Electrical connection
16
A05449_01_Y00_02
HBM: public
C11
8.1
EMC protection
Electrical and magnetic fields can often induce interference voltages in the
measuring circuit. Pay attention to the following points to avoid this:
S
Only use shielded, low-capacitance measurement cables for extensions
(HBM measurement cables fulfill these conditions).
S
Do not route the measurement cable parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with metal
tubing
.
S
Avoid stray fields from transformers, motors and contact switches
.
S
Do not ground transducers, amplifiers and indicators more than once.
S
Connect all the devices in the measuring chain to the same protective
conductor.
S
Always connect the cable shield extensively on the amplifier side and on the
transducer side, to create the best possible Faraday cage.