background image

w w w . h a n n a i n s t . c o m

HI 991300

HI 991301

Range

pH

0.00 to 14.00 pH

0.00 to 14.00 pH

EC

0 to 3999 

µ

S/cm

0.00 to 20.00 mS/cm

TDS 0 to 2000 ppm (mg/l)

0.00 to 10.00 ppt (g/l)

Temperature

0.0 to 60.0°C / 32.0 to 140.0°F

Resolution

pH

0.01 pH

0.01pH

EC

µ

S/cm

0.01 mS/cm

TDS

1 ppm (mg/l)

0.01 ppt (g/l)

Temperature

0.1°C / 0.1°F

Accuracy

±0.01 pH

(@20°C/68°F)

±2 % F.S. (EC / TDS)

±0.5°C / ±1°F

pH Calibration

Automatic, 1 or 2 point, choose between 2 sets
of buffers (pH 4.01/7.01/10.01 or
Nist pH 4.01/6.86/9.18)

EC/TDS Calibration

Automatic, 1 point at

1413 

µ

S/cm

12.88 mS/cm

or 1382 ppm (CONV=0.5) or 6.44 ppt (CONV=0.5)
or 1500 ppm (CONV=0.7) or 9.02 ppt (CONV=0.7)

Temperature

pH: Automatic

Compensation

EC/TDS: Automatic with

β

 selectable from 0.0 to 2.4 %/°C with     0.1 increments

TDS Conversion Factor

Selectable from 0.45 to 1.00 with

0.01 increments (default 0.50)

Probe (included)

HI 1288 

pH/EC/TDS/temperature

with DIN connector and 1 m (3.3’) cable

Battery Type

3 x 1.5V AAA

Battery Life

Approximately 500 hours of continuous use

Auto-off

After 8 minutes of non-use

Environment

0 to 50°C (32 to 122°F);

RH max. 100%

Dimensions

152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)

Weight

205 g (7.2 oz.)

Instruction Manual

HI 991300 • HI 991301

Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the
instrument. If you need additional technical information, do not
hesitate to e-mail us at 

[email protected]

.

All Hanna Instruments     

meters are warranted for two years

against defects in workmanship and materials when used for
their intended purpose and maintained according to instructions.

The probes are warranted for a period of six months.

This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of
prescribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the
problem. If the repair is not covered by the warranty, you will
be notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department
and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete
protection.

All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.

Remove the instrument from the packing material and examine
it carefully to make sure that no damage has occurred during
shipment. If noticeable damage is evident, notify your dealer.

Note:

Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing together with the
supplied accessories.

HI 991300 

and

 HI 991301 

have been designed to offer you

the combination of pH, conductivity, total dissolved solids and
temperature measurements.
To increase precision, two models are available, with different
conductivity ranges, for applications from purified to brackish waters.
All operations and settings, including calibration buffers and
temperature scale selections, are made through only 2 buttons.
The housing is waterproof and rated for IP 67 conditions.
The supplied multi-parameter probe, 

HI 1288

, includes pH, EC/TDS

and temperature in one convenient, rugged probe. In addition, to
ensure against interference from transient electrical noise to pH, a
solid-state preamplifier is integrated into the probe.
Other user selectable features include different TDS factors from
0.45 to 1.00, and a range of temperature coefficients (

β

) from 0.0

to 2.4% for greater consistency and reproducibility.

Each meter is supplied with:

HI 1288

 pH/EC/TDS probe with built-in temperature

sensor, DIN connector and 1 m (3.3’) cable

• Calibration and cleaning solution sachets
• 3 x 1.5V AAA batteries
• Instruction manual.

SPECIFICATIONS

Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance
notice.

HI  1288

pH/conductivity probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable

HI  7004M

pH 4.01 buffer solution, 230 mL bottle

HI  7006M

pH 6.86 buffer solution, 230 mL bottle

HI  7007M

pH 7.01 buffer solution, 230 mL bottle

HI  7009M

pH 9.18 buffer solution, 230 mL bottle

HI  7010P

pH 10.01 buffer solution, 230 mL bottle

HI  7030M

12.88 mS/cm solution, 230 mL bottle

HI  7031M

1413 

µ

S/cm solution, 230 mL bottle

HI  7032M

1382 ppm solution, 230 mL bottle

HI  7038M

6.44 ppt solution, 230 mL bottle

HI  70442M

1500 ppm solution, 230 mL bottle

1. Automatic Temperature Compensation indicator
2. Battery percentage (visible at power up)
3. Secondary display
4. Low battery indicator
5. Stability indicator
6. Calibration indicator
7. Selectable temperature unit
8. Primary display
9. Measuring unit for primary display

  

  IST991300Z  01/11

Waterproof

pH/EC/TDS/Temperature Meters

with Advanced Features

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

ACCESSORIES

LCD DESCRIPTION

WARRANTY

 

Содержание HI 991300

Страница 1: ...w w w h a n n a i n s t c o m Instruction Manual HI 991300 HI 991301 Waterproof pH EC TDS Temperature Meters with Advanced Features ...

Страница 2: ...se precision two models are available with different conductivityranges forapplicationsfrompurifiedtobrackishwaters All operations and settings including calibration buffers and temperature scale selections are made through only 2 buttons The housing is waterproof and rated for IP 67 conditions The supplied multi parameter probe HI 1288 includes pH EC TDS and temperature in one convenient rugged p...

Страница 3: ...n 230 mL bottle HI 7030M 12 88 mS cm solution 230 mL bottle HI 7031M 1413 µS cm solution 230 mL bottle HI 7032M 1382 ppm solution 230 mL bottle HI 7038M 6 44 ppt solution 230 mL bottle HI 70442M 1500 ppm solution 230 mL bottle 1 Automatic Temperature Compensation indicator 2 Battery percentage visible at power up 3 Secondary display 4 Low battery indicator 5 Stability indicator 6 Calibration indic...

Страница 4: ...ONV 0 5 or 6 44 ppt CONV 0 5 or 1500 ppm CONV 0 7 or 9 02 ppt CONV 0 7 Temperature pH Automatic Compensation EC TDS Automaticwith βselectablefrom0 0to2 4 Cwith0 1increments TDS Conversion Factor Selectable from 0 45 to 1 00 with 0 01 increments default 0 50 Probe included HI 1288 pH EC TDS temperature with DIN connector and 1 m 3 3 cable Battery Type 3 x 1 5V AAA Battery Life Approximately 500 hou...

Страница 5: ...ressed while turning the meter on The meter will display all segments as long as the button is pressed To select the measurement range Whileinnormalmeasurementmode presstheSET HOLDbutton quickly to select pH EC or TDS value on the primary LCD while temperature will be simultaneously displayed on the secondary LCD E g pH 5 73 and 22 5 C To freeze the display While in measurement mode press and hold...

Страница 6: ... 10 01 The meter will automatically recognize the buffer value If the buffer is not recognized or the calibration offset is out of the accepted range WRONG is displayed Ifthebufferisrecognised REC isdisplayeduntilthereadingis stable and the calibration is accepted If using pH 7 01 or pH 6 86 from NIST buffer set after acceptance ofthebufferpressany key toexit OK1 message is displayed and meter ret...

Страница 7: ...d and return to normal measurement mode If the standard is not recognised or the slope is out of accepted range Wrong is display Change the calibration solution or the electrode or press any key to exit calibration When the calibration procedure is completed the Calibrated tag is turned on Notes Beta should be set to 1 9 during calibration Since there is a known relationship between the EC and TDS...

Страница 8: ...emperature unit press the ON OFF MODE button again to show the current conversion factor eg 0 50 CONV then press SET HOLD to change the value PresstheON OFF MODEbuttontoshowthecurrenttemperature compensation coefficient b eg 2 1 BETA then press SET HOLD to change the temperature coefficient value Press the ON OFF MODE button to return to normal opera tion The meter is supplied with batteries The m...

Страница 9: ... from whom you purchased the instrument If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with ...

Отзывы: