w w w . h a n n a i n s t . c o m
HI 991300
HI 991301
Range
pH
0.00 to 14.00 pH
0.00 to 14.00 pH
EC
0 to 3999
µ
S/cm
0.00 to 20.00 mS/cm
TDS 0 to 2000 ppm (mg/l)
0.00 to 10.00 ppt (g/l)
Temperature
0.0 to 60.0°C / 32.0 to 140.0°F
Resolution
pH
0.01 pH
0.01pH
EC
1
µ
S/cm
0.01 mS/cm
TDS
1 ppm (mg/l)
0.01 ppt (g/l)
Temperature
0.1°C / 0.1°F
Accuracy
±0.01 pH
(@20°C/68°F)
±2 % F.S. (EC / TDS)
±0.5°C / ±1°F
pH Calibration
Automatic, 1 or 2 point, choose between 2 sets
of buffers (pH 4.01/7.01/10.01 or
Nist pH 4.01/6.86/9.18)
EC/TDS Calibration
Automatic, 1 point at
1413
µ
S/cm
12.88 mS/cm
or 1382 ppm (CONV=0.5) or 6.44 ppt (CONV=0.5)
or 1500 ppm (CONV=0.7) or 9.02 ppt (CONV=0.7)
Temperature
pH: Automatic
Compensation
EC/TDS: Automatic with
β
selectable from 0.0 to 2.4 %/°C with 0.1 increments
TDS Conversion Factor
Selectable from 0.45 to 1.00 with
0.01 increments (default 0.50)
Probe (included)
HI 1288
pH/EC/TDS/temperature
with DIN connector and 1 m (3.3’) cable
Battery Type
3 x 1.5V AAA
Battery Life
Approximately 500 hours of continuous use
Auto-off
After 8 minutes of non-use
Environment
0 to 50°C (32 to 122°F);
RH max. 100%
Dimensions
152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
Weight
205 g (7.2 oz.)
Instruction Manual
HI 991300 • HI 991301
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the
instrument. If you need additional technical information, do not
hesitate to e-mail us at
.
All Hanna Instruments
meters are warranted for two years
against defects in workmanship and materials when used for
their intended purpose and maintained according to instructions.
The probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of
prescribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the
problem. If the repair is not covered by the warranty, you will
be notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments, first obtain a Returned Goods
Authorization Number from the Customer Service department
and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete
protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Remove the instrument from the packing material and examine
it carefully to make sure that no damage has occurred during
shipment. If noticeable damage is evident, notify your dealer.
Note:
Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing together with the
supplied accessories.
HI 991300
and
HI 991301
have been designed to offer you
the combination of pH, conductivity, total dissolved solids and
temperature measurements.
To increase precision, two models are available, with different
conductivity ranges, for applications from purified to brackish waters.
All operations and settings, including calibration buffers and
temperature scale selections, are made through only 2 buttons.
The housing is waterproof and rated for IP 67 conditions.
The supplied multi-parameter probe,
HI 1288
, includes pH, EC/TDS
and temperature in one convenient, rugged probe. In addition, to
ensure against interference from transient electrical noise to pH, a
solid-state preamplifier is integrated into the probe.
Other user selectable features include different TDS factors from
0.45 to 1.00, and a range of temperature coefficients (
β
) from 0.0
to 2.4% for greater consistency and reproducibility.
Each meter is supplied with:
•
HI 1288
pH/EC/TDS probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable
• Calibration and cleaning solution sachets
• 3 x 1.5V AAA batteries
• Instruction manual.
SPECIFICATIONS
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance
notice.
HI 1288
pH/conductivity probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable
HI 7004M
pH 4.01 buffer solution, 230 mL bottle
HI 7006M
pH 6.86 buffer solution, 230 mL bottle
HI 7007M
pH 7.01 buffer solution, 230 mL bottle
HI 7009M
pH 9.18 buffer solution, 230 mL bottle
HI 7010P
pH 10.01 buffer solution, 230 mL bottle
HI 7030M
12.88 mS/cm solution, 230 mL bottle
HI 7031M
1413
µ
S/cm solution, 230 mL bottle
HI 7032M
1382 ppm solution, 230 mL bottle
HI 7038M
6.44 ppt solution, 230 mL bottle
HI 70442M
1500 ppm solution, 230 mL bottle
1. Automatic Temperature Compensation indicator
2. Battery percentage (visible at power up)
3. Secondary display
4. Low battery indicator
5. Stability indicator
6. Calibration indicator
7. Selectable temperature unit
8. Primary display
9. Measuring unit for primary display
IST991300Z 01/11
Waterproof
pH/EC/TDS/Temperature Meters
with Advanced Features
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
ACCESSORIES
LCD DESCRIPTION
WARRANTY