6
H a r v a r d
A
p
p a r a t u s S y r i n g e P
u
m p M o d e l ‘ 1 1 ’
Fuse Replacement
Make sure the power cord is disconnected from the main supply before servicing
the fuse.
Fuse Replacement Procedure
Turn off power, remove power cord from power module. Use straight blade screw
driver, pry open access door. Remove fuse holder, remove fuse(s) from holder.
Replace fuse(s) and fuse holder. (Replace holder with correct voltage marking is
showing through access door window) For continued fire protection, replace fuse
only with 250V fuse of the specified type and rating. (T 1/4 A 250V)
Fuse not replaceable by the operator is located on the power supply. Fuse
F 2.0A 250V.
Fuse Replacement
WARNING:
USE IN MANNER NOT SPECIFIED
BY THE MANUFACTURER MAY
IMPAIR THE PROTECTION
PROVIDED BY THE EQUIPMENT.