery The voltage converter is suit able for operation on 9 35V unstabilized Pin assignment positive pole pin 4a negative pole pin 5a The DC DC converter decouples the internal operating voltages from the sup ply voltage When using th...

Страница 19: ...igned to each other The required fine calibration can be carried out from the front panel as described in chapter 3 5 4 Safety Barriers for use in Hazardous Areas For transducer operation in potentially explosive atmospheres the safety barri ers SD01 are used conforming to the protection classes EEx ib IIC accord ing to VDE0171 EN 50 014 and EN 50 020 see data sheet 5 5 Additional Units The follow...

Страница 20: ...bserved in plants with strongly disturbed supply mains systems The most important points are listed up in the HBM leaflet Hints for connection disturbances An extensive summary is given in the installation regulations VDI VDE 3551 The extent of certain measures to be taken depends on the specific applications ...

Страница 21: ...V 0 2 Calibration signal mV V 1 0 1 Bridge balance range coarse balance adjustable in 16 steps and polarity mV V 2 fine balance with screwdriver potentiometer mV V 0 08 Measuring frequency Butterworth low pass 3rd order switchable at 1dB Hz 0 2 0 60 at 3dB Hz 2 5 80 phase delay ms 135 4 8 rise time ms 170 7 surge voltage overshoot 10 10 residual carrier 0 1 0 2 typ 0 1 Input symmetrical input resi...

Страница 22: ... Operating temperature range C 20 60 Storage temperature range C 25 70 Stabilized Voltage for the operation of additional units V 15 max power consumption mA 50 Supply current standard stab KM001 DC DC converter Supply voltage V 14 5 15 5 15 6 25 9 35 max current consumption without additional units mA 70 65 80 70 340 140 influence of supply voltage for changes in the relevant range the measuring ...

Страница 23: ...23 ME 30 8 Component Layout ...

Страница 24: ...24 ME 30 9 Copy of Declaration of Conformity ...

Страница 25: ...25 ME 30 ...

Страница 26: ...26 ME 30 ...

Страница 27: ...27 ME 30 ...

Страница 28: ...iefen See 45 D 64293 Darmstadt Tel 49 61 51 8 03 0 Fax 49 61 51 89 48 96 www hbm de e mail TSC hottinger baldwin com Modifications reserved All details describe our products in general form only They are not to be understood as express warranty and do not constitute any liability whatsoever IM D 09 99 POD ...

Отзывы: