Product Information

 

The Hardy IncuPRO Incubators are designed for laboratories and clinics that 
require contemporary styling, accuracy, and economy in a basic incubator 
package.  The incubator uses a dial controlled thermostat and has the added 
features of a digital temperature display and a double walled chamber door.  The 
incubator has a chamber volume of 0.8 cu. ft. (0.02 cu.m) and a chamber 
temperature range of a5

˚

C to 60˚C.  The incubator has a chamber 

uniformity of +/- 1

˚

C at 37

˚

C within the bounds of the shelf area. 

 

3.0

    

Assembly

 

3.1   

Unpacking

 

Remove packing materials carefully, and retain for future shipment or 
storage of the unit.  Inspect for damage.  Report all shipping damage to 
the carrier immediately.  Shipping damage is covered by the carrier and 
repair/replacement for shipping damages must be coordinated through the 
carrier. Complete and return the Warranty Registration Card.  Packs 
should contain: 

  133000H Incubator 

  Power Line Cord 

  Operating Instructions 

  Two Shelves 

  Eight Shelf Brackets 

3.2 

Installation

 

Place the Incubator on a flat and stable surface, preferably away from 
drafts.  Ensure that the surface on which the unit is placed will withstand 
the radiated heat produced by typical laboratory incubators.  Fit the power 
line cord into the IEC power socket on the rear of the unit.  Plug power 
cord into a power supply that matches the voltage listed on the serial 
number label on the rear of the unit. 
 
Insert the wire rack shelves at the desired heights.  Refer to the sketch 
below for proper shelf bracket installation. 
 
 
 
 
 
 
 
 
 
 
 

Содержание IncuPRO 133000H

Страница 1: ...IFU XXXXX IncuPRO Incubator Model 133000H Operating Instructions...

Страница 2: ...packing 3 3 2 Installation 3 4 0 Operation 4 4 1 Controls and Indicator Lamps 4 4 2 Setting the Temperature 4 4 3 Operating the Incubator 5 5 0 Accessories 5 6 0 Fault Diagnosis 5 7 0 Technical Specif...

Страница 3: ...supply that provides a safety ground terminal Before moving disconnect at the power supply socket Do not remove the plug from the rear of the unit Do not check the temperature by touch but instead use...

Страница 4: ...age Report all shipping damage to the carrier immediately Shipping damage is covered by the carrier and repair replacement for shipping damages must be coordinated through the carrier Complete and ret...

Страница 5: ...erature adjust knob is used to control the chamber temperature Temperature Adjust Knob Power Switch O I Heater Lamp Temperature Display 4 2 Setting the Temperature Use the temperature adjust knob to s...

Страница 6: ...le Shelf Catalog Number C1902053 5 2 Shelf Bracket Catalog Number A1903208 6 0 Fault Diagnosis Symptom Possible Cause Action Required 1 Unit does not operate a Unit not switched on b Unit not plugged...

Страница 7: ...aning If a spill occurs use appropriate clean up procedures as required for radiation or biohazard control The outer casing may be cleaned with water and a damp cloth Do not submerge or immerse the In...

Страница 8: ...harge for parts or labor 10 0 Service It is required to obtain a Returned Material Authorization RMA number before any Hardy Diagnostics products are returned for any reason A Decontamination Certific...

Отзывы: