
T
+44 (0)1709 377172
F
+44 (0)1709 377173
E
www.harvesthealthcare.co.uk
32
33
SERVICING
11.4
MEASUREMENT OF OVERALL ELECTRICAL SYSTEM
The measurements described here must only be performed by a
qualified electrician or by an electrotechnically trained person, (using
suitable measuring and testing devices).
The measurements shall include as a minimum the testing of the housing leakage current and
the measurement of the isolation resistance.
The following measured values must be attained:
*Housing leakage current
<= 0.2mA
*Isolation resistance
<= 7M
Ω
During testing the corresponding button on the handset must be constantly pressed.
The measurement is to be performed between:
• *The control unit and the bed frame
• *The control unit and the handset
11.3 CHECKING THE INITIAL FAULT SAFETY BY MEANS OF THE INTEGRATED
CONTROL BOX IN THE HANDSET
The switching positions I and II are testing settings
used only to check the safety during the annual
inspection, or after repair work, or each time bed is
put into service again.
To check the safety equipment, proceed as follows:
Setting switch position 4 (padlock symbol open
).
Move all bed adjustments to a slightly raised position.
Setting switch position 3 (padlock symbol closed ).
When operating the adjustment buttons, no motorised adjustments should be possible.
Set switch on the back of the handset to testing position 1 (symbol I).)
When operating the adjustment buttons, no motorised adjustments should be possible.
Set switch on the back of the handset to testing position 2 (symbol II).)
When operating the adjustment buttons, no motorised adjustments should be possible.
•
•
•
•
SERVICING
12 GUARANTEE
As stated in our Standard Terms and Conditions, we provide a manufacturer’s warranty of 5
years from the date of purchase.
To take advantage of the 5 year warranty, the bed must be serviced (without exception)
every 12 months by a Harvest Healthcare Ltd approved technician using only Harvest
Healthcare Ltd original spare parts. A service record must be completed (an example can be
found on
pages 28-29)
.
13
SERVICE LIFE & DISPOSAL
The service life of our care beds in domestic use is assumed to be approximately
5 years. This naturally depends upon the manner of use. The carebed is suitable
for reuse if all measures of section 6.3 and 10 are taken. Frequent transportation,
setting up and adjustment reduce the service life, as do improper treatment,
irregular servicing and exceeding the safe working load or the admissible load
cycle of the electric motors. The care bed must not be disposed of as normal
household waste after the end of its service life. To ensure that it is disposed of in
an environmentally compatible way please contact Harvest Healthcare Ltd.