Connection
BN100A
A0643-6.1
HBM: public
11
Microfuse replacement after blowing:
S
Use screwdriver to squeeze out fuse carrier
S
Replace fuses
S
Reinsert fuse carrier
CAUTION
Be sure to have selected the appropriate mains voltage
connection
230V: 0,2AT fuse
115V: 0,4AT fuse
The bridge calibration unit is connected to the measuring
chain in place of the transducer with the socket Bu1.
A:Measuring signal (+)
white
B:Bridge excitation voltage (-)
black
C:Bridge excitation voltage( +)
blue
D:Measuring signal (-)
red
E:Three inner cable shields (measurement ground)
F:Sensor circuit (+)
green
G:Sensor circuit (-)
grey
A
F
B
G
E
C
D
Fuses:
230 VAC
2 x 0,2 A sl
115 VAC
2 x 0,4 A sl
Connector housing: Outer cable shield (ground potential)
Fig. 3.2
Socket assignment