Setup/Installation

16

Delete Logging Type/Lot

Press 

RCL

 and select the parameter log.

Use  the 

pq

  keys  to  select  the  Manual/Stability 

records  or  Interval  lots  to  delete.  Press 

CLR

.  The 

instrument  will  display  “CLEAR  MANUAL”  for  Manual 
Records, “CLEAR STAB” for Stability Records. 

For Interval lots, the message “CLEAR”, followed by the 
selected lot will be displayed with “CFM” tag blinking.
Press the 

pq

 keys to select a diff erent lot. Press 

CFM

The instrument will display “PLEASE WAIT”.

“CLEAR DONE” is displayed for a few seconds after the 
selected Interval lot is deleted.

Delete Records (Manual and Stability log on demand)

To  delete  individual  records  (Manual  and  Stability  logs  only),  enter  Manual 
(Stability)  log  by  pressing 

CFM

  when  Manual  (Stability)  is  displayed.  Use  the 

ARROW keys to select the record to be deleted and then press 

CLR

.

The  instrument  will  display  “CLEAR  REC.“  and  record 
number along with “CFM” tag blinking. Use the ARROW 
keys to select another record if necessary.

Press 

CFM

.  The  instrument  will  display  “PLEASE  WAIT”  and  then  “CLEAR 

DONE”  message.  When  individual  logs  are  deleted  within  saved  MANUAL  or 
STABILITY logs, the logs will renumber, fi lling in the deleted data but staying in 
chronological order.
To delete all records of the MANUAL (STABILITY) log, proceed as described on 
page 16 for LOTS.

Select  the  Manual  (Stability)  lot  and  press 

CLR

The  ”CLEAR”  message  will  be  displayed  along  with 
”MANUAL”  or  “STABILITY”  and 

CFM

  tag  blinking  on 

the LCD. Press the 

CFM

 key to confi rm the deleting of 

the selected lot (MANUAL or STABILITY) or all records. 
Press 

CLR

 to exit without deleting. 

Viewing 

Logged 

Data

Operational Guide

25

Preparation

Pour a small quantity of the calibration solution into a beaker. If possible, use a 
plastic beaker to minimize any EMC interferences.
Before pressing 

CAL

 verify in SETUP:

• 

Basic mode is off .

• 

Salinity Scale is set to NaCl%.

In measurement mode use the RANGE key to select the Salinity measurement. 
The NaCl% tag will be on. NaCl calibration is a single point calibration at 100.0% 
NaCl. Use the HI 7037L calibration solution (sea water solution) as a 100% NaCl 
calibration solution.

Procedure

Rinse  the  probe  with  some  of  the  HI  7037L  calibration  solution  or  deionized 
water. Shake off  excess solution. Submerse the probe in the calibration solution. 
The sleeve holes must be completely submersed. If possible, center the probe in 
the beaker away from the bottom or beaker walls. Raise and lower the probe to 
refi ll the center cavity and tap the probe repeatedly to remove any air bubbles 
that may be trapped inside the sleeve.

Press 

CAL

  to  enter  calibration.  The  ” “,  ”STIR”  and 

“CAL”  tags  will  turn  on  and  the  fi rst  LCD  line  will 
display the percent NaCl reading, the second LCD line 
will display the temperature and the third LCD line will 
display  ”WAIT”  message  blinking  until  the  reading  is 
stable.

When the reading is stable and close to the standard, 
“CFM”  tag  will  blink  and      “SOLUTION  STANDARD” 
message will be displayed.

Press 

CFM

 to confi rm calibration.

The Instrument displays “SAVING”, stores the calibration values and returns to 
measurement mode.

Note

:  If a new EC calibration is performed, the NaCl calibration is automatically 

cleared. A new NaCl calibration is required.

Wrong Standard

If the reading is too far from the expected value, the 
message  “WRONG  STANDARD”  will  scroll.  Calibration 
cannot  be  confi rmed.  In  this  case  check  if  the 
calibration  solution  has  been  used  correctly  or  clean 
the probe by following the instructions for EC PROBE 
MAINTENANCE (see page 33).

NaCl  % 
Calibration

Calibration 
Messages

Содержание edge

Страница 1: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice MANEDGEEC 01 15 EC ...

Страница 2: ...nsent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Included 3 Safety Measures 3 Description 4 Diagram 4 8 Product Diagram 4 5 Probe Diagram 6 Keypad Function 7 Guide to Indicators 8 Setup Installation 9 19 Setting Up edgeEC 9 10 Electrode Probe Connections 10 General Setup 11 12 Basic Mode 13 Logging Function 13 15 Viewing Logged Data 15 17 PC Storage Interface 18...

Страница 3: ...fy damage has not occurred during shipping Remove protective film from meter Notify your nearest Hanna Customer Service Center if damage is observed Each instrument is supplied with edge EC HI2003 Bench cradle Wall cradle Electrode holder USB cable 5 Vdc Power Adapter Instruction Manual Quality Certificate HI 763100 Digital 4 ring conductivity probe with integrated temperature sensor 1413 µS cm an...

Страница 4: ...with logged data Basic mode for simplified operation Simplified data transfer to a PC Up to 8 hour battery life when used as a portable device Diagram Accessories 37 Accessories Probes HI 763100 EC temperature probe EC Conductivity Solutions HI 70030P 12880 μS cm 20 mL sachets 25 pcs HI 70031P 1413 μS cm 20 mL sachets 25 pcs HI 70039P 5000 μS cm 20 mL sachets 25 pcs HI 7030M 12880 μS cm 230 mL bot...

Страница 5: ... condensing Dimensions 202 x 140 x 12 mm 7 9 x 5 5 z 0 5 Weight 250g 8 82 oz Specifications Diagram 5 1 Liquid Crystal Display LCD 2 Capacitive Touch Keypad 3 3 mm jack input for edgeEC digital probes 4 Top mounted ON OFF button 5 Micro USB device connection for power or PC interface 6 Standard USB host connection for data transfer to a USB thumb drive Product Diagram ...

Страница 6: ...0 01 g L Accuracy 25 C 77 F 1 of reading 0 05 μS cm or 1 digit whichever is greater 1 of reading 0 03 ppm or 1 digit whichever is greater 1 of reading Calibration Single Cell Factor calibration 6 standards available 84 1413 μS cm 5 00 12 88 80 0 111 8 mS cm 1 point offset 0 00 μS cm Single point with HI 7037L Standard Temperature compensation Automatic 5 to 100º C 23 to 212º F NoTC none absolute c...

Страница 7: ...he keys is stuck Check the keyboard or contact your local Hanna Office CAL Prod message at startup Instrument was not factory calibrated or lost factory calibration Contact Hanna Technical Support for help Troubleshooting Guide Diagram 7 Keypad Function 1 CAL MODIFY Used to enter and exit calibration mode In SETUP used to initiate changes of a configuration setting 2 GLP CFM Used to display GLP ca...

Страница 8: ...sired message Options include date time calibration data battery charge or no message If a measurement error or log status change occurs during measurement the third line will display a pertinent message Maintenance 33 Rinse the probe with clean water after measurements If a more thorough cleaning is required remove the probe sleeve and clean the probe with a cloth or a nonabrasive detergent Make ...

Страница 9: ...lowing topics are expanded upon in the sections that follow in this manual 1 Familiarize yourself with the design features of this unique meter 2 Decide how the meter will be used and set up the wall or bench cradle in a clean area near line power 3 Turn edgeEC on using the ON OFF button located on the top of the meter 4 Plug in the probe required for measurement 5 SETUP the measurement parameters...

Страница 10: ...standard PSU Practical Salinity Units The practical salinity S of seawater relates the ratio of electrical conductivity of a normal seawater sample at 15 ºC and 1 atmosphere to a potassium chloride solution KCl with a mass of 32 4356 g Kg water at the same temperature and pressure Under these conditions the ratio is equal to 1 and S 35 The practical salinity scale may be applied to values to throu...

Страница 11: ... reading on the second LCD line If the reading is out of range the full scale value 100 0 g L for ATC mode or 400 0 g L for uncompensated TDS will be displayed blinking along with a message on the display Salinity Measurements Not Available In Basic Mode Press the RANGE key twice to switch from conductivity to the configured Salinity scale Verify the desired scale is configured in SETUP The meter ...

Страница 12: ...er Available Note Options that are seen under special conditions only General Setup Operational Guide 29 If enabled the number of days since the calibration expired i e CAL EXPIRED 2 DAYS AGO The serial number of the probe Note Press GLP at any moment and the instrument will return to measurement mode The RANGE key will change measurement from conductivity to TDS to Salinity Conductivity Measureme...

Страница 13: ...EC through the micro USB connector to a PC a SETUP option will require the choice LOG ON EDGE or EXPORT TO PC 1000 log records can be stored into edgeEC memory This memory is shared between all logging types Manual Manual Stability Interval logs The maximum number of records for an Interval lot is 600 records provided log space is available A record is a stored reading and a lot is a group of reco...

Страница 14: ...in will stop the Interval logging session The LOG STOPPED message will be displayed for a few seconds If a sensor failure occurs during interval logging the message OUT OF SPEC will alternate with logging information Manual Logging Select Manual in the SETUP menu To initiate a Manual log press the LOG key while the instrument is in measurement The PLEASE WAIT screen will be displayed briefly follo...

Страница 15: ...ct Stability and choose measurement stability criteria in the SETUP menu Only Stability Medium is available in Basic mode To initiate the Stability log press the LOG key while the instrument is in measurement The PLEASE WAIT screen will be displayed briefly followed by a screen showing the stability tag LOG tag and a WAITING message The log can be stopped while the WAITING message is displayed by ...

Страница 16: ...alibration solution into a beaker If possible use a plastic beaker to minimize any EMC interferences Before pressing CAL verify in SETUP Basic mode is off Salinity Scale is set to NaCl In measurement mode use the RANGE key to select the Salinity measurement The NaCl tag will be on NaCl calibration is a single point calibration at 100 0 NaCl Use the HI 7037L calibration solution sea water solution ...

Страница 17: ...278 76 72400 102100 4523 21 69 8 11910 1305 78 74000 104000 4617 22 71 6 12150 1332 79 75200 105900 4711 23 73 4 12390 1359 81 76500 107900 4805 24 75 2 12640 1386 82 78300 109800 4902 25 77 12880 1413 84 80000 111800 5000 26 78 8 13130 1440 86 81300 113800 5096 27 80 6 13370 1467 87 83000 115700 5190 28 82 4 13620 1494 89 84900 117700 5286 29 84 2 13870 1521 90 86300 119700 5383 30 86 14120 1548 ...

Страница 18: ...surement range and resolution if you know ahead of time where your samples will fall If AUTO is selected the meter will select the scale with the highest possible resolution but may change in the middle of a series of measurements changing units and displayed resolution Note The selected range is only active during measurements Auto ranging is used during calibration If a fixed range is selected a...

Страница 19: ...5 Use the pq keys to change C F cm 1 until the display reads the Custom Standard value Setup Installation 19 Logged data on the edgeEC can be transferred from the meter to a PC by following these simple directions Suitable operating systems include Windows Xp minimum OS X or Linux 1 Connect edgeEC to the PC using the supplied micro USB cable 2 Turn on edgeEC 3 Press SETUP and select LOG ON EDGE 4 ...

Страница 20: ...e Default values used Log types Manual Log on demand Manual Log on stability Fast Medium Accurate Interval Logging Manual Log on demand Manual Log on stability Medium EC Electrolytic Conductivity meter operation is configured using the SETUP key with an EC probe connected to the meter The EC specific parameters will be seen inserted into the menu If Basic mode is On the EC parameter list will be s...

Отзывы: