First time you use the instrument open the battery compartment
and put the batteries inside observing the polarity.

To connect the probe

With the meter turned off, connect the 

HI 1288 

probe to the DIN

socket on the bottom of the meter by aligning the pins and
pushing in the plug. Tighten the nut to ensure a good connection.
Remove the protective cap from the probe before taking any
measurements.

To turn the meter ON and check the battery status

Press the 

ON/OFF/MODE

 button until the display lights up. At

start-up, all the LCD segments are displayed for 1 second, then the
percent indication of the remaining battery life is displayed for
another second (E.g. “100 % BATT“). The meter then enters the
normal measuring mode.

Note

:

If the display needs to be checked, keep the 

ON

 button

pressed while turning the meter on. The meter will display
all segments as long as the button is pressed.

To select the measurement range

While in normal measurement mode, press the 

SET/HOLD

 button

quickly to select pH, EC or TDS value on the primary LCD, while
temperature will be simultaneously displayed on the secondary
LCD (E.g. “pH 5.73“ and “22.5°C“).

To freeze the display

While in measurement mode, press and hold the 

SET/HOLD

button until “HOLD“ appears on the secondary display and the
reading will be frozen on the LCD (E.g. “pH 5.73 HOLD“). Press
any button to return to normal mode.

To turn the meter OFF

While in normal measurement mode, press the 

ON/OFF/MODE

button. “OFF“ will appear on the secondary display. Release the
button.

Note

: When the meter detects the absence of probe at its input,

the message “Probe not connected“ appears on LCD and ''-
--'' blinking on LCD lines. When a probe is connected, the
“Probe not connected“ tag is turned off, and the readings
are displayed on LCD lines.

While in measurement mode, press and hold the 

ON/OFF/MODE

button until “TEMP” and the current temperature unit are
displayed on the lower LCD (eg. “TEMP °C“). Then:

To select the temperature unit (°C / °F)

•  Press the 

SET/HOLD

 button to change the temperature unit,

and then press the 

ON/OFF/MODE

 button until the meter

returns to normal measurement mode.

To change the type of calibration buffer set,
• While in pH measurement mode, after setting the temperature

unit press the 

ON/OFF/MODE

 button again to show the

current buffer set: “pH 7.01 BUFF” (for standard buffer set:
4.01/7.01/10.01) or “pH 6.86 BUFF” (for NIST buffer set:
4.01/6.86/9.18). Change the set with the 

SET/HOLD

 button,

then press 

ON/OFF/MODE

 to return to normal mode.

To change the EC/TDS conversion factor (CONV) and the
temperature compensation coefficient (BETA)
• While in EC or TDS measurement mode after setting the

temperature unit press the 

ON/OFF/MODE

 button again to

show the current conversion factor (eg. “0.50 CONV“), then
press 

SET/HOLD

 to change the value.

• Press the 

ON/OFF/MODE

 button to show the current temperature

compensation coefficient b (eg. “2.1 BETA“), then press

SET/HOLD

 to change the temperature coefficient value.

• Press the 

ON/OFF/MODE

 button to return to normal opera-

tion.

The meter is supplied with batteries.
The meter displays the remaining battery percentage when turned
on. When the level is below 5%, the “

-    +

“ symbol on the LCD

blinks to indicate a low battery condition. If the battery level is low
enough to cause erroneous readings, the Battery Error Prevention
System (BEPS) turns the meter     off. It is recommended to replace
the batteries as soon as the display will flash the battery symbol.
To replace the batteries, follow the next steps:
• Open the battery compartment cap (on the bottom of the

instrument).

• Remove old batteries.
• Replace new batteries observing the polarity on the rear of

the instrument.

• Close the battery compartment cap.

• Make sure the meter has been calibrated before use.
• If the probe is dry, soak it in 

HI 70300

 storage solution for

30 minutes to reactivate it.

• Submerge the probe in the sample to be tested while stirring it

gently. Wait until the “Not Stable“ tag on the LCD is turned off.

• The pH value automatically compensated for temperature is

displayed on the primary LCD, while the secondary LCD displays
the sample temperature.

• If measurements are taken in different samples successively,

rinse the probe tip thoroughly to eliminate cross-contamination.
After cleaning, rinse the probe tip with some deionized water
and some of the sample to be measured.

pH calibration

• While in pH measurement mode, press and hold the

ON/OFF/MODE

 button until “OFF“ on the secondary display

is replaced by “CAL“. Release the button. Place the sensor
into the first calibration buffer.

• The meter enters the calibration mode, displaying “pH 7.01

USE” (or “pH 6.86 USE” if the NIST buffer set was selected).

• For a 

single-point calibration place the probe in any buffer from

the selected buffer set (E.g. pH 4.01, 9.18 or 10.01). The meter
will automatically recognize the buffer value.

• If the buffer is not recognized or the calibration offset is out of

the accepted range ''---WRONG''

 

is displayed.

• If the buffer is recognised ''REC" is displayed until the reading is

stable and the calibration is accepted.

• If using pH 7.01 (or pH 6.86 from NIST buffer set), after

acceptance of the buffer press any key to exit. ''OK1'' message is
displayed and meter returns to pH measurment mode.

• If using 4.01 or 10.01 (9.18) buffer the ''OK1'' message is

displayed and meter returns to pH measurment mode.

• For a 

two-point calibration, proceed as for the single point

calibration selecting 7.01/6.86 pH buffer.

• The ''pH 4.01 USE'' message is displayed.
• Place the probe in the second calibration buffer (pH 4.01 or

10.01, or, if using NIST, pH 4.01 or 9.18). When the second
buffer is accepted, the LCD will display “OK2” for 1 second and
the meter will return to the normal measurement mode.

• If the buffer is not recognised or the slope is out of accepted

range ''---WRONG'' is displayed. Change the buffer, clean the
electrode or press any key to exit calibration.

• When the calibration procedure is completed, the “Calibrated“

tag is turned on.

• It is always recommended to carry out a two-point calibration for

better accuracy.

OPERATIONAL GUIDE

pH MEASUREMENT & CALIBRATION

• Place the probe in the sample to be tested. Use plastic beakers

or containers to minimize any electromagnetic interference.

• Tap the probe lightly on the bottom of the container to remove

air bubbles that may be trapped inside the tip.

• Wait for a few minutes for the temperature sensor to reach

thermal equilibrium (i.e. until the “Not Stable“ tag disappears).

• The LCD displays the EC or TDS value (automatically compen-

sated for temperature) on the primary LCD, while the secondary
LCD displays the temperature of the solution.

EC calibration

• From EC or TDS normal mode, press and hold the 

ON/OFF/

MODE

 button until “OFF“ on the secondary display is replaced

by “CAL“. Release the button.

• The meter enters the calibration mode. Immerse the probe in

the 

HI 7031

 calibration solution (“

µ

S 1413 USE“) for

HI 991300

 or in 

HI 7030

 calibration solution (“mS 12.88

USE“) for 

HI 991301

.

• If the standard value is recognised ''REC'' is displayed until the

reading is stable and calibration is accepted.

• The LCD will display “OK” for 1 second and return to normal

measurement mode.

• If the standard is not recognised or the slope is out of accepted

range ''---Wrong'' is display. Change the calibration solution
or the electrode or press any key to exit calibration.

• When the calibration procedure is completed, the “Calibrated“

tag is turned on.

Notes

: • Beta should be set to 1.9 during calibration.

• Since there is a known relationship between the EC

and TDS reading, it is not necessary to calibrate the
meter in TDS. If the conversion factor is either 0.5 or
0.7, the meter will allow a direct calibration in TDS
by using the Hanna calibration solutions.

To exit calibration and reset default values

• After entering the calibration mode and before the point is

accepted, it is possible to quit the procedure and return to the
last calibration data by pressing the 

ON/OFF/MODE

 button.

The LCD displays “---ESC“ for 1 second and the meter returns to
normal mode.

• To reset the default values and clear a previous calibration,

press the 

SET/HOLD

 button after entering the calibration

mode, before the first point is accepted.
The LCD displays “---CLR“ for 1 second, the meter resets to the
default calibration and the “Calibrated” tag on the LCD disap-
pears.

METER SETUP

EC MEASUREMENT & CALIBRATION

BATTERY REPLACEMENT

 

Содержание HI 991300

Страница 1: ...w w w h a n n a i n s t c o m Instruction Manual HI 991300 HI 991301 Waterproof pH EC TDS Temperature Meters with Advanced Features ...

Страница 2: ...se precision two models are available with different conductivityranges forapplicationsfrompurifiedtobrackishwaters All operations and settings including calibration buffers and temperature scale selections are made through only 2 buttons The housing is waterproof and rated for IP 67 conditions The supplied multi parameter probe HI 1288 includes pH EC TDS and temperature in one convenient rugged p...

Страница 3: ...n 230 mL bottle HI 7030M 12 88 mS cm solution 230 mL bottle HI 7031M 1413 µS cm solution 230 mL bottle HI 7032M 1382 ppm solution 230 mL bottle HI 7038M 6 44 ppt solution 230 mL bottle HI 70442M 1500 ppm solution 230 mL bottle 1 Automatic Temperature Compensation indicator 2 Battery percentage visible at power up 3 Secondary display 4 Low battery indicator 5 Stability indicator 6 Calibration indic...

Страница 4: ...ONV 0 5 or 6 44 ppt CONV 0 5 or 1500 ppm CONV 0 7 or 9 02 ppt CONV 0 7 Temperature pH Automatic Compensation EC TDS Automaticwith βselectablefrom0 0to2 4 Cwith0 1increments TDS Conversion Factor Selectable from 0 45 to 1 00 with 0 01 increments default 0 50 Probe included HI 1288 pH EC TDS temperature with DIN connector and 1 m 3 3 cable Battery Type 3 x 1 5V AAA Battery Life Approximately 500 hou...

Страница 5: ...ressed while turning the meter on The meter will display all segments as long as the button is pressed To select the measurement range Whileinnormalmeasurementmode presstheSET HOLDbutton quickly to select pH EC or TDS value on the primary LCD while temperature will be simultaneously displayed on the secondary LCD E g pH 5 73 and 22 5 C To freeze the display While in measurement mode press and hold...

Страница 6: ... 10 01 The meter will automatically recognize the buffer value If the buffer is not recognized or the calibration offset is out of the accepted range WRONG is displayed Ifthebufferisrecognised REC isdisplayeduntilthereadingis stable and the calibration is accepted If using pH 7 01 or pH 6 86 from NIST buffer set after acceptance ofthebufferpressany key toexit OK1 message is displayed and meter ret...

Страница 7: ...d and return to normal measurement mode If the standard is not recognised or the slope is out of accepted range Wrong is display Change the calibration solution or the electrode or press any key to exit calibration When the calibration procedure is completed the Calibrated tag is turned on Notes Beta should be set to 1 9 during calibration Since there is a known relationship between the EC and TDS...

Страница 8: ...emperature unit press the ON OFF MODE button again to show the current conversion factor eg 0 50 CONV then press SET HOLD to change the value PresstheON OFF MODEbuttontoshowthecurrenttemperature compensation coefficient b eg 2 1 BETA then press SET HOLD to change the temperature coefficient value Press the ON OFF MODE button to return to normal opera tion The meter is supplied with batteries The m...

Страница 9: ... from whom you purchased the instrument If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with ...

Отзывы: