• Immerse the probe in the HI 70032 

HI 70032 

HI 70032 

HI 70032 

HI 70032 calibration

solution without exceeding the maximum im-
mersion level.

• Stir gently and wait for about 20 seconds to

confirm the reading.

• Once the display stops blinking, the meter is

calibrated and ready for use.

BATTERIES REPLACEMENT:

The symbol "Eb" on the LCD warns the user when
the batteries have to be replaced.
Open the compartment and
replace the two 1.5V batter-
ies while paying attention to
their polarity.
Batteries should be replaced only in a non hazard-
ous area using the battery type specified in this
instruction sheet.
Note

Note

Note

Note

Note: : : : : Once the batteries have been replaced, turn

the meter OFF and then ON again.  Wait for about
30 seconds to allow a complete self-test, and then
perform a new calibration.

ERROR CODES:

Error in calibration, wrong calibration solution or
damaged probe.

 Over range.

Low battery indication.

OPERATION:

• Turn the meter on by pressing the

ON/OFF button.

• Immerse the probe in the sample without

exceeding the maximum immersion level.

• Stir gently and wait for reading to stabilize.

The meter automatically compensates for tem-
perature variations.

• The TDS value of the sample will be shown on

the LCD in ppm unit.

• To display the temperature mea-

surement of the sample, press the
indicated button. The temperature
will be shown on the LCD in °C unit.

• Press the same button again to return to the

ppm measurement mode.

• To extend the batteries life, after 5 minutes of

nonuse the meter provides for auto-off.

CALIBRATION:

• Switch the meter on.
• Press and hold the ON/OFF

button for about 3 seconds.
"1382" will start blinking on
the display to indicate that
the calibration mode has
been entered.

Warranty

The 

Primo 3

 meter 

is warranted for one year

 against defects in

workmanship and materials when used for their intended purpose and

maintained according to instructions. This warranty is limited to repair or

replacement free of charge.

Damages due to accident, misuse, tampering or lack of prescribed main-

tenance are not covered.

If service is required, contact the dealer from whom you purchased the

instrument. If under warranty, report the model number, date of pur-

chase, serial number and the nature of the failure. If the repair is not

covered by the warranty, you will be notified of the charges incurred. If

the instrument is to be returned to Hanna Instruments, first obtain a

Returned Goods Authorization number from the Customer Service depart-

ment and then send it with shipping costs prepaid. When shipping any

instrument, make sure it is properly packaged for complete protection.

To validate your warranty, fill out and return the enclosed warranty card

within 14 days from the date of purchase.

Отзывы: