18
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance
notice.
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
HI 764
09
/4
Galvanic DO probe with built-in temperature sensor and
4 m cable
HI 764
09
/10
Galvanic DO probe with built-in temperature sensor and
10 m cable
HI 76410-0
Protective sleeve for
HI 76410
probes series
HI 76410A/P
Membrane for
HI 76410
probes series, 5 pcs.
HI 7042S
Electrolyte solution, 30 ml bottle
HI 710015
Shockproof rubber boot, blue color
HI 710016
Shockproof rubber boot, orange color
HI 731326
Calibration screwdriver (20 pcs)
PROBE MAINTENANCE & CLEANING
PROBE MAINTENANCE & CLEANING
PROBE MAINTENANCE & CLEANING
PROBE MAINTENANCE & CLEANING
PROBE MAINTENANCE & CLEANING
For a top performance probe, it is recommended to replace the membrane
every 2 months and the electrolyte once a month.
Proceed as follows:
•
Unscrew the membrane by turning it counterclockwise.
•
Rinse the supplied spare membrane (
HI 764
09
A
/P
) with some
electrolyte solution while shaking it gently. Refill with clean
electrolyte.
•
Gently tap the membrane over a surface to ensure that no air
bubbles remain trapped.
•
With the sensor facing down screw the cap clockwise to the end
of the threads. Some electrolyte will overflow.
If any deposit scales the sensors, gently brush the sensor surface with the
supplied scouring pad, while paying attention to not damage the plastic
body.