background image

8

A100h 
JANUARY 2020

State

Switch 1

(Tap)

Switch 1

(Hold)

Switch 2

(Tap)

Switch 2

(Hold)

Warm-up

-

-

Normal

Start Zero 

Calibration

Start Scan Calibration

Low Alarm

Mute Buzzer

Ack. Latched Alarm

High Alarm

Mute Buzzer

Ack. Latched Alarm

Offline

-

-

Fault

Mute Buzzer

Ack. Latched Alarm

Negative Gas Fault

Mute Buzzer

Start San Calibration

Zero Cal. Fault

Acknowledge Fault

-

Span Cal. Fault

-

Acknowledge Fault

Enable

Bluetooth

®

Connectivity

Enable

Bluetooth

®

Connectivity

   4. Once alarm thresholds have been exceeded,  

       relays should activate and digital signal outputs  

       should transmit the gas concentration.
   5. Turn off gas flow from cylinder and remove   

       calibration adapter from detector.
   6. Allow detector to recover/stabilize before  

       returning to normal operation (green LED lit).

VERIFYING ANALOG SIGNAL

The Hansen Gas Detector features a single 

configurable analog output. During normal operation 

the analog output signal of the detector is proportional 

to the detected gas concentration.

SWITCH FUNCTIONS

 
User interaction with the gas detector is accomplished 

through the use of two magnetic switches located on the 

bottom of each unit. To actuate a magnetic switch (referred 

to as MAG#1 or MAG#2), apply the supplied magnetic 

wand) to the relevant switch location as indicated below:

GENERAL CALIBRATION PROCEDURE

Ambient air can be used to zero the detector instead of 

synthetic air only if the area is known to be free of the 

target gas to which any detector may be cross-sensitive. 

In this case, no cylinder or calibration adapter is needed 

for zero adjustment. Otherwise follow the steps below 

 

NOTE:

 The gas detector MAY NOT be in an alarm or fault 

condition during calibration. Acknowledge any alarms or 

faults BEFORE attempting to begin the calibration process. 

Except for CO2, calibration gas must be in a balance of air, 

NOT nitrogen (Ni).

  1. Fit calibration adapter to the sensing element on  

       the gas detector lid.
   2. If using a variable flow regulator, adjust the gas    

       flow to approximately 0.3 L/min.
   3. Hold magnet at MAG#l for >5 seconds
   4. Apply zero gas
   5. Tap magnet at MAG#l within 30 seconds or gas    

       detector will time-out 
   6. The LED will blink green-red, green-red-red, etc.  

       until calibration is complete.
   7. Turn off gas flow from zero gas  

 

Zero calibration can be aborted during step 6 by 

holding the magnet at MAG#l for >5 seconds. If 

calibration is unsuccessful orange LED blinks, in this 

case tap MAG#l with magnet to discard calibration 

attempt and start over. 

After calibrating the zero adjustment, the span 

adjustment should be calibrated. With calibration 

adapter still connected and flow regulator still 

adjusted to approximately 0.3 L/min: 

 

   1. Hold magnet at MAG#2 for >5 seconds. The LED  

       will blink green-green-orange when ready
   2. Apply target gas at appropriate concentration  

       level
   3. Tap magnet at MAG#2 within 30 seconds or  

       sensor will time-out
   4. The LED will blink green-orange, green-orange- 

       orange, etc. until calibration is complete.
   5. Turn off gas flow from calibration gas 

 

Span calibration can be aborted during step 4 by 

holding the magnet at MAG#1 for >5 seconds. If 

calibration is unsuccessful LED blinks orange, in this 

case tap magnet at MAG#2 to discard calibration 

attempt. 

After the calibration process is complete, allow the 

sensor to recover and stabilize before the sensor 

returns to normal operation. The green LED should be 

lit for normal operation. 

FIG. 10

START-UP SEQUENCE 

After applying power, the instrument will go through a start-up sequence 

{initialization, audible/visual test and 

self-test sequence). 

After start-up sequence completes, the instrument will enter a warm-up period to allow the 

sensor element to stabilize before reporting a valid output. 

1. Switch power on.

2. Observe the start-up sequence and warm-up phase:

• Green LED will blink at 0.5 HZ for about 5 minutes.

• Modbus flag for warm-up is set.

Buzzer is off.

• Relay state is "no alarm."

• Gas reading is invalid.

3. Observe normal operation:

• Green LED is steady on.

Modbus flag for warm-up is cleared.

Buzzer is off.

• Relay state is "no alarm."

• Gas reading is valid.

The sensor may also enter several special states, these are indicated below by the specific analog output levels 

indicated: 

MODE OF OPERATION 

4-20mA

1-SV

0-SV

2-l0V

0-l0V

Instrument Fault 

< 1.2 mA 

< 0.3 V 

N/A 

<0.6V 

N/A 

Offline Mode/ Maintenance 

3 mA 

0.75 V 

N/A 

1.5 V 

N/A 

Drift below Zero 

3.8mA 

0.95V 

N/A 

1.9 V 

N/A 

Measuring Range Exceeded 

20.5 mA 

5.12 V 

5.12 V  10.25 V 

10.25 V 

Fault on Analog Interface 

>21 mA

> 5.25 V

> 5.25 V

> 10.5 V

> 10.5 V

VERIFYING ANALOG SIGNAL 

The Hansen Gas Detector features a single configurable analog output. During normal operation the analog output 

signal of the sensor is proportional to the detected gas concentration. 

GAS CONCENTRATION 

4-20mA

1-SV

0-SV

2-l0V

0-l0V

0% 

4mA 

lV 

ov 

2V 

ov 

50% 

12 mA 

3V 

2.5 V 

6V 

5V 

100% 

20 mA 

5V 

5V 

l0V 

lOV 

TABLE 6

FIG. 11

8b.  Zero Adjustment

8a.  General Calibration Procedure

8c.  Span Adjustment

9.  Bump Test

5.  Installation 

(Continued)

6.  Connect Hansen Gas Detector to App 

(User Discretion)

7.  Operation Overview

STEP 3 | 

Reinstall Sensor & Connect Lid

WARNING:

 DO NOT leave excess cable inside of the gas detector 

housing. Failure to comply may result in damage to the product.

IMPORTANT:

 To achieve proper seal in the IP66 version, the lid 

screws should be torqued to 15 to 20 lb

f

 in (1.5 to 2.0 Nm).

1.  Reinstall the rubber gasket 

(IP66 version only)

. Ensure that it 

is correctly seated by placing the side with two grooves face 

down and the edge with two bumps on the top.

WARNING:

 Except for CO

2

 or O

2

 sensors, ambient air may be used 

instead of zero gas if the area is know to be free of the target gas 

or any gases to which the sensor may be cross-sensitive.

3.

  Begin zero adjustment:

 

APP:

 

Home Tab

 

 

Calibrate

 

 scan barcode on gas 

cylinder or manually enter values for zero gas.

 

MAG SWITCH:

 hold 

MAG#1

 for >5-seconds. The LED 

will  blink  green-green-red  when  the  instrument  is 

ready.

4.

  Apply zero gas 

(or ambient air per warning above)

.

5.

  Confirm the start of calibration:

 

APP:

 press the 

Start Zero 

button.

  MAG SWITCH: 

tap 

MAG#1

 within 30-seconds or the 

instrument will time-out and return to normal 

operation.

6.

  Complete zero adjustment:

 

APP:

 app will countdown to completion. If calibration is 

successful, proceed to Step 12.

 

MAG SWITCH:

  the  LED  will  blink  green-red,  green-

red-red, green-red-red-red, etc. until calibration is 

complete. To abort, hold 

MAG#1

 for >5-seconds, turn 

off gas flow and remove the calibration adapter. If 

calibration is successful 

(green LED)

, proceed to Step 12. 

If calibration is unsuccessful 

(LED blinks orange @ 2 Hz)

tap 

MAG#1

 to discard the calibration attempt and see 

the user manual for troubleshooting.

7.

  Turn off gas flow from zero gas.

8

.  Replace  zero  gas  with  calibration  gas  in  preparation  for 

span adjustment.

9.

  Begin span adjustment:

 

APP:

 scan barcode on gas cylinder or manually enter 

values for calibration gas.

 

MAG SWITCH:

 hold 

MAG#2

 for >5-seconds. The LED will 

blink green-green-orange when the instrument is ready.

10.

  Apply calibration gas at the concentration listed on the cal 

gas concentration label 

(located on top of the instrument)

WARNING:

 The Hansen Gas Detector MAY NOT be in an 

alarm or fault condition during calibration. Acknowledge 

any alarms or faults BEFORE attempting to begin the 

calibration process.

WARNING:

 Except for CO

2

 or O

2

 sensors, calibration gas must be 

in a balance of air, not nitrogen (N

2

).

IMPORTANT:

 Calibration and / or bump testing requires the 

calibration adapter kit (P/N 31-1037).

IMPORTANT:

  At  elevations  higher  than  6,560′  (2,000  m), 

calibration will result in a lower reading. See the user manual 

for additional information.

1.  Fit calibration adapter to the gas detector lid.

IMPORTANT:

 Default alias, passkey and unlock code can be 

changed via the app’s configuration menu.

1.  Enable Bluetooth

®

 discovery by tapping 

MAG #1

 for 1-second. 

(After 10-seconds, device will indicate that it is discoverable with 

audible heartbeat until it has been paired, discovery has timed-out 

or has been cancelled.)

2.  Launch the app and click the Bluetooth

®

 icon at the bottom of 

the screen to initiate a scan.

3.  Select the instrument 

(default is “18TMAH”)

 from the list of 

available Bacharach gas detectors.

4.  When prompted, enter the passkey 

(default is “123456”)

.

5.  Go to configure tab to setup device. When prompted, enter 

unlock code to access device configuration. 

(default is “1234”)

IMPORTANT:

 The manufacturer of this product requires that a 

bump test or calibration be performed following installation to 

verify instrument functionality.

1.

  Connect adapter and gas cylinder according to the 

instructions in the General Calibration Procedure.

2.

  If desired, disable / silence external annunciators 

(e.g., 

shutdown valves, notification of authorities, etc.)

:

 

APP:

 

Home Tab

 

 

Calibrate

 

 

Bump

 

 toggle 

Take Offline

 

to disable communications to external devices.

 

MAG SWITCH:

 Inform building personnel of test so that 

external devices can be disabled / silenced.

3.

  Apply a sufficiently high concentration of the target gas to 

trigger alarms, but NOT pure refrigerant or hydrocarbons 

(e.g., do not use a butane lighter)

.

4.

  Once thresholds have been exceeded, relays should 

activate, digital outputs should transmit the gas 

concentration and:

 

APP:

 gas concentration should be displayed, the 

instrument status should be “Low Alarm” or “High Alarm” 

and alarms states should be “On.”

 

MAG SWITCH:

 LED status should display “Low Alarm” or 

“High Alarm.”

5.

  Turn off gas flow and remove the calibration adapter.

6.

  Allow sensor to recover / stabilize before the instrument 

returns to normal operation 

(green LED)

.

2.  Reconnect the ribbon cable from the sensor to the PCBA.
3.  Ensure no cables are interferring with the sensor module and 

close the lid.

4.  Using a 5/32″ 

(4 mm)

 hex key / allen wrench, tighten the lid 

screws in an “X” tightening pattern:

2.  If  using  a  variable  flow  regulator,  adjust  the  gas  flow  to 

approximately 0.3 L/min.

The Hansen Gas Detector uses a smartphone application to 

allow users to interface with the gas detector. To download the 

app, scan here or visit 

www.hantech.com/gas-detection

.

#

APP DESCRIPTION

1

Main Menu (App Settings)

2

Status (Gas Concentration)

3

Calibrate (Calibration / Bump Test)

4

Details (Instrument Information)

5

Disconnect Bluetooth

®

6

Restart Connected Device

7

Test Mode (LED / Buzzer / Relays / 

Analog Output)

8

Device Configuration

9

Logs

11.

  Confirm the start of calibration:

 

APP:

 press the 

Start Span

 button.

  MAG SWITCH: 

tap 

MAG#2

 within 30-seconds or the 

instrument will time-out and return to normal operation.

12.

  Complete span adjustment:

 

APP:

 app will countdown to completion. If calibration is 

successful, proceed to Step 18.

  MAG SWITCH: 

the  LED  will  blink  green-orange,  green-

orange-orange, green-orange-orange-orange, etc. 

until calibration is complete. To abort, hold 

MAG#2

 for 

>5-seconds, turn off gas flow and remove the calibration 

adapter. If calibration is successful 

(LED blinks green-orange-

red)

, proceed to Step 18. If calibration is unsuccessful 

(LED 

blinks orange @ 2 Hz)

, tap 

MAG#2

 to discard the calibration 

attempt and see the user manual for troubleshooting.

13.

  Turn  off  gas  flow  from  calibration  gas  and  remove  the 

calibration adapter.

14.

  Allow sensor to recover / stabilize before the instrument 

returns to normal operation 

(green LED)

.

Part Number

Serial Number

Sensor Type

Max Range

PN:   1100-XXXX

SN:   YYWW####

GAS:   CO2

 

10000 PPM

STATE

OUTPUT

INPUT

LED

RELAY 1

RELAY 2

RELAY 3

Buzzer

MAG #1 TAP

MAG #1 HOLD

MAG #2 TAP

MAG #2 HOLD

Warm-up

OFF

OFF

OFF

Enable Bluetooth

®

 

Connectivity

Disable Bluetooth

®

 

Connectivity

Normal

OFF

OFF

OFF

Start Zero Cal.

Start Span Cal.

Low Alarm

ON

OFF

OFF

Mute Buzzer

Ack. Latched Alarm

High Alarm

ON

ON

OFF

Mute Buzzer

Ack. Latched 

Alarm

Offline

OFF

OFF

OFF

Fault

OFF

OFF

ON

Mute Buzzer

Ack. Latched Fault

Negative Gas Fault

OFF

OFF

ON

Mute Buzzer

Start Zero Cal.

Zero Cal. Fault

OFF

OFF

OFF

Ack. Fault

Span Cal. Fault

OFF

OFF

OFF

Ack. Fault

TABLE 7

Содержание HGD-EC-NH3-100

Страница 1: ...sent in the surrounding environment The analog outputs 4 20mA or 0 10V DC can interface with nearly any existing monitor computer or PLC controller The sensor elements are mounted internally on the en...

Страница 2: ...120 F 50 C 4 F 20 C to 122 F 50 C 13 F 25 C to 131 F 55 C VOLTAGE REQUIREMENTS 24 VDC 18 24 VAC 20 100 240 VAC 50 60 Hz 80W max D Cell BaVery HUMIDITY RANGE NON CONDENSING 5 90 RH 15 90 RH electroche...

Страница 3: ...e most likely sources of leakage Perimeter detection is where detectors completely surround the area in question The size and nature of the area will help to decide which method is the most appropriat...

Страница 4: ...D AVOID FALSE ALARMS REDUCING BUSHING SCREEN STAND ALONE CONTROL The Hansen Gas Detector can operate as a stand alone device All that is needed is a 24V AC DC power supply LOCAL AND REMOTE VISUAL AUDI...

Страница 5: ...nly 3 Internal Alarm Buzze 4 Power Connections x2 5 Digital Connection MODBUS 6 Analog Connection 7 Tactile Switch 1 8 Ribbon Cable to Sensor 9 Tactile Switch 2 10 Relay 3 Connection FAULT 11 Relay 2...

Страница 6: ...f the detector measures the gas dependent light transmission while the other channel is used as a reference The ratio between measurement and reference signal is used to determine the gas concentratio...

Страница 7: ...el or branching multiple units from the main bus may introduce impedance mismatches reflections and or signal distortions Avoid long stubs when connecting detectors to the bus should be less than 3 De...

Страница 8: ...is proportional to the detected gas concentration GAS CONCENTRATION 4 20mA 1 SV 0 SV 2 l0V 0 l0V 0 4mA lV ov 2V ov 50 12 mA 3V 2 5V 6V 5V 100 20 mA 5V 5V l0V lOV TABLE 6 FIG 11 8b Zero Adjustment 9 Bu...

Страница 9: ...n user would need to download the Hansen Gas Detector App from either Google Play Store or Apple Store SENSOR SETUP In order to utilize the Bluetooth option sensors will need to be setup individually...

Страница 10: ...verify the instrument functionality The steps to perform a bump test with the Hansen Gas Detector APP 1 Connect adapter and gas cylinder according to the instructions in the General Calibration Proce...

Страница 11: ...ogrammable Logic Controller PLC Data logging is available on the HGD C 8 via the integrated SD card which can be removed to allow download of the logged data to a computer External Component Descrip2o...

Страница 12: ...scrip on 1 Integrated Visual Alarm 2 Controller Power and Fault LEDs 3 Channel 1 and 2 alarm mute bu ons 4 Power Fault Low High Alarm LEDs 5 M16 Cable Glands x6 6 M20 Cable Glands x2 Component Descrip...

Страница 13: ...UARY 2020 INSTALL DIMENSIONS 8 CHANNEL CONTROLLER FIG 18 5 625 0 0 0 6 0 O 0 ____ __ 16 J 1 47 REFERENCE NOT TO SCALE REFERANCE NOT TO SCALE 47 0 13 2 8 0 16 5 625 6 INSTALL DIMENSIONS 2 CHANNEL CONTR...

Страница 14: ...14 A100h JANUARY 2020 FIG 20 CONTROLLER WIRING CONNECTIONS...

Страница 15: ...he currently selected menu item is indicated with triangle pointers to the left and right of the description If the menu list is longer than can be displayed up and down arrow keys on the right side o...

Страница 16: ...nec ng you must ENABLE Bluetooth from this menu screen From the LCD contrast seqng screen use the arrow keys to adjust the contrast from 1 63 with 30 being the default The brightness of front panel LE...

Страница 17: ...MS device non Amer pted to e to edit ord a 30 s amer enter n 0 seqngs ma on nt since re rd bring assigned accessed either by selec ng CHANNEL CONFIG from the main menu or by pressing the channel numbe...

Страница 18: ...ttings of the Hansen gas detector must match the BMS system to work correctly The 3 form C relays that are included in the Hansen gas detector fault low alarm high alarm can be configured to be fail s...

Страница 19: ...D display open the clear cover and remove the LED display panel by removing the 4 screws On the backside of the LED display adjust the poten ometer labeled Z un l the KEY FEATURES GAS ALERT SYSTEM Pro...

Страница 20: ...N100 etc WIRING SCHEMATIC FOR CONTROLLER AND DETECTOR External Power Supply EARTH GRD CUSTOMER BMS 0 J w w I i i z i 0 0 n r7 r7 r7 I i t i t ppm L J L J L J I 0 w I i i z 0 0 n 0 0 i 0 0 0 z 5 5 w w...

Страница 21: ...connected to relay _______________________________ OK Check Analog Output in Use e g 0 5V 4 20 mA _______________________________ OK 2 On Site Gas Calibration Initial the following encircle OK when co...

Страница 22: ...EP1 Handheld unit without sensor module HGD C 2 31 1012 Electrochemical 0 50 500 PPM HGC C 8 8 Channel controller 115 230 VAC 31 1013 Electrochemical 0 500 2000 PPM 31 1032 Filter 10 pack CALIBRATION...

Отзывы: