- 5 - 

 

calibration is done. 

After zero calibration, one yellow LED will be lighted on which 

means the detector enters into the calibration status of the second point. 

Put the detector into the standard gas (level equals to the second point) 

and keep it steady for about 30 seconds. Then press 

  to save the 

second point information. “

Power” light will be lighted on in yellow 

color, 3 green LED and 3 yellow LED are lighted on and then flashing, 

which means the calibration of second point is done. If there are more 

than two calibration points, the operating method is same with the second 

point calibration. After all points’ calibration finish, “

Power” light 

will be lighted in green color and then turns off. The detector will turns 

off then. 

Note:   

There  are  more  than  two  calibration  points.  Please  carry  out 

calibration in order, first zero point calibration and then the second point, 

the third point. 

During  calibration,  if  the  sensor  responses  abnormally,  the 

Fault” light and “

OL” light will be lighted on. Please note it.   

6. Charging 

Please  charge  the  battery  after  the  detector  is  turned  off.  The  color  of 

Power” light on the charger will show the voltage level of the battery. 

The  charging  will  last  for  4  to  7  hours. When  “

Power”  light  doesn’t 

flash  and  it  is  in  green  color,  the  charging  is  done.  Please  disconnect  the 

charger from the power source.   

Charging Attention 

 

It’s forbidden to charge the detector in dangerous area. Otherwise, it 

will damage the detector or cause fire & explosion.   

 

During charging, it’s normal that the area around the battery will be 

heated. 

Содержание JL269LED

Страница 1: ...JL269LED Portable gas leak detector Operation manual Please read this manual carefully before operation Ver HWWM161229CG...

Страница 2: ......

Страница 3: ...g range gas for long time Otherwise it will seriously affect the performance or even damage the device If exposed to the environment consisting of leaded compound sulfur compound organic phosphorus co...

Страница 4: ...s if they have been used for 2 years or charged for 300 times It s forbidden to disassemble adjust or repair the device without permission Please avoid the device falling down from high place or serio...

Страница 5: ...NCTIONS 2 4 OPERATION 2 4 1 Power on 2 4 2 Impact Test 3 4 3 Detecting 3 4 4 Battery voltage check 3 4 5 Open close audible alarm 4 4 6 Set present level as reference basis 4 4 7 Turn off the detector...

Страница 6: ...Automatic zero point adjustment 2 Main technical Specifications Target gas Sensor type Detecting range CH4 semiconductor 0 30000ppm 0 60 LEL 0 3 0 VOL C3H8 semiconductor 0 20000ppm 0 95 LEL 0 2 0 VOL...

Страница 7: ...1 Power on Press and hold for more than 2 seconds the detector will power on and enters into self testing status All LED lights will be lighted on and then off alternately At the same time the buzzer...

Страница 8: ...ble test bar into the gas environment The quantity of LED lights and sound frequency will show the gas concentration With gas concentration increasing first more and more green LED lights will be turn...

Страница 9: ...button 4 7 Turn off the detector Press both and buttons the detector will be turned off 5 Calibration In order to assure the accuracy it s suggested to re calibrate the detector once every 6 months Ca...

Страница 10: ...lor and then turns off The detector will turns off then Note There are more than two calibration points Please carry out calibration in order first zero point calibration and then the second point the...

Страница 11: ...se don t charge them 7 Sensor using and replacement In normal working status the sensor life is 2 years When the sensor is overdue or defective please contact the seller for correct replacement instru...

Страница 12: ......

Отзывы: