15. TAG INSTALLATION
The tag is housed in a rugged metal that can withstand harsh environments. However, it is better to
protect the tag from direct rain.
1. Place the tag near a sampling point.
2. Fix it securely with the provided screws.
3. The metallic iButton must be easily accessible for tag reading.
There’s no limit to the number of tags that can be installed. Additional tags can be ordered (
HI920005
- five tag holders with tags).
16. LED REPLACEMENT
In case of LED failure, “no L” error message is displayed. Replace the LED and recalibrate the meter.
1. Remove the battery cover.
2. Unscrew the LED connection using a screwdriver.
3. Unlock the LED and pull it out from the LED holder handler.
4. Place the new LED in the right position and push it until is securely locked.
5. Insert the LED leads into the connector and tighten using a screwdriver.