9
4 Setting up and Connecting the MEA Signal Generator
A MEA-SG is handled in the same manner as a MEA probe for experiments. For detailed information please
read the MEA Manual.
Mounting instructions:
1.
Open the lid of the amplifier.
2.
Place the respective MEA-SG inside. Please orientate the chip as shown on the picture.
3.
Replace the lid and close it carefully.
4.
Ground the MEA-SG with a CB-GND cable.
4.1 Handling a MEA-SG
After starting the software, in this case MC_Rack and adding a data display you see your selected signal,
in this case 12.5 Hz sinus waves. Please adjust the axes via drop down menus or slider in the header of the
display: X-axis 1000 ms, y-axis +/- 50 µV.
MC_Rack data display
On this MC_Rack data display channel 66 gets saturated. Check whether this channel is grounded.
If not, the gold spring contacts in the lid of your amplifier may be defective. Please see chapter
"Troubleshooting".