34
OUR Configuration
Highlight
OUR configuration.
Press
Select
.
Parameters:
•
Min time
‑ the minimum time for the OUR test.
Range: 1 to 3600 seconds.
•
Max time
‑ the maximum time for the OUR test. The test will stop automatically when the maximum
time has elapsed.
Range: 1 to 3600 seconds.
•
Min start
DO
‑ the minimum accepted DO value for starting the OUR test. If the DO reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
•
Min end DO
‑ the minimum accepted DO value at the end of the test. If the DO reading at the
end of the OUR test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
•
Total volume
‑ the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
•
Sample volume
‑ the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
Press
Prev/Next
to select a different parameter.
Press
/
keys to modify the selected parameter’s value.
Press
Save
to save the new OUR configuration.
Press
ESC
to leave without changing.