AFB-PRO User Manual                                                             

Page 5 of 8

 

 
 

 

 

3.0

 

Instrument Setup 

 

3.1

 

Initial Start-Up 
 

3.1.1

 

Before plugging the wall transformer into an AC outlet, connect the round    
barrel receptacle to the instrument. The Connector is located on the left-
hand side of the instrument directly under the on/off switch. 

3.1.2

 

Turn the power switch off and plug the wall transformer into the unit. 

3.1.3

 

Place the Reagent Pack on the shelf behind the instrument and connect the 
Reagent Pack phone type connector to the receptacle on the right-hand 
side of the instrument. If the Reagent Pack is not plugged in when the unit 
is turned on, the message "SELF-TEST FAILURE: Pack not connected" 
will appear. Turn unit off, connect Reagent Pack and turn unit on to 
continue. 

 

3.1.4

 

Remove the caps from the Reagent Pack and insert the properly labeled 
reagent cannulas in the appropriate bottles. Customer needs to supply DI 
water container and place DI labeled line in the water for the rinse cycles. 

3.1.5

 

Place the waste tube in the waste container. 

3.1.6

 

Turn the instrument on. The instrument will go through a self-test and then 
display “prime” or “continue” 

3.1.7

 

Press <F1> to prime. Two primes are required on initial set up. 
 

4.0

 

Operating Instructions 

 

4.1

 

General Guidelines 

4.1.1

 

Make sure the specimen smears are thoroughly dry before placing them 
into the cuvette.

 

4.1.2

 

Use high quality clean slides. Slides that have variable thickness may 
have variable staining results.

 

 

CAUTION:

  

The AFB-PRO will only accept stain packs with an included module sold by 
Hardy Diagnostics’ QuickSlide™ to ensure optimal results. The instrument is 
calibrated only for these reagents. The use of other stain packs or solutions will 
void the warranty. 

 

Always remove the tubing kit cannulas from bottles before 
moving unit  

 

Transport unit with care 

 

Dropping the unit may cause internal and external damage to the 
instrument 

 

Never operate the unit with damaged power supply 

 

Repairs are to be carried out only by qualified service personnel 

Содержание QuickSlide AFB-PRO

Страница 1: ......

Страница 2: ...ty Safety Instructions 4 2 1 Explanation of Safety Notes 4 2 2 Safety Instructions 4 3 0 Instrument Setup 5 4 0 Operating Instructions 5 4 1 General Guidelines 5 4 2 Stain Procedure 6 4 3 Reagent Replacement 6 4 4 Menu Options 7 4 5 Setting Fill Levels 7 5 0 Maintenance 7 6 0 Service and Supplies 8 7 0 Safety Data Sheets 8 ...

Страница 3: ... 9001 2008 All products are shipped to you from our manufacturing facility located at 3411 McNiel Ave Suite 302 Wichita Falls Texas 76308 Unpacking and Inspecting Carefully unpack the AFB PRO and accessories Check for damage incurred during transit Keep all packing material until you are sure the unit operates properly Any damage to the shipping box should be reported to the responsible carrier or...

Страница 4: ...personnel is also advised to refer to this Operating Manual and or call the QuickSlide Technical Support phone number for additional assistance 2 1 Explanation of Safety Notes 2 2 Safety Instructions CAUTION Describes a possibly dangerous situation If this is not avoided slight or minor injuries could result A warning of possible property damage may also be contained in the text CAUTION Instrument...

Страница 5: ... water container and place DI labeled line in the water for the rinse cycles 3 1 5 Place the waste tube in the waste container 3 1 6 Turn the instrument on The instrument will go through a self test and then display prime or continue 3 1 7 Press F1 to prime Two primes are required on initial set up 4 0 Operating Instructions 4 1 General Guidelines 4 1 1 Make sure the specimen smears are thoroughly...

Страница 6: ...d remove the empty Reagent Pack Don t forget to empty the waste container at this time 4 3 4 Place the new Reagent Pack on the instrument connect the NEW phone connector from the NEW Reagent Pack and place the reagent cannulas in the new reagent bottles and insure DI water line is in a source for DI water 4 3 5 Turn the instrument on and the Self Test menu will be displayed The cuvette is automati...

Страница 7: ...nes must be replaced every 6 to 12 months depending on use To perform this procedure do the following 5 1 Remove the lines from the stain bottles 5 2 Prime the unit twice to remove fluid from the old tubes 5 3 Shut the unit off 5 4 Unscrew the two thumbscrews holding the pump cover in place 5 5 Remove the lines one at a time and replace starting from left to right The longest lines are to be posit...

Страница 8: ...ort team at 800 295 9588 If additional supplies are needed reagents and tubing kits can be ordered through Hardy Diagnostics Customer Service Call 800 266 2222 or go to www HardyDiagnostics com Alternatively you can contact your preferred distributor Catalog Number Description AFB 0070 AFB PRO Stain Kit AFB 0020 AFB PRO Replacement Tubing Kit AFB 0800 AFB QC Slides Box of 50 PF72P ProSlide Frosted...

Отзывы: