Setup/Installation
16
Delete Logging Type/Lot
Press
RCL
and select the parameter log.
Use the
pq
keys to select the Manual/Stability
records or Interval lots to delete. Press
CLR
. The
instrument will display “CLEAR MANUAL” for Manual
Records, “CLEAR STAB” for Stability Records.
For Interval lots, the message “CLEAR”, followed by the
selected lot will be displayed with “CFM” tag blinking.
Press the
pq
keys to select a diff erent lot. Press
CFM
.
The instrument will display “PLEASE WAIT”.
“CLEAR DONE” is displayed for a few seconds after the
selected Interval lot is deleted.
Delete Records (Manual and Stability log on demand)
To delete individual records (Manual and Stability logs only), enter Manual
(Stability) log by pressing
CFM
when Manual (Stability) is displayed. Use the
ARROW keys to select the record to be deleted and then press
CLR
.
The instrument will display “CLEAR REC.“ and record
number along with “CFM” tag blinking. Use the ARROW
keys to select another record if necessary.
Press
CFM
. The instrument will display “PLEASE WAIT” and then “CLEAR
DONE” message. When individual logs are deleted within saved MANUAL or
STABILITY logs, the logs will renumber, fi lling in the deleted data but staying in
chronological order.
To delete all records of the MANUAL (STABILITY) log, proceed as described on
page 16 for LOTS.
Select the Manual (Stability) lot and press
CLR
.
The ”CLEAR” message will be displayed along with
”MANUAL” or “STABILITY” and
CFM
tag blinking on
the LCD. Press the
CFM
key to confi rm the deleting of
the selected lot (MANUAL or STABILITY) or all records.
Press
CLR
to exit without deleting.
Viewing
Logged
Data
Operational Guide
25
Preparation
Pour a small quantity of the calibration solution into a beaker. If possible, use a
plastic beaker to minimize any EMC interferences.
Before pressing
CAL
verify in SETUP:
•
Basic mode is off .
•
Salinity Scale is set to NaCl%.
In measurement mode use the RANGE key to select the Salinity measurement.
The NaCl% tag will be on. NaCl calibration is a single point calibration at 100.0%
NaCl. Use the HI 7037L calibration solution (sea water solution) as a 100% NaCl
calibration solution.
Procedure
Rinse the probe with some of the HI 7037L calibration solution or deionized
water. Shake off excess solution. Submerse the probe in the calibration solution.
The sleeve holes must be completely submersed. If possible, center the probe in
the beaker away from the bottom or beaker walls. Raise and lower the probe to
refi ll the center cavity and tap the probe repeatedly to remove any air bubbles
that may be trapped inside the sleeve.
Press
CAL
to enter calibration. The ” “, ”STIR” and
“CAL” tags will turn on and the fi rst LCD line will
display the percent NaCl reading, the second LCD line
will display the temperature and the third LCD line will
display ”WAIT” message blinking until the reading is
stable.
When the reading is stable and close to the standard,
“CFM” tag will blink and “SOLUTION STANDARD”
message will be displayed.
Press
CFM
to confi rm calibration.
The Instrument displays “SAVING”, stores the calibration values and returns to
measurement mode.
Note
: If a new EC calibration is performed, the NaCl calibration is automatically
cleared. A new NaCl calibration is required.
Wrong Standard
If the reading is too far from the expected value, the
message “WRONG STANDARD” will scroll. Calibration
cannot be confi rmed. In this case check if the
calibration solution has been used correctly or clean
the probe by following the instructions for EC PROBE
MAINTENANCE (see page 33).
NaCl %
Calibration
Calibration
Messages