 Overtime dosing

When enabled this feature ensures that overdosage is avoided.
Select a desired maximum dosage period. If the dosage relay is
active more than the selected period an alarm condition is activated
and the dosage relay is deactivated.
To leave the overtime dosage alarm status change the position of the

OFF/AUTO/ON

 switch to OFF and then to AUTO again. The overtime

counter is restarted.
To set the overtime dosing period rotate the overtime knob to the
proper position. The time can be set between approx. 5 min to 60 min
(from MIN to MAX position).
To disable the overtime dosage feature
make a short with a jumper wire between
the terminals of the proper connectors. (see
picture, terminals 1, 2)

 Hysteresis knob

Put the hysteresis knob in the desired position.
The dosage will be active according to the DO
reading, set point value and the hysteresis set
value.

DOSING MODE SELECTION

The DOSAGE switch and the corresponding right side LED on the front
panel are used to select and indicate the dosing mode.

OFF

 mode

The dosage is disabled. The corresponding
DOSAGE MODE LED blinks (right side of the
switch).

Connect the “+” wire from the recorder to terminal 1 on the
instrument and the other wire (common) to terminal 2 for 4-20 mA
recorder output or terminal 3 for 0-20 mA recorder output.

Note:

Only one recorder output connection is permited. In order to avoid
malfunction leave unconnected the unused terminal.

Содержание HI8410

Страница 1: ...10 Dissolved Oxygen Dissolved Oxygen Dissolved Oxygen Dissolved Oxygen Dissolved Oxygen P P P P Process Controller rocess Controller rocess Controller rocess Controller rocess Controller w w w h a n n...

Страница 2: ...applications If you need additional technical information do not hesitate to e mail us at tech hannainst com TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS P...

Страница 3: ...sated for the temperature effects on the oxygen solubility and membrane permeability Moreover thesalinity compensationfeatureallowsthedeterminationofDissolvedOxygeneven in salty waters The instruments...

Страница 4: ...e working dosing point SALINITY g L Todisplayandsetthesalinityfactor activeonly in mg L range CAL To enter in calibration mode 100 TEST Diagnosticfunction activeonlyin DOrange C To display the tempera...

Страница 5: ...ndicate that the dosage is active ALARM To indicate an alarm condition DOSAGEMODESWITCH To indicate that the continuous ON or OFF mode is selected from dosing switch SWITCHES OFF AUTO ON To select the...

Страница 6: ...8410 1 DO probe connection terminals 2 Range selection mg L or DO 3 SET terminals for connection to a dosing pump 4 ALARM terminals for connection to an external alarm device 5 Power supply terminals...

Страница 7: ...tout size for the installation Side view of the panel mounted unit Adjustablelocationbrackets suppliedwiththemeter allowtheindicator to slide into the cutout and will hold the unit securely in place 1...

Страница 8: ...rane see figure 3 Rinse the supplied membrane HI 76409A P with electrolyte while shaking it gently Refill with clean electrolyte Gently tap the membrane over a surface to ensure that no air bubbles re...

Страница 9: ...d a o l e v i t s i s e r V 0 4 2 x a M A 2 d e t a l o s i 1 s e k o r t s 0 0 0 0 0 0 1 e g n a R t n i o p t e S O 0 0 6 o t 5 2 O m p p L g m 0 0 5 o t 5 0 2 e g n a R m r a l A O m p p L g m 0 5...

Страница 10: ...orothercontrolequipment may be connected to the SET Max 2A 240 V terminals 3 4 These contacts act only as a dry switch allowing electrical continuity not as a power supply Alarm Contacts terminals 5 6...

Страница 11: ...hort with a jumper wire between the terminals of the proper connectors see picture terminals 1 2 Hysteresis knob Putthehysteresisknobinthedesiredposition The dosage will be active according to the DO...

Страница 12: ...Ensure the DO probe is immersed in the sample to be tested The DO reading can be displayed in air saturation or in ppm mg L according to the range selection The corresponding LED is ON For accurate di...

Страница 13: ...set point set point hysteresis overpassed stop dosing set point start dosing SALINITY g L SET mg L range only Oxygenmeasurementscanbecompensatedforsalinityconcentrationand this correction value can be...

Страница 14: ...he ALARM LED lights up PresstheSALINITYg Landthedisplaywillshowthesalinity Useasmall screwdrivertoadjusttheSALINITYtrimmertodisplaythedesiredsalinity value within the 0 to 51 g L range ALTITUDE COMPEN...

Страница 15: ...notwithinthetoleranceofthesetvalue the alarm contact will be open HYSTERESIS SET Turn the hysteresis knob rear panel in the desired position from 0 5 to 2 4 mg L range Thedosagewillbeactiveaccordingto...

Страница 16: ...turn the CAL trimmer to display 100 If the instrument is set on the mg L range press and hold down the CAL key to enter DO rangeandturntheCAL trimmertodisplay100 After CAL key is released the instrume...

Страница 17: ...ve active In SET mode it is ON or blinking 75 ON in accordance with the dosage relay status inactive active ALARM LED InMEASUREmodeitisOFForblinking 25 ON in accordance with the alarm status inactive...

Страница 18: ...tion screwdriver 20 pcs PROBE MAINTENANCE CLEANING PROBE MAINTENANCE CLEANING PROBE MAINTENANCE CLEANING PROBE MAINTENANCE CLEANING PROBE MAINTENANCE CLEANING Foratopperformanceprobe itisrecommendedto...

Страница 19: ...All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Recommendations for Users Before using these products make sure that they are...

Страница 20: ...et RI 02895 USA Local Sales and Customer Service office Hanna Instruments United States Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Tel 800 426 6287 Fax 401 765 7575 www h...

Отзывы: