background image

16

17

Data transmission from the instrument to the PC can be done with

the 

HI 92000

 Windows

®

 compatible software (optional). 

HI 92000

also offers graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for

further analysis.
To connect your instrument to a PC, use the optional Hanna 

HI 920010

cable connector. Make sure that your instrument is switched off and

plug one  connector to the instrument RS232C socket and the other into

the serial port of your PC.

Notes:

• Other cables than 

HI 920010

 may use a different

configuration. In this case communication between in-

strument and PC may not be possible.

• If you are not using Hanna Instruments 

HI 92000

software, please see the following instructions.

SETTING THE BAUD RATE AND THE COMMAND

SETTING THE BAUD RATE AND THE COMMAND

SETTING THE BAUD RATE AND THE COMMAND

SETTING THE BAUD RATE AND THE COMMAND

SETTING THE BAUD RATE AND THE COMMAND
PREFIX

PREFIX

PREFIX

PREFIX

PREFIX

• To set the instrument baud rate, press 

COMM

while in measurement mode. The primary LCD
will show the current 

baud rate

.

• Use the 

ARROW

 keys to change the displayed value (150; 300;

600; 1200; 2400 – factory setting; 4800; 9600).

• Press 

CFM

 to confirm. The primary LCD will show the current

command prefix

 (factory setting is 16).

• Use the 

ARROW

 keys to change the displayed value (between 0

and 47).

• Press 

CFM

 to confirm. The meter returns to

measurement mode.

Note:

The command prefix must not be changed if

HI 92000 

PC software is used.

PC INTERFACE

PC INTERFACE

PC INTERFACE

PC INTERFACE

PC INTERFACE

(p

(p

(p

(p

(pH 212 & 

H 212 & 

H 212 & 

H 212 & 

H 212 & pppppH 213 

H 213 

H 213 

H 213 

H 213 only

only

only

only

only)))))

      m

m

m

m

mV CALIBRATION (

V CALIBRATION (

V CALIBRATION (

V CALIBRATION (

V CALIBRATION (pppppH 213 

H 213 

H 213 

H 213 

H 213 only

only

only

only

only)))))

(for technical personnel only)

C

C

CFM

COMM

All the instruments are factory calibrated for mV.
Hanna’s ORP electrodes are interchangeable and no mV calibration
is needed when they are replaced.
If the mV measurements are inaccurate, mV recalibration should be
performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions below.
A two or three-point calibration can be performed at 0.0 mV, 600.0
mV and 1800.0 mV.

• Attach to the BNC connector a mV simulator with an accuracy of

±0.1 mV.

• With the instrument off, press and hold down the 

MR&CFM

keys, then power on the instrument. The “CAL” tag will appear
and the secondary LCD will show 0.0 mV.

• Set 0.0 mV on the simulator.

When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.

• Press 

CFM

 to confirm. The secondary LCD will display 600 mV.

• Set 600.0 mV on the simulator.

When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.

• Press 

CFM

 to confirm. The secondary LCD will display 1800 mV.

• Set 1800.0 mV on the simulator.

When the reading is stable and close to the selected calibration
point, “READY” tag will appear and “CFM” tag will blink.

• Press 

CFM

 to confirm. The 

instrument 

returns to measurement

mode.

Notes:

• If the reading is not close to the selected calibration point,

“WRONG” tag will blink. Verify calibration condition or
contact your vendor if you can not calibrate.

• Press 

CAL 

in any moment of the calibration process. The

instrument will return to measurement mode. If calibra-
tion process is stopped after 600 mV is confirmed, the
600 mV range is calibrated and calibration parameters
are memorized.

Содержание pH 211

Страница 1: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL Instruction Ma...

Страница 2: ...your Dealer or the nearest Hanna Customer Service Center Each instrument is supplied with HI 1131B Glass body Combination pH Electrode HI 7669 2W Temperature Probe HI 76404 Electrode Holder pH 4 01 7...

Страница 3: ...1 2 H p V m 1 C 1 0 Y C A R U C C A F 8 6 C 0 2 H p 1 0 0 V m 2 0 V m 9 9 9 3 V m 1 V m 0 0 0 2 C 0 0 0 1 0 0 C 5 0 1 e d i s t u o C r o r r e e b o r p g n i d u l c x e n o i t a r b i l a C H p s...

Страница 4: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL E G N A R H p...

Страница 5: ...on the secondary LCD If the pH reading is out of range will be displayed on the LCD If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly wit...

Страница 6: ...ccuracy is required The instrument should be recalibrated Whenever the pH electrode is replaced At least once a week After testing aggressive chemicals PREPARATION PREPARATION PREPARATION PREPARATION...

Страница 7: ...rn to measurement mode Notes If the value measured by the meter is not close to the selected buffer WRONG and WRONG tags will blink alternately In this case check if the correct buffer has been used o...

Страница 8: ...er With the instrument off press and hold down the CAL MEM keys then power on the instrument The CAL tag will appear and the secondary LCD will show 0 0 C Immerse the temperature probe in the vessel w...

Страница 9: ...LIBRATION V CALIBRATION V CALIBRATION V CALIBRATION V CALIBRATION p p p p pH 213 H 213 H 213 H 213 H 213 only only only only only for technical personnel only C C CFM COMM All the instruments are fact...

Страница 10: ...set to pH Err 6 is sent TM Causes the instrument to send the temperature reading Err 3 is sent if out of range Notes Either small or capital letters can be used The characters sent by the meter are a...

Страница 11: ...Storage Solution or in its absence Filling Solution HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction electrodes Follow the Preparation Procedure on page 20 before taki...

Страница 12: ...r r o f y l n o e h t n i e g n a r f o t u O e l a c s V m y r D n o i t c n u j e n a r b m e m r o 0 0 3 0 7 I H n i k a o S e g a r o t s 0 0 3 0 8 I H n o i t u l o s t o n s e o d r e t e m e h...

Страница 13: ...FDA approved bottle 4x50 mL for single junction electrodes HI 8072 1M KNO3 Electrolyte in FDA approved bottle 4x50 mL HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x50 mL Typical Electrode Lif...

Страница 14: ...DIA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution 460 mL bottle HI 7092L Oxidizing Pretreatment Solution 460 mL bottle pH ELECTRODES All electrodes part num...

Страница 15: ...12 mm 0 5 S VERSION HI 3131 120 mm 4 7 12 mm 0 5 HI 3230 S VERSION 120 mm 4 7 12 mm 0 5 HI 4430 S VERSION FC 100B Plastic body Kynar double junction refillable combination pH electrode Use general pu...

Страница 16: ...E ELECTRODES CONNECT TO THE BNC SOCKET OF THE METER HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3 0 mm 0 12 CABLE WITH BNC Windows is registered Trademark of Microsoft Co RECOMMENDATIONS FOR USERS B...

Страница 17: ...02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739...

Отзывы: