Electrical connection
24
A03729_03_YI0_01
HBM: public
C9C
Pin
Version VA 1
(voltage out
put)
Version VA 2
(current out
put)
KAB168
connection
cable wire
assignment
1
Supply voltage 0 V (GND)
white
2
Not in use
brown
3
Control input zero setting
green
4
Not in use
yellow
5
Output signal
0 ... 10 V
Output signal
4 ... 20 mA
gray
6
Output signal
0 V
Not in use
pink
7
Not in use
blue
8
Voltage 19 ... +30 V
red
The cable connecting the inline amplifier to the next link in the measurement
chain must not exceed 30 m in length.
7.2.3
Operating the inline amplifier/zeroing the measurement chain
The measurement starts as soon as the sensor is connected to a supply
voltage and the output of the amplifier is connected to the next link in the
measurement chain.
If you apply a voltage > 10 V to the “Zero” input, a one-time zeroing is
performed. After this zeroing, the device continues to measure, even if you
apply a voltage above 10 V to the input.
To trigger a new zeroing operation, the input must first be set to 0 V and then
be reset by applying a voltage of over 10 V.
Notice
Please note that you can zero the measurement chain with any force applied. If
an initial load is already acting on the force transducer, it is essential to
consider it, otherwise the force transducer may be overloaded.
Содержание C9C
Страница 36: ...Specifications 34 A03729_03_YI0_01 HBM public C9C...
Страница 39: ...C9C A03729_03_YI0_01 HBM public 3 8 Abmessungen 26 9 Technische Daten 29 9 1 Ausf hrungen und Bestellnummern 33...
Страница 70: ...Technische Daten 34 A03729_03_YI0_01 HBM public C9C...
Страница 107: ...C9C A03729_03_YI0_01 HBM public 3 8 Dimensioni 26 9 Dati tecnici 29 9 1 Versioni e No Ordine 33...
Страница 138: ...Dati tecnici 34 A03729_03_YI0_01 HBM public C9C...
Страница 139: ...Dati tecnici C9C A03729_03_YI0_01 HBM public 35...