Harvest Healthcare WOBURN 6 Скачать руководство пользователя страница 6

T

 +44 (0)1709 377172     

F

 +44 (0)1709 377173

E

 [email protected]      

www.harvesthealthcare.co.uk

10

11

GENERAL INFORMATION

GENERAL INFORMATION

Unplug the mains plug from the socket before moving the care bed and take care 

to avoid dragging the mains plug across the floor when moving the bed. 

The mains plug must always remain accessible to enable immediate cut-off by
unplugging the mains plug from the wall socket in case of emergency. The mains
cable must be free and not caught in anything, as it gets carried along when the
bed height is adjusted. Otherwise, the mains cable may be torn out and damaged.
In addition, the mains plug may be pulled out of its socket and electric leads
exposed as a result. If the mains cable or the mains plug are damaged, the relevant
part must be replaced. This work should only be carried out by the manufacturer or
authorised professionals.

When connecting the mains plug do not use multiple sockets since liquids may 

penetrate into these (fire hazard and electric shock).

Before cleaning and disinfection the mains plug must be unplugged 
and hung up safely. Plugs for the handset and the motors which are 
inserted into the mattress base control box and motor unit must 
remain plugged in. This is necessary to prevent water ingress into the 
control box.

When the bed is stationary the castors must always be in the braked position. If
the castors are not braked, the bed can move when the occupant gets into and
out of bed, since the occupant uses the bed for support. Injury can result if the
care bed rolls away.

In order to move the care bed, the brakes on all four castors must be released and
the mattress base be adjusted to the lowest horizontal position.

The maximum duty cycle and the safe working load must not be exceeded
otherwise safe operation cannot be guaranteed (please refer to the Technical Data
in section 14).

The bed must not be used in rooms where there is a risk of explosion.

4.2  

SAFETY INFORMATION FOR THE OPERATOR 

With the help of this Instruction Manual, instruct each user in the safe operation of 

this care bed before it is put into service for the first time.

Advise the user of any hazards which may occur if not handled correctly. 

Only persons who have been properly instructed may operate this care bed. This also applies 
for persons who only operate the care bed on a temporary basis. 

According to the Medical Products Act (German abbreviation: MPG, Medizinproduktgesetz), 
care beds are Class I active medical products.

Please observe your obligations as the operator in accordance with the Operators
of Medical Products Ordinance (Medizinprodukte-Betreiberverordnung, German
abbreviation: MPBetreibV), in order to ensure the permanently safe operation of this medical 
product with no risk of danger to patients, users or third parties. If the care bed is used on a 
long-term basis, checks for proper functioning and for any visible damage must be performed 
and documented at least once a year. There is section 

10.2

 for this purpose.

4.3 

SAFETY INFORMATION FOR THE USER

Ensure that the operator instructs you in the safe operation of this bed.

In addition, pay particular attention to the general safety information as described in 

4.1

.

Adjustments of the bed must only be carried out by suitably instructed persons or in the 
presence of an instructed person. 

Make sure that the mattress base has travelled to its lowest position before leaving the patient 
unattended. This will minimise the risk of injury to the patient when getting in or out of bed.

If there is a suspected fault or damage, unplug the mains plug from the socket. Clearly mark 
the care bed as “Out of Order” and take it immediately out of service. Please inform the 
person in charge without delay. 

Содержание WOBURN 6

Страница 1: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 1 General User Safety Guide WOBURN 6 AND WOBURN LOW 4 PROFILING BEDS www harvesthealthcare co uk ...

Страница 2: ...ualification of Users Safety Instructions General Safety Instructions Safety Information for the Operator Safety Information for the User Cleaning and Disinfection Servicing and Maintenance Accessories Electromagnetic Compatibility Transport and Storage Service Life and Disposal PRE INSTALLATION CHECK INSTALLATION AND COMMISSIONING Removal from the Transporting Device Assembly of the Care Bed Conn...

Страница 3: ...ely Bruce McEwan Sales Director Harvest Healthcare Ltd READ THIS INSTRUCTION MANUAL AND OBSERVE SAFETY INSTRUCTIONS CAUTION Please read and observe this instruction manual before each use In the event the care bed changes owners please supply this instruction manual to the new owner When the bed is moved on the transport frame take care not to allow it to overbalance The narrow tall design saves s...

Страница 4: ...is symbol carefully and urgently follow instructions This information is safety relevant This symbol indicates hazards due to electrical voltage There is mortal danger This symbol indicates general hazards There is danger to life and health Conformity mark in accordance with the Medical Device Directive 93 42 EEC The electrical equipment is splash proof Symbol for Protection Class II device double...

Страница 5: ...patients or occupants who have not been instructed in its use Use of the care bed for children Attempting to move the care bed when castors are braked Use of the care bed on a non horizontal surface max incline 5 3 GENERAL REGULATIONS FOR USERS The care bed must only be used for the purpose intended When setting up operating and using the care bed respect the regulations in your country and the ge...

Страница 6: ... exceeded otherwise safe operation cannot be guaranteed please refer to the Technical Data in section 14 The bed must not be used in rooms where there is a risk of explosion 4 2 SAFETY INFORMATION FOR THE OPERATOR With the help of this Instruction Manual instruct each user in the safe operation of this care bed before it is put into service for the first time Advise the user of any hazards which m...

Страница 7: ...nded the battery is replaced every two years if not used Ensure it is a type 6LR61 alkaline manganese battery and not any other type Used batteries must be disposed of in an environmentally compatible way Please check all fixings on your bed at least once a month Pay special attention to sleeping platform connections GENERAL INFORMATION GENERAL INFORMATION 4 6 ACCESSORIES The optional accessories ...

Страница 8: ...6 INSTALLATION COMMISSIONING Harvest Healthcare Ltd recommends that a risk assessment is completed by the Operator before this bed is assembled G Lift the cover from the bed unit and transporting device Please do not dispose of the cover It can be used again as a dust cover in the event that the care bed is later stored in the transport rack Do not remove the cardboard protection from the top of t...

Страница 9: ...s on the bed castors by pressing the brake pedal and then release the spring loaded catches see 6 2 at the height adjustable head and foot end panels so the mattress base can be slotted onto the mounting lugs Remove the height adjustable head and foot end panels from the transporting device To do this release the spring loaded catches that are later used to fix the height adjustable head and foot ...

Страница 10: ...the mattress base off the mounting lug and pull the mattress base back far enough to allow you to push the side guards onto the location bolts Slide both side guards onto the location bolts Lift one corner of the mattress base slightly Pull back the bed end panel with mounting lug With the side guards slotted into place the end panel can be pushed back up to the mattress base and secured with the ...

Страница 11: ...embly Clean and disinfect the bed as described in chapter 8 before putting into service and before each further use 6 5 DISASSEMBLY OF THE CARE BED Remove the mains plug from the socket before disassembly Disassembly of the care bed is carried out in reverse order of assembly When reassembling the care bed sections onto the transport brackets please ensure you align the red dots on the headboard s...

Страница 12: ...ITH LOCKING FUNCTION To avoid damage the handset should always be hung up by the handset hook e g on mattress base when not in use Press only one button at a time as the system could overload and become damaged 7 3 LOCKING FUNCTION FOR THE HANDSET On the back of the handset there is a locking device All electric adjustment functions can be blocked at the same time using the nurses key supplied The...

Страница 13: ... failure Please note however that this is only possible once per 9V battery as the capacity of the 9V battery is limited After the emergency lowering has been used once the 9V battery must be replaced Type 6LR61 alkaline manganese battery The 9V battery should however be replaced every 2 years even if it has not been used 7 5 2 BATTERY CHANGE To replace check or remove before lengthy storage of th...

Страница 14: ...ion piece until it is fully released Please note when the extension piece has fully released it will not separate from the bed Once fully extended tighten all 4 hand screws If you have side rails fitted these will need to be removed before extending the platform It is important to regularly check all hand screws and snap pins for security In Built Extension Retracted In Built Extension Extended OP...

Страница 15: ...tor Motor unit not plugged in Plug motor unit into mattress base motor No Response Connecting cables on the connectors transposed Check plugs and connectors and change over plugging in locations Adjustment functions transposed 9V block battery is discharged Replace 9V block battery No function after power failure Bed only adjusted via the battery Mains plug not plugged in Plug in mains plug and re...

Страница 16: ...y signs of wear 12 Have castors including locking brake been tested for safe functioning 13 Mains cable connecting cables and plugs without damage 14 Fixture available for safe transportation of mains plug 15 Strain relief of the mains cable and handset securely attached 16 Are all plug in connections securely attached Washers without damage 17 SERVICE RECORD Are cables laid correctly and safely N...

Страница 17: ...the adjustment buttons no motorised adjustments should be possible Set switch on the back of the handset to testing position 1 symbol I When operating the adjustment buttons no motorised adjustments should be possible Set switch on the back of the handset to testing position 2 symbol II When operating the adjustment buttons no motorised adjustments should be possible SERVICING 12 GUARANTEE As stat...

Страница 18: ...06 kg Operating noise 53 db A at a distance of 1m 14 2 TECHNICAL DATA ELECTRONIC Power supply unit LIMOSS control unit MC220 SMPS MC125 Voltage rating 230 240V Frequency rating 50 60 Hz Type of current AC Nominal consumption during operation 70 Watt Nominal consumption in idle state 0 5 Watt Nominal operating time Nominal idle time 2 Min 18 Min max 5 switching cycles min Primary safety fuse 2 0 A ...

Страница 19: ...uch an environment Group 1 The care bed uses RF energy only for its internal functioning Therefore its RF emissions are very low and it is unlikely that nearby electronic devices will be disturbed Class B The care bed is designed for use in all establishments including domestic establishments and those determined to be directly connected to a public supply network that supplies buildings used for ...

Страница 20: ...ions in energy supply demands it is recommended to feed the care bed from an uninterruptible power supply or a battery 3 A m Magnetic field of power frequency 50 60 Hz according to IEC 61000 4 8 3 A m 0 3 A m Magnetic fields of power supply frequency should comply with the typical values as can be found in a business and hospital environment INTERFERENCE IEC 60601 ELECTROMAGNETIC ENVIRONMENT GUIDE...

Страница 21: ...l output of the transmitter in watts W according to specifications of the transmitter manufacturer NOTE 1 An additional factor of 10 3 is applied when calculating the recommended working clearance between transmitters in the 80 MHz to 2 5 GHz frequency range in order to reduce the probability that a mobile portable communication device unintentionally brought into the patient area could lead to in...

Страница 22: ...T 44 0 1709 377172 F 44 0 1709 377173 E sales harvesthealthcare co uk www harvesthealthcare co uk 42 43 DATE OF PURCHASE Date of purchase Distributor stamp You can fix your receipt here NOTES ...

Страница 23: ...T 44 0 1709 377172 F 44 0 1709 377173 E sales harvesthealthcare co uk www harvesthealthcare co uk 44 45 NOTES ...

Страница 24: ...rvest Healthcare Harvest Healthcare Limited Company No 07210261 Our Full Terms Conditions are available by request or can be found on our Website Harvest Healthcare reserves the right to alter or amend this document without notice Serial No DOCUMENT REFERENCE HLB791 03 Woburn 6 HLB797 04 Woburn Low 4 January 2017 All parts and data continually undergo further development and may therefore deviate ...

Отзывы: