• No Slope:
- Check the electrode for cracks in glass stem or bulb
(replace the electrode if cracks are found).
- Make sure cable and connections are not damaged
nor lying in a pool of water or solution.
• Slow Response/Excessive Drift: soak the tip in
HI7061
solution for 30 minutes, rinse thoroughly in distilled water
and then follow the Cleaning Procedure above.
• For ORP electrodes: polish the metal tip with a lightly
abrasive paper (paying attention not to scratch the surface)
and wash thoroughly with water.
Note
With industrial applications, it is always recommended to
keep at least one spare electrode handy. When anomalies
are not resolved with a simple maintenance, change the
electrode (and recalibrate the controller) to see if the problem
is alleviated.