Note
Cables other than
HI920010
may use a different configu-
ration. In these cases, no communication between the meter
and the PC is possible. If you are not using the
HI920010
cable, contact the nearest Hanna Instruments Customer
Service Center or proceed as follows for a proper electrical
connection:
pH500/ mV600
9-pin DSUB male cennector
PC
9-pin DSUB female connector
Pin 2
Pin 3 (Txd)
Pin 3
Pin 2 (Rxd)
Pin 4
Pin 6 (Txd)
Pin 5
Pin 5 (Gnd)
Pin 6
Pin 4 (DTR)
Pin 7 short circuit with 8 (RTS + CTS)
pH500/ mV600
9-pin DSUB male cennector
PC
25-pin DSUB female connector
Pin 2
Pin 2 (Txd)
Pin 3
Pin 3 (Rxd)
Pin 4
Pin 6 (Txd)
Pin 5
Pin 7 (Gnd)
Pin 6
Pin 20 (DTR)
Pin 4 short circuit with 5 (RTS + CTS)
Excel
©
Copyright of “Microsoft Co.”
Lotus 1-2-3
©
Copyright of “Lotus Co.”
Windows
®
and Windows Terminal
®
are registered Trademark of “Microsoft Co.”
SETTING THE BAUD RATE
The transmission speed (baud rate) of the meter and the
external device must be identical.
The meter is factory set to 4800. If you wish to change this
value, use item 71 in the setup mode (see page 14).