background image

5

Publication 5502-018-REV 1.0

1. Prior to use of electrode, spray with 

Rain-X and gently shake over platinum 

electrode surface of electrode for 

10 seconds. Discard Rain-X. Sterilize 

Oocyte Petri Electrode by spraying 

three times with 75% ethanol, then 

drying with a Kimwipe. Rinse with 

sterile MilliQ water, then dry with a 

Kimwipe. Electrode is now ready to use.

  Note: Sigmacote may be used as an 

alternative to Rain-X to coat electrodes 

prior to use.

2. Connect the Mini Micro-Grabber cables 

to the Oocyte Petri Electrode, plug the 

banana cable into the Tweezertrode 

Adapter Cables and then connect the 

Adapter Cables into the voltage output 

of the BTX Electroporator. 

3. Following instructions for the BTX 

Electroporator, set the appropriate 

parameters for electroporation.

4. Use a wide bore pipette tip to transfer 

oocytes or zygotes to the 1 mm gap 

between the two electrodes on the 

slide. Suggested fill volumes are 7 µl 

for 70 embryos and 10 µl for 100–120 

embryos.

5. Deliver the electroporation pulse(s) to 

the sample.

6.    Recover sample with wide bore 

 

pipette tip. During a session of 

multiple electroporations using the 

same transfectant, excess buffer may 

be removed with a sterile eye spear in 

between rounds of electroporation.

7.    Disconnect from Mini Micro-Grabber 

cables and clean electrode by flushing 

with distilled water.

8.    Sterilize Oocyte Petri Electrode by 

spraying with 75% ethanol three times 

and allow to dry.

9.   Electrode is ready for next sample. 
10. If finished with electroporation, 

remove Mini Micro-grabber cables 

 

from Oocyte Petri Electrode.

11. Gently flush electrode gap with 

 

a mild detergent and rinse with 

distilled water.

12. Dry electrode with 75% ethanol by 

flushing electrode with alcohol and 

gently shake any remaining drops.

13. Allow to air dry until no alcohol is seen 

and store with glass cover in a dry 

area. 

Make sure the BTX Electroporator is switched off before continuing.

WARNING: High Voltage

Operation: Getting Started

Содержание BTX Oocyte Petri Electrode 45-2059

Страница 1: ...Oocyte Petri Electrode 45 2059 45 2060 15 mm Oocyte Petri Electrode Platinum 1 mm gap electrode only 15 mm Oocyte Petri Electrode Platinum 1 mm gap Kit 45 2059 45 0503 45 0204 User s Manual ...

Страница 2: ...n If the unit is replaced the replacement unit is covered only for the remainder of the original warranty period dating from the purchase of the original device This warranty gives you specific rights and you may also have other rights which vary from state to state Note BTX electrodes are not recommended for use with power supplies or cables from other manufacturers Such use is completely at the ...

Страница 3: ... medium might lead to flashover between the electrodes ARC and or explosive evaporation of the medium Reduce voltage or pulse length to avoid repeating this condition DO NOT OPERATE WITH SUSPECTED FAILURES If you suspect there is damage to the product have it inspected by qualified BTX service personnel DO NOT CONTACT ELECTRODES To avoid fire or shock hazard observe all ratings and markings on the...

Страница 4: ...es can be visualized during electroporation This electrode is particularly useful for the generation of transgenic animals including use with CRISPR Cas 9 constructs to both knock in and knock out targeted genes The Oocyte Petri Electrode may be cleaned with a mild detergent and sterilized with 75 ethanol This electrode requires the use of Mini Micro Grabber Cables 45 0503 and Tweezertrode Adapter...

Страница 5: ...s on the slide Suggested fill volumes are 7 µl for 70 embryos and 10 µl for 100 120 embryos 5 Deliver the electroporation pulse s to the sample 6 Recover sample with wide bore pipette tip During a session of multiple electroporations using the same transfectant excess buffer may be removed with a sterile eye spear in between rounds of electroporation 7 Disconnect from Mini Micro Grabber cables and...

Страница 6: ...20 to 80 Physical Characteristics Item No Electrode Length Electrode Gap Electrode Material 45 2059 15 mm 1 mm Medical Grade Polished Platinum Compatibility Generators BTX Gemini X2 ECM 830 ECM 2001 Please contact BTX Technical Service at any of the numbers listed below in the event of any failure BTX A Division of Harvard Bioscience 84 October Hill Road Holliston MA 01746 1388 U S A Toll Free 1 8...

Отзывы: