10
H a r v a r d
A
p
p a r a t u s S y r i n g e P
u
m p M o d e l ‘ 1 1 ’
3. Display Intensity
For varying light conditions, four levels of intensity can be set on the vacuum
fluorescent display.
To set the desired display intensity, at the RATE VOL CONFIG display, move
the underline cursor, using the
▲
▲
or
▼
▼
key, to CONFIG. Press the SET key
to enter the CONFIG mode options. Use the
▲
▲
or
▼
▼
key to scroll the CON-
FIG options until you reach the SET:INTENSITY:. Press the SET key again
moving the underline cursor to the right. Use the
▲
▲
or
▼
▼
key to scroll the
SET:INTENSITY choices; “1”, “2”, “3” or “4” (4 is highest intensity, 1 is low-
est intensity) Press the ENTER key to select the desired display intensity. Exit
the CONFIG mode by pressing the ENTER key again and save the INTEN-
SITY setting in memory.
4. Changing Rates
If the pump is running at an existing rate it will continue to do so until a new
rate is entered. Except for volume mode, the flow rate can be changed while
the pump is running. As soon as the ENTER key is pressed the pump will
change to the new flow rate. To change rates from the keypad, while in vol-
ume mode, the pump must be stopped first.
5. Maintenance
No special maintenance is required other than keeping the pump 11 clean and
dry. Avoid liquid spills that may find their way into the electronics.
A small container of grease is provided for periodic lubrication of the lead
screw. A small bottle of oil is provided for periodic lubrication of the two steel
guide rods. It is important to keep these guide rods clean and oiled.
6. Protecting Small, Fragile Syringes
The Pump 11 will hold microliter size syringes down to 0.5µl size. These
small syringes have fine wire plungers that may be damaged if allowed to bot-
tom out. The Pump 11 is equipped with an adjustable stop collar on one of
the guide rods. Use the provided hex wrench to position and lock the stop col-
lar to prevent damage to small syringes.
Operation: Advanced Features