• When the reading is stable and close to the selected buffer,

“READY” tag will appear and “CFM” tag will blink.

• Press 

CFM

 to confirm calibration.

• The calibrated value is then displayed on the

primary LCD and the secondary LCD will display
the second expected buffer value.

Note: 

The instruments will automatically skip the buffer used for the

first point. It also skips 6.86 if 7.01 buffer was used and

viceversa. Likewise, it will skip 9.18 if 10.01 buffer was used and

viceversa.

• After the first calibration point is confirmed,

immerse the pH electrode and the temperature

probe approximately 4 cm (1½”) into the

second buffer solution and stir gently. The

temperature probe should be close to the pH

electrode.

• If necessary, press the 

ARROW

 keys to select a different buffer value.

• The “ ” tag will blink on the LCD until the reading is stable.
• When the reading is stable, “READY” tag will appear and “CFM”

tag will blink.

• Press 

CFM

 to confirm calibration. The instrument

will return to measurement mode.

Notes:

• If the value measured by the meter is not close to the

selected buffer, “WRONG  ” and “WRONG  ” tags will
blink alternately. In this case check if the correct buffer
has been used, or regenerate the electrode by following
the cleaning procedure (see page 22). If necessary,
change the buffer or the electrode.

• The “WRONG” tag and temperature value are displayed

blinking if the  temperature reading is out of the defined
temperature range of the buffer. Calibration cannot be
confirmed in this situation.

C

C

• Press 

RANGE

 to display the tem-

perature reading on the LCD during
calibration (

pH 211

pH 212

pH

213

 only).

ONE-POINT CALIBRATION

ONE-POINT CALIBRATION

ONE-POINT CALIBRATION

ONE-POINT CALIBRATION

ONE-POINT CALIBRATION

• Proceed as described in “Two-point calibration” section.
• Press 

CAL

 after the first calibration point

was confirmed.

The instrument will return to measurement mode
and will memorize the one-point calibration data.

Note:

Press and hold down 

CFM

 and then press 

CAL

. The instrument

sets default calibration parameters, displays “CLR” message for a
few seconds and then returns to normal measurement mode.

CFM

CFM

CAL

CAL

2

BUF

pH

BUFFER pH

Содержание pH 211

Страница 1: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL Instruction Ma...

Страница 2: ...your Dealer or the nearest Hanna Customer Service Center Each instrument is supplied with HI 1131B Glass body Combination pH Electrode HI 7669 2W Temperature Probe HI 76404 Electrode Holder pH 4 01 7...

Страница 3: ...1 2 H p V m 1 C 1 0 Y C A R U C C A F 8 6 C 0 2 H p 1 0 0 V m 2 0 V m 9 9 9 3 V m 1 V m 0 0 0 2 C 0 0 0 1 0 0 C 5 0 1 e d i s t u o C r o r r e e b o r p g n i d u l c x e n o i t a r b i l a C H p s...

Страница 4: ...ffer nd Confirm 1 Buffer st CFM Select 1 Buffer st Start Calibration CAL COMM RS232 Communication MR Memory Recall Select pH ORP RANGE MEM Memorize Reading MR MEM COMM C C RANGE CFM CAL E G N A R H p...

Страница 5: ...on the secondary LCD If the pH reading is out of range will be displayed on the LCD If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly wit...

Страница 6: ...ccuracy is required The instrument should be recalibrated Whenever the pH electrode is replaced At least once a week After testing aggressive chemicals PREPARATION PREPARATION PREPARATION PREPARATION...

Страница 7: ...rn to measurement mode Notes If the value measured by the meter is not close to the selected buffer WRONG and WRONG tags will blink alternately In this case check if the correct buffer has been used o...

Страница 8: ...er With the instrument off press and hold down the CAL MEM keys then power on the instrument The CAL tag will appear and the secondary LCD will show 0 0 C Immerse the temperature probe in the vessel w...

Страница 9: ...LIBRATION V CALIBRATION V CALIBRATION V CALIBRATION V CALIBRATION p p p p pH 213 H 213 H 213 H 213 H 213 only only only only only for technical personnel only C C CFM COMM All the instruments are fact...

Страница 10: ...set to pH Err 6 is sent TM Causes the instrument to send the temperature reading Err 3 is sent if out of range Notes Either small or capital letters can be used The characters sent by the meter are a...

Страница 11: ...Storage Solution or in its absence Filling Solution HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction electrodes Follow the Preparation Procedure on page 20 before taki...

Страница 12: ...r r o f y l n o e h t n i e g n a r f o t u O e l a c s V m y r D n o i t c n u j e n a r b m e m r o 0 0 3 0 7 I H n i k a o S e g a r o t s 0 0 3 0 8 I H n o i t u l o s t o n s e o d r e t e m e h...

Страница 13: ...FDA approved bottle 4x50 mL for single junction electrodes HI 8072 1M KNO3 Electrolyte in FDA approved bottle 4x50 mL HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x50 mL Typical Electrode Lif...

Страница 14: ...DIA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution 460 mL bottle HI 7092L Oxidizing Pretreatment Solution 460 mL bottle pH ELECTRODES All electrodes part num...

Страница 15: ...12 mm 0 5 S VERSION HI 3131 120 mm 4 7 12 mm 0 5 HI 3230 S VERSION 120 mm 4 7 12 mm 0 5 HI 4430 S VERSION FC 100B Plastic body Kynar double junction refillable combination pH electrode Use general pu...

Страница 16: ...E ELECTRODES CONNECT TO THE BNC SOCKET OF THE METER HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3 0 mm 0 12 CABLE WITH BNC Windows is registered Trademark of Microsoft Co RECOMMENDATIONS FOR USERS B...

Страница 17: ...02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739...

Отзывы: