13 Setup
Temperature Alarm
(Options /
Default
)
Temperature High Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
T E M P H I G
dIS
T E M P H I G
“
TEMP HIGH
” is scrolled on the bottom of the LCD.
Temperature High Value
0.1 to 50.0 °C*
(26.6 °C)
32.1 to 122.0 °F*
(80.0 °F)
26.6
T E M P H I G
28.6
T E M P H I G
With Temperature High Alarm configured as enabled “
TEMP
HIGH
VALUE
” is scrolled on the LCD.
Actual high alarm temperature value (editable) is displayed blinking.
Press
e
to enter edit mode and use
a
/
v
to change value. Press
e
to save.
Temperature Low Alarm
Enabled (En), Disabled (dIS)
(Enabled)
En
T E M P L O W
dIS
T E M P L O W
Temperature Low Value
0.0 to 49.9 °C*
(23.3 °C)
32.0 to 121.9 °F*
(74.0 °F)
23.3
T E M P L O W
25.3
T E M P L O W
With Temperature Low Alarm configured as enabled “
TEMP
LOW
VALUE
” is scrolled on the LCD.
Actual low alarm temperature value (editable) is displayed blinking.
Press
e
to enter edit mode and use
a
/
v
to change value. Press
e
to save.